Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of > 10 years

被引:465
作者
Kleemann, Thomas [1 ]
Becker, Torsten [1 ]
Doenges, Klaus [1 ]
Vater, Margit [1 ]
Senges, Jochen [1 ]
Schneider, Steffen [1 ]
Saggau, Werner [1 ]
Weisse, Udo [1 ]
Seidl, Karlheinz [1 ]
机构
[1] Herzzentrum Ludwigshafen, Med Klin B, Klinikum Ludwigshafen, D-67063 Ludwigshafen, Germany
关键词
defibrillation; cardioverter-defibrillators; implantable; tachyarrhythmia;
D O I
10.1161/CIRCULATIONAHA.106.663807
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background - The number of patients with longer follow-up after implantation of an implantable cardioverter-defibrillator is increasing continuously. Defibrillation lead failure is a typical long-term complication. Therefore, the long-term reliability of implantable cardioverter-defibrillator leads has become an increasing concern. The aim of the present study was to assess the annual rate of transvenous defibrillation lead defects related to follow-up time after lead implantation. Methods and Results - A total of 990 consecutive patients who underwent first implantation of an implantable cardioverter-defibrillator between 1992 and May 2005 were analyzed. Median follow-up time was 934 days (interquartile range, 368 to 1870). Overall, 148 defibrillation leads (15%) failed during the follow-up. The estimated lead survival rates at 5 and 8 years after implantation were 85% and 60%, respectively. The annual failure rate increased progressively with time after implantation and reached 20% in 10-year-old leads (P < 0.001). Lead defects affected newer as well as older models. Patients with lead defects were 3 years younger at implantation and more often female. Multiple lead implantation was associated with a trend to a higher rate of defibrillation lead defects (P=0.06). The major plead complications were insulation defects (56%), lead fractures (12%), loss of ventricular capture (11%), abnormal lead impedance (10%), and sensing failure (10%). Conclusions - An increasing annual lead failure rate is noted primarily during long-term follow-up and reached 20% in 10-year-old leads. Patients with lead defects are younger and more often female.
引用
收藏
页码:2474 / 2480
页数:7
相关论文
共 27 条
  • [1] Sudden cardiac death unresponsive to implantable defibrillator therapy: An urgent target for clinicians, industry and government
    Anderson, KP
    [J]. JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2005, 14 (02) : 71 - 78
  • [2] Insulation lead failure: Is it a matter of insulation coating, venous approach, or both?
    Antonelli, D
    Rosenfeld, T
    Freedberg, NA
    Palma, E
    Gross, JN
    Furman, S
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (02): : 418 - 421
  • [3] Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure
    Bardy, GH
    Lee, KL
    Mark, DB
    Poole, JE
    Packer, DL
    Boineau, R
    Domanski, M
    Troutman, C
    Anderson, J
    Johnson, G
    McNulty, SE
    Clapp-Channing, N
    Davidson-Ray, LD
    Fraulo, ES
    Fishbein, DP
    Luceri, RM
    Ip, JH
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2005, 352 (03) : 225 - 237
  • [4] Defibrillation energy requirements using a left anterior chest cutaneous to subcutaneous shocking vector: Implications for a total subcutaneous implantable defibrillator
    Burke, MC
    Coman, JA
    Cates, AW
    Lindstrom, CC
    Sandler, DA
    Kim, SS
    Knight, BP
    [J]. HEART RHYTHM, 2005, 2 (12) : 1332 - 1338
  • [5] Electrophysiologic testing to identify patients with coronary artery disease who are at risk for sudden death
    Buxton, AE
    Lee, KL
    DiCarlo, L
    Gold, MR
    Greer, GS
    Prystowsky, EN
    O'Toole, MF
    Tang, A
    Fisher, JD
    Coromilas, J
    Talajic, M
    Hafley, G
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2000, 342 (26) : 1937 - 1945
  • [6] Incidence and implications of abrasion of implantable cardioverter-defibrillator leads
    DeLurgio, DB
    Sathavorn, C
    Mera, F
    Leon, A
    Walter, PF
    Langberg, JJ
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1997, 79 (10) : 1409 - &
  • [7] Transvenous defibrillation leads: High incidence of failure during long-term follow-up
    Dorwarth, U
    Frey, B
    Dugas, M
    Matis, T
    Fiek, M
    Schmoeckel, M
    Remp, T
    Durchlaub, I
    Gerth, A
    Steinbeck, G
    Hoffmann, E
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2003, 14 (01) : 38 - 43
  • [8] Detection and management of an implantable cardioverter defibrillator lead failure - Incidence and clinical implications
    Ellenbogen, KA
    Wood, MA
    Shepard, RK
    Clemo, HF
    Vaughn, T
    Holloman, K
    Dow, M
    Leffler, J
    Abeyratne, A
    Verness, D
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2003, 41 (01) : 73 - 80
  • [9] Long-term structural failure of coaxial polyurethane implantable cardioverter defibrillator leads
    Hauser, RG
    Cannom, D
    Hayes, DL
    Parsonnet, V
    Hayes, J
    Ratliff, N
    Tyers, GFO
    Epstein, AE
    Vlay, SC
    Furman, S
    Gross, J
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (06): : 879 - 882
  • [10] MULTICENTER EXPERIENCE WITH A BIPOLAR TINED POLYURETHANE VENTRICULAR LEAD
    HAYES, DL
    GRAHAM, KJ
    IRWIN, M
    VIDAILLET, H
    DISLER, G
    SWEESY, M
    KINCAID, D
    OSBORN, MJ
    SUMAN, VJ
    NEUBAUER, SA
    SEEBANDT, M
    KALLINEN, L
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (05): : 999 - 1004