Diamond detector in absorbed dose measurements in high-energy linear accelerator photon and electron beams

被引:21
|
作者
Ravichandran, Ramamoorthy [1 ]
Binukumar, John Pichy [1 ]
Al Amri, Iqbal [1 ]
Davis, Cheriyathmanjiyil Antony [1 ]
机构
[1] Royal Hosp, Natl Oncol Ctr, Dept Radiat Oncol, Med Phys Unit, P Box 1331, Muscat 111, Oman
来源
关键词
diamond detectors; small field RT; absorbed doses; stereotactic radiotherapy; DOSIMETRY;
D O I
10.1120/jacmp.v17i2.5690
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Diamond detectors (DD) are preferred in small field dosimetry of radiation beams because of small dose profile penumbras, better spatial resolution, and tissue-equivalent properties. We investigated a commercially available 'microdiamond' detector in realizing absorbed dose from first principles. A microdiamond detector, type TM 60019 with tandem electrometer is used to measure absorbed doses in water, nylon, and PMMA phantoms. With sensitive volume 0.004 mm(3), radius 1.1 mm, thickness 1 x10(-3) mm, the nominal response is 1 nC/Gy. It is assumed that the diamond detector could collect total electric charge (nC) developed during irradiation at 0 V bias. We found that dose rate effect is less than 0.7% for changing dose rate by 500 MU/min. The reproducibility in obtaining readings with diamond detector is found to be +/- 0.17% (1 SD) (n = 11). The measured absorbed doses for 6 MV and 15 MV photons arrived at using mass energy absorption coefficients and stopping power ratios compared well with N-d, water calibrated ion chamber measured absorbed doses within 3% in water, PMMA, and nylon media. The calibration factor obtained for diamond detector confirmed response variation is due to sensitivity due to difference in manufacturing process. For electron beams, we had to apply ratio of electron densities of water to carbon. Our results qualify diamond dosimeter as a transfer standard, based on long-term stability and reproducibility. Based on micro-dimensions, we recommend these detectors for pretreatment dose verifications in small field irradiations like stereotactic treatments with image guidance.
引用
收藏
页码:291 / 303
页数:13
相关论文
共 50 条
  • [21] The equivalence of the absorbed dose to water in high energy photon beams
    Stancu, E.
    Kapsch, R-P.
    Badita, E. S.
    Scarlat, F.
    Scarisoreanu, A. M.
    OPTOELECTRONICS AND ADVANCED MATERIALS-RAPID COMMUNICATIONS, 2018, 12 (3-4): : 237 - 239
  • [22] Dosimetric properties of radiophotoluminescent glass rod detector in high-energy photon beams from a linear accelerator and Cyber-Knife
    Araki, F
    Moribe, N
    Shimonobou, T
    Yamashita, Y
    MEDICAL PHYSICS, 2004, 31 (07) : 1980 - 1986
  • [23] EFFECT OF AIR CAVITIES ON DOSE DISTRIBUTION OF HIGH-ENERGY ELECTRON AND PHOTON BEAMS
    NUSSLIN, F
    PHYSICS IN MEDICINE AND BIOLOGY, 1977, 22 (03): : 574 - 574
  • [24] METHODS OF IMPROVING DOSE UNIFORMITY IN HIGH-ENERGY PHOTON AND ELECTRON-BEAMS
    BRAHME, A
    PHYSICS IN MEDICINE AND BIOLOGY, 1977, 22 (03): : 565 - 565
  • [25] A COMPARISON OF THE AAPM PROTOCOL FOR THE DETERMINATION OF ABSORBED DOSE FROM HIGH-ENERGY PHOTON AND ELECTRON-BEAMS WITH CURRENTLY USED PROTOCOLS
    HUNT, MA
    MALIK, S
    THOMASON, C
    MASTERSON, ME
    MEDICAL PHYSICS, 1984, 11 (06) : 806 - 813
  • [26] CONSISTENCY AND SIMPLICITY IN THE DETERMINATION OF ABSORBED DOSE TO WATER IN HIGH-ENERGY PHOTON BEAMS - A NEW CODE OF PRACTICE
    MIJNHEER, BJ
    AALBERS, AHL
    VISSER, AG
    WITTKAMPER, FW
    RADIOTHERAPY AND ONCOLOGY, 1986, 7 (04) : 371 - 384
  • [27] Changes in absorbed dose to water for high-energy electron beams by beam quality correction factor
    Kinoshita, N.
    Takemura, A.
    Toi, A.
    Kita, A.
    Nishimoto, Y.
    Adachi, T.
    Shioura, H.
    Kimura, H.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S56 - S56
  • [28] ABSORBED DOSE AND CAVITY IONIZATION FOR HIGH ENERGY ELECTRON BEAMS
    KESSARIS, N
    RADIATION RESEARCH, 1970, 43 (01) : 235 - &
  • [29] A new procedure for dose measurements in mixed high-energy electron and photon fields
    Gargioni, E
    Büermann, L
    Kramer, HM
    RADIATION PHYSICS AND CHEMISTRY, 2001, 61 (3-6) : 455 - 456
  • [30] DETERMINATION OF THE SOURCE POSITION FOR THE ELECTRON-BEAMS FROM A HIGH-ENERGY LINEAR-ACCELERATOR
    JAMSHIDI, A
    KUCHNIR, FT
    REFT, CS
    MEDICAL PHYSICS, 1986, 13 (06) : 942 - 948