Development of a 3D printed quality control tool for evaluation of x-ray beam alignment and collimation

被引:2
作者
Oliveira, Marcus [1 ]
Barros, Jose Carlos [2 ]
Ubeda, Carlos [3 ]
机构
[1] Fed Inst Bahia, Dept Hlth Technol & Biol, Salvador, BA, Brazil
[2] Fed Inst Bahia, Salvador, BA, Brazil
[3] Tarapaca Univ, Hlth Sci Fac, Med Technol Dept, Arica, Chile
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2019年 / 65卷
关键词
Image quality; X-ray; Quality control; Quality assurance; 3D printer;
D O I
10.1016/j.ejmp.2019.07.026
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this work was to develop a low-cost, 3D printed tool to evaluate X-ray beam alignment and collimation. The study was divided into two phases: 1) the development of 3D printed prototypes; and 2) a comparison with a commercial test object. A 3D printer was used to develop two objects that utilized 40% infill and were each printed with a different filament PLA (polylactic acid) and ABS (acrylonitrile butadiene styrene). Two pieces of X-ray equipment were used for the beam collimation and beam alignment tests. For validation, a standard commercial tool was used, and the evaluation results of the prototypes were compared with those of the commercial tool. The tests performed with both the prototypes and the standard tool showed a deviation of +/- 1 cm between the light field and the radiation field. The central ray's perpendicularity was evaluated through the coincidence between the rod and the metallic circle. The test of central ray alignment conducted with a standard tool revealed an axis perpendicularity of 1.5 degrees, while both prototypes presented axis perpendicularities of less than 3 degrees. The prototypes proved to be effective tools and were easy to handle. The variety of printing materials that can be used and the ease with which the filaments can be acquired contribute to a low cost of production.
引用
收藏
页码:29 / 32
页数:4
相关论文
共 20 条
  • [11] 3D printing X-Ray Quality Control Phantoms. A Low Contrast Paradigm
    Kapetanakis, I.
    Fountos, G.
    Michail, C.
    Valais, I.
    Kalyvas, N.
    [J]. INTERNATIONAL CONFERENCE ON BIO-MEDICAL INSTRUMENTATION AND RELATED ENGINEERING AND PHYSICAL SCIENCES (BIOMEP 2017), 2017, 931
  • [12] Karami V, 2017, J Biomed Phys Eng, V7, P101
  • [13] Medical 3D Printing for the Radiologist
    Mitsouras, Dimitris
    Liacouras, Peter
    Imanzadeh, Amir
    Giannopoulos, Andreas A.
    Cai, Tianrun
    Kumamaru, Kanako K.
    George, Elizabeth
    Wake, Nicole
    Caterson, Edward J.
    Pomahac, Bohdan
    Ho, Vincent B.
    Grant, Gerald T.
    Rybicki, Frank J.
    [J]. RADIOGRAPHICS, 2015, 35 (07) : 1966 - 1989
  • [14] Paquet F., 2017, Annals of the ICRP, V46, P1, DOI [10.1177/0146645317734963, 10.1016/j.icrp.2013.01.001, 10.1016/j.icrp.2008.08.001]
  • [15] Royal Australian and New Zealand College of Radiologists, GEN XRAY QA QC GUID
  • [16] SEFM-SEPR-SERAM, 2011, PROT ESP CONTR CAL R
  • [17] Imaging Properties of 3D Printed Materials: Multi-Energy CT of Filament Polymers
    Shin, James
    Sandhu, Ranjit S.
    Shih, George
    [J]. JOURNAL OF DIGITAL IMAGING, 2017, 30 (05) : 572 - 575
  • [18] A comparative study of collimation in bedside chest radiography for preterm infants in two teaching hospitals
    Stotifuss, J.
    Schneider, K.
    Krueger-Stollfuss, I.
    [J]. EUROPEAN JOURNAL OF RADIOLOGY OPEN, 2015, 2 : 118 - 122
  • [19] The ALARA concept in pediatric CR and DR: dose reduction in pediatric radiographic exams - A white paper conference Executive Summary
    Willis, CE
    Slovis, TL
    [J]. PEDIATRIC RADIOLOGY, 2004, 34 (Suppl 3) : S162 - S164
  • [20] Three dimensional printing technology and materials for treatment of elbow fractures
    Yang, Long
    Grottkau, Brian
    He, Zhixu
    Ye, Chuan
    [J]. INTERNATIONAL ORTHOPAEDICS, 2017, 41 (11) : 2381 - 2387