Use of rapid influenza diagnostic tests under field conditions as a screening tool during an outbreak of the 2009 novel influenza virus: Practical considerations

被引:26
作者
Sambol, Anthony R. [1 ]
Abdalhamid, Baha [2 ]
Lyden, Elizabeth R. [3 ]
Aden, Tricia A. [4 ]
Noel, Rhonda K. [1 ]
Hinrichs, Steven H. [1 ]
机构
[1] Univ Nebraska, Med Ctr, Dept Pathol & Microbiol, Omaha, NE 68198 USA
[2] King Fahad Specialist Hosp, Dammam, Saudi Arabia
[3] Univ Nebraska, Med Ctr, Coll Publ Hlth, Omaha, NE 68198 USA
[4] Assoc Publ Hlth Labs, Silver Spring, MD USA
关键词
Luminex RVP; RIDT; BinaxNOW; TruFlu; QuickVue; Xpect; Specificity; Sensitivity;
D O I
10.1016/j.jcv.2009.12.015
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
Background: Rapid influenza diagnostic tests (RIDTs) are used in various settings as a first-line screen of patient specimens. During the initial outbreak of the 2009 novel influenza A/H1N1 virus, the Nebraska Public Health Laboratory (NPHL) adopted a testing algorithm, attempting to maximize the usefulness of RIDTs. However, it became apparent that a high percentage of the positive specimens received from off-site facilities were negative for influenza viruses by the confirmatory test, the Luminex xTAG Respiratory Viral Panel (RVP) molecular assay. Objectives: To explore the cause of discrepancies between RIDTs results obtained from on-site facility testing versus confirmatory testing performed at NPHL. Study design: Specimens (n = 336) tested with RIDTs at off-site facilities and screened for high-probability of containing H1N1 were sent to the NPHL for confirmatory testing by RVP. Results: Of 336 specimens analyzed, 104 were negative for influenza A or B by both RIDT and RVP; 127 were positive by both tests; 102 were positive by RIDT only; and 3 were positive by RVP only. Using the RVP assay as the gold standard, overall RIDT characteristics in this screened population were: sensitivity = 97.7% (95% CI: 92.5, 99.3); specificity = 48.1% (95% CI: 40.4, 55.8); positive predictive value = 54.3% (95% CI: 47.0, 61.4); and negative predicative value = 97.1% (95% CI: 90.6, 99.1). Conclusions: The results show that the confirmation of RIDT-positive results varied widely by testing site. Possible explanations for the discrepancies in performance characteristics include testing a narrowly defined sample population, test facility characteristics, facility work load, and seasonal timing. (C) 2009 Elsevier B. V. All rights reserved.
引用
收藏
页码:229 / 233
页数:5
相关论文
共 10 条
[1]  
*CDCP, MMWR MORB MORTAL WKL, V58, P826
[2]  
Centers for Disease Control and Prevention, 2009, CDC PROT REAL TIM RT
[3]  
Centers for Disease Control and Prevention (CDC), 2009, MMWR Morb Mortal Wkly Rep, V58, P467
[4]  
FAIX DJ, 2009, NEW ENGL J MED, V10, P1
[5]   Clinical performance of the 3M Rapid Detection Flu A plus B Test compared to R-Mix culture, DFA and BinaxNOW Influenza A&B Test [J].
Ginocchio, Christine C. ;
Lotlikar, Madhavi ;
Falk, Leon ;
Arora, Suman ;
Kowerska, Margaret ;
Bornfreund, Mark ;
Manji, Ryhana .
JOURNAL OF CLINICAL VIROLOGY, 2009, 45 (02) :146-149
[6]   Evaluation of multiple test methods for the detection of the novel 2009 influenza A (H1N1) during the New York City outbreak [J].
Ginocchio, Christine C. ;
Zhang, Frank ;
Manji, Ryhana ;
Arora, Suman ;
Bornfreund, Mark ;
Falk, Leon ;
Lotlikar, Madhavi ;
Kowerska, Margaret ;
Becker, George ;
Korologos, Diamanto ;
de Geronimo, Marcella ;
Crawford, James M. .
JOURNAL OF CLINICAL VIROLOGY, 2009, 45 (03) :191-195
[7]   Performance of six influenza rapid tests in detecting human influenza in clinical specimens [J].
Hurt, Aeron C. ;
Alexander, Robert ;
Hibbert, Jan ;
Deed, Nicola ;
Barr, Ian G. .
JOURNAL OF CLINICAL VIROLOGY, 2007, 39 (02) :132-135
[8]  
Rahman MM, 2007, MILCHWISSENSCHAFT, V62, P6
[9]   Comparison of the NOW Influenza A & B, NOW Flu A, NOW Flu B, and directigen Flu A+B assays, and immunofluorescence with viral culture for the detection of Influenza A and B viruses [J].
Smit, Marita ;
Beynon, Kirsten A. ;
Murdoch, David R. ;
Jennings, Lance C. .
DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2007, 57 (01) :67-70
[10]   Evaluation of three immunoassay kits for rapid detection of influenza virus A and B [J].
Weinberg, A ;
Walker, ML .
CLINICAL AND DIAGNOSTIC LABORATORY IMMUNOLOGY, 2005, 12 (03) :367-370