Bone Temperature Variation Using a 3D-Printed Surgical Guide with Internal Irrigation

被引:9
作者
Stocchero, Michele [1 ,2 ]
Sivolella, Stefano [1 ]
Brunello, Giulia [1 ]
Zoppello, Arianna [1 ]
Cavallin, Francesco [3 ]
Biasetto, Lisa [4 ]
机构
[1] Univ Padua, Dept Neurosci, Sect Dent, Via Giustiniani 2, I-35128 Padua, Italy
[2] Malmo Univ, Fac Odontol, Dept Oral & Maxillofacial Surg & Oral Med, Carl Gustays Vag 34, S-20506 Malmo, Sweden
[3] Independent Statistician, I-36020 Solagna, Italy
[4] Univ Padua, Dept Management & Engn, Stradella San Nicola 3, I-36100 Vicenza, Italy
来源
APPLIED SCIENCES-BASEL | 2021年 / 11卷 / 06期
关键词
dental implants; guided surgery; irrigation system; three-dimensional printing; osteotomy; overheating;
D O I
10.3390/app11062588
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Bone overheating is a possible cause of implants early failure. When a surgical guide is used, the risk of heat injury is greater due to the reduced efficacy of the irrigation. The aim of this ex vivo study was to evaluate the effect of an additional built-in irrigation on bone temperature variation during implant osteotomy. Twelve bovine ribs were used. Cone beam computerized tomography (CBCT) was performed and a 3D-printed surgical guide with additional built-in irrigation tubes was produced for each rib. A total of 48 osteotomies were prepared, to compare the supplementary internal irrigation system (Group A) with external irrigation alone (Group B), no irrigation (Group C) and with free-hand surgery with external irrigation (Group D). Temperature was measured by three thermocouples placed at depths of 1.5, 7, and 12 mm. The largest temperature variation at each thermocouple showed median values of 3.0 degrees C, 1.9 degrees C, and 2.3 degrees C in Group 1; 2.3 degrees C, 1.7 degrees C, and 0.9 degrees C in Group 2; 3.2 degrees C, 1.6 degrees C, and 2.0 degrees C in Group 3; 2.0 degrees C, 2.0 degrees C, and 1.3 degrees C in Group 4, respectively. No differences were found among the four groups. In general, the highest temperature increase was observed with the use of the first drill (cortical perforator). Post-experimental CBCT revealed the presence of radiopaque material clogging the aperture of the internal irrigation channels. Additional internal irrigation was not found to significantly contribute to decrease bone temperature in this ex vivo setting.
引用
收藏
页数:12
相关论文
共 38 条
  • [1] A Thermal and Biological Analysis of Bone Drilling
    Aghvami, Maziar
    Brunski, John B.
    Tulu, U. Serdar
    Chen, Chih-Hao
    Helms, Jill A.
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2018, 140 (10):
  • [2] OSSEOINTEGRATED TITANIUM IMPLANTS - REQUIREMENTS FOR ENSURING A LONG-LASTING, DIRECT BONE-TO-IMPLANT ANCHORAGE IN MAN
    ALBREKTSSON, T
    BRANEMARK, PI
    HANSSON, HA
    LINDSTROM, J
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 1981, 52 (02): : 155 - 170
  • [3] Guided Implant Placement Using an Internally Cooling Surgical Template: A Technical Note
    Alevizakos, Vasilios
    Mitov, Gergo
    von See, Constantin
    [J]. JOURNAL OF ORAL IMPLANTOLOGY, 2020, 46 (05) : 533 - 536
  • [4] Cortical bone drilling and thermal osteonecrosis
    Augustin, Goran
    Zigman, Tomislav
    Davila, Slavko
    Udilljak, Toma
    Staroveski, Tomislav
    Brezak, Danko
    Babic, Slaven
    [J]. CLINICAL BIOMECHANICS, 2012, 27 (04) : 313 - 325
  • [5] Flapless implant surgery: Review of the literature and report of 2 cases with computer-guided surgical approach
    Azari, Abbas
    Nikzad, Sakineh
    [J]. JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2008, 66 (05) : 1015 - 1021
  • [6] Intraosseous Heat Generation During Osteotomy Performed Freehand and Through Template With an Integrated Metal Guide Sleeve: An In Vitro Study
    Barrak, Ibrahim
    Joob-Fancsaly, Arpad
    Braunitzer, Gabor
    Varga, Endre, Jr.
    Boa, Kristof
    Piffko, Jozsef
    [J]. IMPLANT DENTISTRY, 2018, 27 (03) : 342 - 350
  • [7] Bulloch SE, 2012, INT J ORAL MAX IMPL, V27, P1456
  • [8] CLATTENBURG R, 1975, J BIOMED MATER RES, V9, P169
  • [9] Measurement of thermal conductivity of bovine cortical bone
    Davidson, SRH
    James, DF
    [J]. MEDICAL ENGINEERING & PHYSICS, 2000, 22 (10) : 741 - 747
  • [10] How Accurate Are Implant Surgical Guides Produced With Desktop Stereolithographic 3-Dimentional Printers?
    Deeb, George R.
    Allen, Riley K.
    Hall, V. Patrick
    Whitley, Daniel, III
    Laskin, Daniel M.
    Bencharit, Sompop