Development of a Ring PET Insert for MRI

被引:0
作者
Bauer, Chris [1 ]
Stolin, Alexander [1 ]
Proffitt, James [2 ]
Martone, Peter [1 ]
Brefczynski-Lewis, Julie [1 ]
Lewis, James [1 ]
Hankiewicz, Janusz [3 ]
Raylman, Ray [1 ]
Majewski, Stan [1 ]
机构
[1] W Virginia Univ, Morgantown, WV 26505 USA
[2] Ait Technol, Newport News, VA 23606 USA
[3] Univ Colorado, Colorado Springs, CO 80918 USA
来源
2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC) | 2013年
关键词
SYSTEM; HAT;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Our research group constructed a 12-module PET detection ring composed of Hamamatsu multi-pixel photon counter (MMPC) silicon photomultiplier (SiPM) detectors placed in a ring that is fully MRI compatible. This brain imager can be placed around the head of a patient (clinical setting) or subject (research setting) and allow for comfortable upright imaging. However, an alternative way to use this device, as enabled by the technology, is to indeed scan individuals in the supine position in conjuncture with current MRI systems. This PET prototype is able to image simultaneously as the MRI scan is occurring, thus maximizing co-registration and the accuracy of the assignment of metabolically active voxels to their anatomically correct counterparts as identified by the MR image. In this study, we first conducted some basic instrumentation tests to ensure the device was functioning properly outside of the MRI, and additionally scan some phantoms (Derenzo; Hoffman Brain) to assess the quality in which our brain imager is able to produce adequate images. After these initial studies, we conducted multiple different experiments inside a 3 Tesla MRI in order to see how the magnetic field would influence the operation of the PET imager and vice versa. Through simultaneous PET/MRI scanning of a Hoffman brain phantom filled with F-18 radioactivity and water, it was shown that the quality of the MR image was largely unaffected by the PET imager. Furthermore, although the quality of PET imaging was affected by the RF pulsing, an acceptable PET image was nevertheless produced. As we discuss following the results, the success of this study shows that our brain imager is indeed MR compatible, and that next generation devices based on its concepts will continue to improve combined PET/MRI functionality. As two of the major advantages of this imager are the potential for low-dose scanning and its adaptability to any MRI scanner, this study suggests that PET/MRI brain imaging with low dose is in principle possible using an insert which could be adapted to any MRI scanner, using standard RF coils for that scanner.
引用
收藏
页数:9
相关论文
共 50 条
  • [41] Evaluation of RF shielding materials to improve MR image of hybrid PET/MRI
    Lim, Hyun Keong
    Choi, Yong
    Jung, Jin Ho
    Jung, Jiwoong
    Oh, Changheun
    Park, Hyun-wook
    Kim, Jong Guk
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2022, 80 (07) : 640 - 646
  • [42] PET/MRI: Paving the Way for the Next Generation of Clinical Multimodality Imaging Applications
    Pichler, Bernd J.
    Kolb, Armin
    Naegele, Thomas
    Schlemmer, Heinz-Peter
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51 (03) : 333 - 336
  • [43] Development of the Helmet-Chin PET Prototype
    Tashima, Hideaki
    Yoshida, Eiji
    Nishikido, Fumihiko
    Wakizaka, Hidekatsu
    Nitta, Munetaka
    Ahmed, Abdella M.
    Mohammadi, Akram
    Tazawa, Shusaku
    Kimura, Yasuyuki
    Suhara, Tetsuya
    Fujibayashi, Yasuhisa
    Yamaya, Taiga
    2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,
  • [44] Dual layer dol detector modules for a dedicated mouse brain PET/MRI
    Parl, C.
    Kolb, A.
    Stricker-Shaver, D.
    Pichler, B. J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (05)
  • [45] Comparison of Standardized Uptake Values in Normal Structures Between PET/CT and PET/MRI in a Tertiary Pediatric Hospital: A Prospective Study
    Lyons, Karen
    Seghers, Victor
    Sorensen, James I. L.
    Zhang, Wei
    Paldino, Michael J.
    Krishnamurthy, Rajesh
    Rohren, Eric M.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 205 (05) : 1094 - 1101
  • [46] Diagnostic value of integrated 18F-PSMA-1007 PET/MRI compared with that of biparametric MRI for the detection of prostate cancer
    Zeng, Yuping
    Leng, Xiaoming
    Liao, Hengbin
    Jiang, Guihua
    Chen, Ping
    PROSTATE INTERNATIONAL, 2022, 10 (02) : 108 - 116
  • [47] Theoretical and Numerical Analysis of the Single-Ring OpenPET geometry for in-beam PET
    Tashima, Hideaki
    Hirano, Yoshiyuki
    Kinouchi, Shoko
    Yoshida, Eiji
    Ito, Hiroshi
    Yamaya, Taiga
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 3871 - 3873
  • [48] Partial-ring PET image restoration using a deep learning based method
    Liu, Chih-Chieh
    Huang, Hsuan-Ming
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (22)
  • [49] The white elephant revived: A new marriage between PET and MRI Comment to Cumming: "PET Neuroimaging: The White Elephant Packs His Trunk?"
    Siebner, Hartwig R.
    Strafella, Antonio P.
    Rowe, James B.
    NEUROIMAGE, 2014, 84 : 1104 - 1106
  • [50] FPGA-based RF interference reduction techniques for simultaneous PET-MRI
    Gebhardt, P.
    Wehner, J.
    Weissler, B.
    Botnar, R.
    Marsden, P. K.
    Schulz, V.
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (09) : 3500 - +