Changes in adhesive ingredients in continuous glucose monitoring systems may induce new contact allergy pattern

被引:22
作者
Svedman, Cecilia [1 ]
Ulriksdotter, Josefin [1 ,2 ]
Lejding, Tina [1 ]
Bruze, Magnus [1 ]
Mowitz, Martin [1 ]
机构
[1] Lund Univ, Skane Univ Hosp, SUS, Dept Occupat & Environm Dermatol, Malmo, Sweden
[2] Sweden Univ Lund, Helsingborg Hosp, Dept Dermatol, Malmo, Sweden
关键词
2,2 '-methylenebis(6-tert-butyl-4-methylphenol) monoacrylate; allergic contact dermatitis; contact allergy; diabetes type I; gas chromatography-mass spectrometry glucose monitoring (CGM) systems; isobornyl acrylate medical device; DERMATITIS;
D O I
10.1111/cod.13781
中图分类号
R392 [医学免疫学];
学科分类号
100102 ;
摘要
Background: Medical devices (MD) in close skin-contact for a prolonged time, such as glucose monitoring (CGM) systems, are a risk factor for contact allergy, and there has been an increase in patients using these. Correct diagnosis demands targetted testing. Objectives: We report a new allergen in a continuous CGM system in which the adhesive was changed. The allergy pattern of the patients diagnosed is reported. Methods: The three patients reported were patch tested with an MD series, own material, and possible allergens found through analysis with gas chromatography-mass spectrometry, comparing analysis from the CGM system before and after change. Results: The patients were sensitized to isobornyl acrylate (IBOA), found in previously used devices and the present CGM. Apart from IBOA, the culprit allergen was found to be 2,2 '-methylenebis(6-tert-butyl-4-methylphenol) monoacrylate. Conclusion; Allergic contact dermatitis due to CGM systems and insulin pumps are difficult to investigate and require chemical analysis. Because of the lack of information on substances used in the production, and when changes with MDs are initiated, it is difficult to advise patients, especially since they risk sensitization to several allergens. The use of MDs has increased and, thus, the need for collaboration between manufacturers, clinicians, and patient organizations.
引用
收藏
页码:439 / 446
页数:8
相关论文
共 29 条
[11]   Intense skin reaction to a new glucose monitoring and insulin pump system [J].
Hamnerius, Nils ;
Mowitz, Martin .
CONTACT DERMATITIS, 2020, 83 (06) :524-527
[12]   Unexpected positive patch test reactions to sesquiterpene lactones in patients sensitized to the glucose sensor FreeStyle Libre [J].
Herman, Anne ;
Mowitz, Martin ;
Aerts, Olivier ;
Pyl, Jeroen ;
de Montjoye, Laurence ;
Goossens, An ;
Bruze, Magnus ;
Baeck, Marie .
CONTACT DERMATITIS, 2019, 81 (05) :354-367
[13]   Allergic contact dermatitis caused by medical devices for diabetes patients: A review [J].
Herman, Anne ;
de Montjoye, Laurence ;
Tromme, Isabelle ;
Goossens, An ;
Baeck, Marie .
CONTACT DERMATITIS, 2018, 79 (06) :331-335
[14]   Flare-up reactions after oral challenge with nickel in relation to challenge dose and intensity and time of previous patch test reactions [J].
Hindsén, M ;
Bruze, M ;
Christensen, OB .
JOURNAL OF THE AMERICAN ACADEMY OF DERMATOLOGY, 2001, 44 (04) :616-623
[15]   European Society of Contact Dermatitis guideline for diagnostic patch testing - recommendations on best practice [J].
Johansen, Jeanne D. ;
Aalto-Korte, Kristiina ;
Agner, Tove ;
Andersen, Klaus E. ;
Bircher, Andreas ;
Bruze, Magnus ;
Cannavo, Alicia ;
Gimenez-Arnau, Ana ;
Goncalo, Margarida ;
Goossens, An ;
John, Swen M. ;
Liden, Carola ;
Lindberg, Magnus ;
Mahler, Vera ;
Matura, Mihly ;
Rustemeyer, Thomas ;
Serup, Jorgen ;
Spiewak, Radoslaw ;
Thyssen, Jacob P. ;
Vigan, Martine ;
White, Ian R. ;
Wilkinson, Mark ;
Uter, Wolfgang .
CONTACT DERMATITIS, 2015, 73 (04) :195-221
[16]  
Mayzo Inc, PROD DAT SHEET BNX 3
[17]  
Mowitz, 2021, CONTACT DERMATITIS, P1
[18]   Patch Testing with Isobornyl Acrylate in 16 Swedish Patients with Contact Dermatitis from Glucose Sensors and/or Insulin Pumps [J].
Mowitz, Martin ;
Fornander, Louise ;
Hosseiny, Sara ;
Ryberg, Kristina ;
Bruze, Magnus .
ACTA DERMATO-VENEREOLOGICA, 2019, 99 (13) :1286-1287
[19]  
Paul, 2002, ADHESION SCI ENG
[20]   Documentation of Skin-Related Issues Associated with Continuous Glucose Monitoring Use in the Scientific Literature [J].
Pleus, Stefan ;
Ulbrich, Sina ;
Zschornack, Eva ;
Kamann, Stefanie ;
Haug, Cornelia ;
Freckmann, Guido .
DIABETES TECHNOLOGY & THERAPEUTICS, 2019, 21 (10) :538-545