Highly precise acoustic calibration method of ring-shaped ultrasound transducer array for plane-wave-based ultrasound tomography

被引:14
作者
Terada, Takahide [1 ]
Yamanaka, Kazuhiro [1 ]
Suzuki, Atsuro [1 ]
Tsubota, Yushi [1 ]
Wu, Wenjing [1 ]
Kawabata, Ken-ichi [1 ]
机构
[1] Hitachi Ltd, Res & Dev Grp, Kokubunji, Tokyo 1858601, Japan
关键词
Calibration - Wave propagation - Elastic waves - Computerized tomography - Signal to noise ratio - Signal receivers - Ultrasonic imaging - Ultrasonic transducers;
D O I
10.7567/JJAP.56.07JF07
中图分类号
O59 [应用物理学];
学科分类号
摘要
Ultrasound computed tomography (USCT) is promising for a non-invasive, painless, operator-independent and quantitative system for breast-cancer screening. Assembly error, production tolerance, and aging-degradation variations of the hardwire components, particularly of plane-wave-based USCT systems, may hamper cost effectiveness, precise imaging, and robust operation. The plane wave is transmitted from a ring-shaped transducer array for receiving the signal at a high signal-to-noise-ratio and fast aperture synthesis. There are four signal-delay components: response delays in the transmitters and receivers and propagation delays depending on the positions of the transducer elements and their directivity. We developed a highly precise calibration method for calibrating these delay components and evaluated it with our prototype plane-wave-based USCT system. Our calibration method was found to be effective in reducing delay errors. Gaps and curves were eliminated from the plane wave, and echo images of wires were sharpened in the entire imaging area. (C) 2017 The Japan Society of Applied Physics
引用
收藏
页数:8
相关论文
共 29 条
[1]  
[Anonymous], HV2901 DAT
[2]  
[Anonymous], VANTAGE 256 PROD BRO
[3]  
[Anonymous], 2013, GLOBOCAN 2012 ESTIMA
[4]  
Dick D. E., 1977, P IEEE ULTRASONICS S, P176
[5]  
Duric N., 2013, P IEEE INT ULTR S, P382
[6]   THE ULTRASOUND MACROSCOPE - INITIAL STUDIES OF BREAST-TISSUE [J].
FOSTER, FS ;
STRBAN, M ;
AUSTIN, G .
ULTRASONIC IMAGING, 1984, 6 (03) :243-261
[7]   COMPREHENSIVE COMPILATION OF EMPIRICAL ULTRASONIC PROPERTIES OF MAMMALIAN-TISSUES [J].
GOSS, SA ;
JOHNSTON, RL ;
DUNN, F .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1978, 64 (02) :423-457
[8]  
GREENLEAF J F, 1978, Ultrasound in Medicine and Biology, V3, P327, DOI 10.1016/0301-5629(78)90075-3
[9]  
Haka AS, 2002, CANCER RES, V62, P5375
[10]  
Hopp T., 2013, P IEEE INT ULTR S, P647