Sensitive high-resolution analysis of biological molecules by capillary zone electrophoresis coupled with reflecting time-of-flight mass spectrometry

被引:38
|
作者
McComb, ME
Krutchinsky, AN
Ens, W
Standing, KG
Perreault, H [1 ]
机构
[1] Univ Manitoba, Dept Chem, Winnipeg, MB R3T 2N2, Canada
[2] Univ Manitoba, Dept Phys, Winnipeg, MB R3T 2N2, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
capillary electrophoresis mass spectrometry; interfaces; CE-MS; electrospray ionization; detection; electrophoresis; peptides; proteins;
D O I
10.1016/S0021-9673(97)01158-8
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Off-line and on line capillary zone electrophoresis-electrospray ionization time-of-flight mass spectrometry (CZE-ESI-TOF-MS) experiments were conducted using uncoated fused-silica capillaries coupled to a reflecting TOF mass spectrometer via a gold-coated sheathless interface. Off-line and on-line experiments were performed on standard mixtures of proteins and peptides. Samples collected off-line electrokinetically in plastic vials were analyzed by standard ESI-TOF-MS at the pmol level. Sheathless CZE-ESI-TOF-MS was first simulated in an off-line experiment, using a test bench, in order to select a suitable running electrolyte, to find the optimal electrospray potential, and also to test the gold-coated capillary tips. This enabled an ease of transition to on-line measurements. On-line CZE-ESI-TOF-MS measurements of the total ion electropherogram (TIE) and of selected ion electropherograms (SIE) on peptide mixtures demonstrated fmol-level sensitivity, with S/N values of 250-400 on raw data (TIE mode) and of 30-760 (SIE mode). The use of reflecting TOF-MS afforded mass resolution values R>6000 (m/Delta m(FWHM)) and enabled isotopic resolution of peptide components as well as mass accuracy in the 10 ppm range. These results were comparable with values observed with the usual ESI source on the same mass spectrometer, and thus demonstrated no loss in spectral quality from using the sheathless CE interface. On-line CE separation efficiency was equivalent to that obtained off-line for the separation of a peptide mixture, with N=35000-87000 theoretical plates. Separations of standard proteins yielded equivalent mass spectral resolution and accuracy with separation efficiencies of N=2800-5500 and S/N values of 110-225 on raw data. The gold-coated sheathless CE-ESI interface was found to be relatively easy to prepare with the use of gold vapour deposition. The interface produced a stable electrospray for extended periods of time and was found to be robust and reliable. (C) Elsevier Science B.V.
引用
收藏
页码:1 / 11
页数:11
相关论文
共 50 条
  • [1] High-resolution mass spectrometry of large molecules in a linear time-of-flight mass spectrometer
    Kinsel, G.R.
    Grundwuermer, J.M.
    Grotemeyer, J.
    Journal of the American Society for Mass Spectrometry, 1993, 4 (01):
  • [2] Recent Advances in the Use of Capillary Electrophoresis Coupled to High-Resolution Mass Spectrometry for the Analysis of Small Molecules
    Klampfl, Christian W.
    Buchberger, Wolfgang
    CURRENT ANALYTICAL CHEMISTRY, 2010, 6 (02) : 118 - 125
  • [3] Analysis of flavonoids in citrus fruits by capillary zone electrophoresis coupled with quadrupole time-of-flight mass spectrometry using chemometrics
    Zhen, Xiao-Ting
    Zhu, Si-Chen
    Shi, Min-Zhen
    Yu, Ya-Ling
    Yan, Tian-Ci
    Yue, Zi-Xuan
    Gu, Yu-Xin
    Zheng, Hui
    Cao, Jun
    JOURNAL OF FOOD COMPOSITION AND ANALYSIS, 2022, 106
  • [4] Metabolomics Platform with Capillary Electrophoresis Coupled with High-Resolution Mass Spectrometry for Plasma Analysis
    Sasaki, Kazunori
    Sagawa, Hitoshi
    Suzuki, Makoto
    Yamamoto, Hiroyuki
    Tomita, Masaru
    Soga, Tomoyoshi
    Ohashi, Yoshiaki
    ANALYTICAL CHEMISTRY, 2019, 91 (02) : 1295 - 1301
  • [5] High throughput analysis of tryptophan metabolites in a complex matrix using capillary electrophoresis coupled to time-of-flight mass spectrometry
    Arvidsson, Bjorn
    Johannesson, Nina
    Citterio, Attilio
    Righetti, Pier Giorgio
    Bergquist, Jonas
    JOURNAL OF CHROMATOGRAPHY A, 2007, 1159 (1-2) : 154 - 158
  • [6] Capillary electrophoresis time-of-flight mass spectrometry of an opium powder
    Lazar, IM
    Naisbitt, G
    Lee, ML
    CHROMATOGRAPHIA, 1999, 50 (3-4) : 188 - 194
  • [7] Capillary electrophoresis time-of-flight mass spectrometry of an opium powder
    I. M. Lazar
    G. Naisbitt
    M. L. Lee
    Chromatographia, 1999, 50 : 188 - 194
  • [8] HIGH-RESOLUTION MASS-SPECTROMETRY OF LARGE MOLECULES IN A LINEAR TIME-OF-FLIGHT MASS-SPECTROMETER
    KINSEL, GR
    GRUNDWUERMER, JM
    GROTEMEYER, J
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1993, 4 (01) : 2 - 10
  • [9] Position sensitive detection coupled to high-resolution time-of-flight mass spectrometry: Imaging for molecular beam deflection experiments
    Abd El Rahim, M
    Antoine, R
    Arnaud, L
    Barbaire, M
    Broyer, M
    Clavier, C
    Compagnon, I
    Dugourd, P
    Maurelli, J
    Rayane, D
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (12): : 5221 - 5227
  • [10] A high-resolution time-of-flight mass spectrometer for the detection of ultracold molecules
    S.D. Kraft
    J. Mikosch
    P. Staanum
    J. Deiglmayr
    J. Lange
    A. Fioretti
    R. Wester