Cardiac output measurement using an ultrasound dilution method: A validation study in ventilated piglets

被引:28
作者
de Boode, Willem P. [1 ]
van Heijst, Arno F. J. [1 ]
Hopman, Jeroen C. W. [3 ]
Tanke, Ronald B. [2 ]
van der Hoeven, Hans G. [4 ]
Liem, Kian D. [1 ]
机构
[1] Radboud Univ Nijmegen, Med Ctr, Dept Neonatol, NL-6525 ED Nijmegen, Netherlands
[2] Radboud Univ Nijmegen, Med Ctr, Dept Pediat Cardiol, NL-6525 ED Nijmegen, Netherlands
[3] Radboud Univ Nijmegen, Med Ctr, Dept Pediat, NL-6525 ED Nijmegen, Netherlands
[4] Radboud Univ Nijmegen, Med Ctr, Dept Intens Care Med, NL-6525 ED Nijmegen, Netherlands
关键词
cardiac output; infant; hemodynamics; ultrasound dilution; systemic blood flow; validation; BLOOD-VOLUME; ACCESS FLOW; HEMODIALYSIS; LOOP;
D O I
10.1097/PCC.0b013e3181b064ea
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To assess agreement between a new method of cardiac. output monitoring, using ultrasound dilution technology and ultrasound transit time-based measurement of pulmonary blood flow in a piglet model. Design: Prospective, experimental juvenile animal study. Setting: Animal laboratory of a university hospital. Subjects: Nine random-bred piglets. Interventions: After the animals received general anesthesia, we placed intravascular arterial and central venous catheters with the tip positioned in the abdominal aorta and the right atrium, respectively. The catheters were connected to the ultrasound dilution cardiac output monitor. An ultrasound transit time perivascular flow probe was positioned around the common pulmonary artery and served as the standard reference measurement. Cardiac output was manipulated during the experiment by creating hemorrhagic hypotension. Ultrasound dilution cardiac output was measured intermittently with injection volumes of 0.5 mL/kg and 1.0 mL/kg of isotonic saline at body temperature. Measurements and Main Results: Ultrasound dilution cardiac Output (Q(UDCO)) measurement was compared with pulmonary blood flow (Q(APC)). Bias, defined as Q(UDCO) minus Q(APC) was calculated for each measurement. Mean bias with standard deviation was calculated for measurements with volumes of injected saline, 0.5 mL/kg and 1.0 mL/kg, and compared using the Mann-Whitney U test. Mean bias (SD) between Q(UDCO) and Q(APC) was 0.040 (0.132) and 0.058 (0.136) L/min for measurement with 0.5 mL/kg and 1.0 mL/kg of isotonic saline, respectively (no statistically significant difference). Conclusions: Ultrasound dilution cardiac output measurement is reliable in piglets with the use of a small volume of a nontoxic indicator (isotonic saline). (Pediatr Crit Care Med 2010; 11:103-108)
引用
收藏
页码:103 / 108
页数:6
相关论文
共 34 条
[1]   Hemodynamic monitoring in shock and implications for management International Consensus Conference, Paris, France, 27-28 April 2006 [J].
Antonelli, Massimo ;
Levy, Mitchell ;
Andrews, Peter J. D. ;
Chastre, Jean ;
Hudson, Leonard D. ;
Manthous, Constantine ;
Meduri, G. Umberto ;
Moreno, Rui P. ;
Putensen, Christian ;
Stewart, Thomas ;
Torres, Antoni .
INTENSIVE CARE MEDICINE, 2007, 33 (04) :575-590
[2]   Restricted versus liberal water intake for preventing morbidity and mortality in preterm infants [J].
Bell, E. F. ;
Acarregui, M. J. .
COCHRANE DATABASE OF SYSTEMATIC REVIEWS, 2008, (01)
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   Minimally invasive hemodynamic monitoring for the intensivist: Current and emerging technology [J].
Chaney, JC ;
Derdak, S .
CRITICAL CARE MEDICINE, 2002, 30 (10) :2338-2345
[5]   A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques [J].
Critchley, LAH ;
Critchley, JAJH .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (02) :85-91
[6]   Cardiac output measurement using a modified carbon dioxide fick method:: A validation study in ventilated lambs [J].
De Boode, Willem P. ;
Hopman, Jercen C. W. ;
Daniels, Otto ;
Van der Hoeven, Hans G. ;
Liem, K. Wien .
PEDIATRIC RESEARCH, 2007, 61 (03) :279-283
[7]   Validation study of a new transit time ultrasonic flow probe for continuous great vessel measurements [J].
Dean, DA ;
Jia, CX ;
Cabreriza, SE ;
DAlessandro, DA ;
Dickstein, ML ;
Sardo, MJ ;
Chalik, N ;
Spotnitz, HM .
ASAIO JOURNAL, 1996, 42 (05) :M671-M676
[8]   Heart blood volume by dilution in patients on hemodialysis [J].
Dobson, A ;
Kislukhin, VV .
ASAIO JOURNAL, 2004, 50 (03) :278-284
[9]  
Dwyer-Matzky K, 2007, CRIT CARE MED, V35, pA75
[10]   Use of an extracorporeal arteriovenous tubing loop to measure cardiac output in intensive care unit patients by ultrasound velocity dilution [J].
Eremenko, A ;
Balykov, I ;
Chaus, N ;
Kislukhin, V ;
Krivitski, N .
ASAIO JOURNAL, 1998, 44 (05) :M462-M464