A new automated screening assay for the diagnosis of von Willebrand disease

被引:29
作者
Salem, Raneem O.
Van Cott, Elizabeth M.
机构
[1] Massachusetts Gen Hosp, Clin Labs, Div Lab Med, Dept Pathol, Boston, MA 02114 USA
[2] Harvard Univ, Sch Med, Boston, MA USA
关键词
Automated; Ristocetin cofactor; Von willebrand factor activity;
D O I
10.1309/CEPND3LFHQ87XU4D
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
A new, automated assay for von Willebrand factor (vWF) activity has recently become commercially available (HemosIL vWF activity assay, Instrumentation Laboratories, Lexington, MA). We prospectively studied 61 specimens from 58 patients undergoing laboratory testing for suspicion of von Willebrand disease with this new method, in comparison with the established ristocetin cofactor method. Assays for factor VIII and vWF antigen were also performed using an established method on an MDA-180 coagulation analyzer (bioMerieux, Durham, NC) and a new method on an ACL TOP coagulation analyzer (Instrumentation Laboratories). Blood types were determined. The results showed no significant difference between the assays for factor VIII (mean, 97% for MDA-180 and ACL TOP; P =.494) or vWF antigen (mean, MDA-180,109%; ACL TOP, 111%; P=766). The mean result for the ristocetin cofactor assay was 106% vs 93% with the automated vWF activity (P =.007). The automated activity assay was 100% sensitive and 86% specific for detecting vWF abnormalities and seems to be a suitable screening test. Abnormal results should be followed up with a ristocetin cofactor activity assay for confirmation. Further study is recommended to confirm these conclusions.
引用
收藏
页码:730 / 735
页数:6
相关论文
共 8 条
[1]  
De Vleeschamer A, 2006, BLOOD COAGUL FIBRIN, V17, P353
[2]   Detection of von Willebrand disorder and identification of qualitative von Willebrand factor defects - Direct comparison of commercial ELISA-based von Willebrand factor activity options [J].
Favaloro, EJ .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2000, 114 (04) :608-618
[3]  
GILL JC, 1987, BLOOD, V69, P1691
[4]  
*NAT COMM CLIN LAB, 2003, COLL TRANSP PROC BLO
[5]  
*NAT COMM CLIN LAB, 2002, ASS VONW FACT ANT RI
[6]   Determination of von Willebrand factor activity with collagen-binding assay and diagnosis of von Willebrand disease: Effect of collagen source and coating conditions [J].
Paczuski, R .
JOURNAL OF LABORATORY AND CLINICAL MEDICINE, 2002, 140 (04) :250-254
[7]  
PINOL M, 2003, J THROMB HAEMOST S, V1, P1679
[8]  
RODEGHIERO F, 1987, BLOOD, V69, P454