Analyte noise in the flame photometric detector

被引:1
|
作者
Singh, H [1 ]
Aue, WA [1 ]
机构
[1] DALHOUSIE UNIV,DEPT CHEM,HALIFAX,NS B3H 4J3,CANADA
基金
加拿大自然科学与工程研究理事会;
关键词
detectors; GC; flame photometric detector; noise; analyte noise;
D O I
10.1016/0021-9673(95)00916-7
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
As the concentration of analyte increases in the flame photometric detector (FPD), its observed noise changes from a predominantly fundamental to a predominantly multiplicative variety. This change-over occurs typically in the 10(-9) to 10(-8) g/s region, with the precise position apparently dependent on the nature of the analyte. In the multiplicative region, slow fluctuations can be seen appearing simultaneously on both channels of a dual-channel FPD. This suggests the possibility of suppressing such correlated noise episodes.