Commissioning of a clinical pencil beam scanning proton therapy unit for ultra-high dose rates (FLASH)

被引:53
作者
Nesteruk, Konrad P. [1 ]
Togno, Michele [1 ]
Grossmann, Martin [1 ]
Lomax, Anthony J. [1 ,2 ]
Weber, Damien C. [1 ,3 ,4 ]
Schippers, Jacobus M. [5 ]
Safai, Sairos [1 ]
Meer, David [1 ]
Psoroulas, Serena [1 ]
机构
[1] Paul Scherrer Inst, Ctr Proton Therapy, Villigen, Switzerland
[2] Swiss Fed Inst Technol, Dept Phys, Zurich, Switzerland
[3] Univ Hosp Zurich, Dept Radiat Oncol, Zurich, Switzerland
[4] Univ Hosp Bern, Dept Radiat Oncol, Bern, Switzerland
[5] Paul Scherrer Inst, Div Large Res Facil, Villigen, Switzerland
基金
瑞士国家科学基金会;
关键词
FLASH; gantry; pencil beam scanning; proton therapy; ultra‐ high dose rates; LINEAR-ACCELERATOR; IRRADIATION; VALIDATION; DOSIMETRY; FACILITY; DESIGN;
D O I
10.1002/mp.14933
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The purpose of this work was to provide a flexible platform for FLASH research with protons by adapting a former clinical pencil beam scanning gantry to irradiations with ultra-high dose rates. Methods PSI Gantry 1 treated patients until December 2018. We optimized the beamline parameters to transport the 250 MeV beam extracted from the PSI COMET accelerator to the treatment room, maximizing the transmission of beam intensity to the sample. We characterized a dose monitor on the gantry to ensure good control of the dose, delivered in spot-scanning mode. We characterized the beam for different dose rates and field sizes for transmission irradiations. We explored scanning possibilities in order to enable conformal irradiations or transmission irradiations of large targets (with transverse scanning). Results We achieved a transmission of 86% from the cyclotron to the treatment room. We reached a peak dose rate of 9000 Gy/s at 3 mm water equivalent depth, along the central axis of a single pencil beam. Field sizes of up to 5 x 5 mm(2) were achieved for single-spot FLASH irradiations. Fast transverse scanning allowed to cover a field of 16 x 1.2 cm(2). With the use of a nozzle-mounted range shifter, we are able to span depths in water ranging from 19.6 to 37.9 cm. Various dose levels were delivered with precision within less than 1%. Conclusions We have realized a proton FLASH irradiation setup able to investigate continuously a wide dose rate spectrum, from 1 to 9000 Gy/s in single-spot irradiation as well as in the pencil beam scanning mode. As such, we have developed a versatile test bench for FLASH research.
引用
收藏
页码:4017 / 4026
页数:10
相关论文
共 26 条
[1]   Feasibility of proton FLASH effect tested by zebrafish embryo irradiation [J].
Beyreuther, Elke ;
Brand, Michael ;
Hans, Stefan ;
Hideghety, Katalin ;
Karsch, Leonhard ;
Lessmann, Elisabeth ;
Schuerer, Michael ;
Szabo, Emilia Rita ;
Pawelke, Jorg .
RADIOTHERAPY AND ONCOLOGY, 2019, 139 :46-50
[2]   Treatment of a first patient with FLASH-radiotherapy [J].
Bourhis, Jean ;
Sozzi, Wendy Jeanneret ;
Jorge, Patrik Goncalves ;
Gaide, Olivier ;
Bailat, Claude ;
Duclos, Frederic ;
Patin, David ;
Ozsahin, Mahmut ;
Bochud, Francois ;
Germond, Jean-Francois ;
Moeckli, Raphael ;
Vozenin, Marie-Catherine .
RADIOTHERAPY AND ONCOLOGY, 2019, 139 :18-22
[3]   Clinical translation of FLASH radiotherapy: Why and how? [J].
Bourhis, Jean ;
Montay-Gruel, Pierre ;
Jorge, Patrik Goncalves ;
Bailat, Claude ;
Petit, Benoit ;
Ollivier, Jonathan ;
Jeanneret-Sozzi, Wendy ;
Ozsahin, Mahmut ;
Bochud, Francois ;
Moeckli, Raphael ;
Germond, Jean-Francois ;
Vozenin, Marie-Catherine .
RADIOTHERAPY AND ONCOLOGY, 2019, 139 :11-17
[4]   Medical physics aspects of the synchrotron radiation therapies: Microbeam radiation therapy (MRT) and synchrotron stereotactic radiotherapy (SSRT) [J].
Braeuer-Krisch, Elke ;
Adam, Jean-Francois ;
Alagoz, Enver ;
Bartzsch, Stefan ;
Crosbie, Jeff ;
DeWagter, Carlos ;
Dipuglia, Andrew ;
Donzelli, Mattia ;
Doran, Simon ;
Fournier, Pauline ;
Kalef-Ezra, John ;
Kock, Angela ;
Lerch, Michael ;
McErlean, Ciara ;
Oelfke, Uwe ;
Olko, Pawel ;
Petasecca, Marco ;
Povoli, Marco ;
Rosenfeld, Anatoly ;
Siegbahn, Erik A. ;
Sporea, Dan ;
Stugu, Bjarne .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (06) :568-583
[5]  
Brown K. L., 1980, CERN YELLOW REPORTS
[6]   Biological effects in normal cells exposed to FLASH dose rate protons [J].
Buonanno, Manuela ;
Grilj, Veljko ;
Brenner, David J. .
RADIOTHERAPY AND ONCOLOGY, 2019, 139 :51-55
[7]  
Busold S, P CYCLOTRONS 2016
[8]   Al2O3:C optically stimulated luminescence dosimeters (OSLDs) for ultra-high dose rate proton dosimetry [J].
Christensen, Jeppe Brage ;
Togno, Michele ;
Nesteruk, Konrad Pawel ;
Psoroulas, Serena ;
Meer, David ;
Weber, Damien Charles ;
Lomax, Tony ;
Yukihara, Eduardo G. ;
Safai, Sairos .
PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (08)
[9]   Feasibility of proton FLASH irradiation using a synchrocyclotron for preclinical studies [J].
Darafsheh, Arash ;
Hao, Yao ;
Zwart, Townsend ;
Wagner, Miles ;
Catanzano, Daniel ;
Williamson, Jeffrey F. ;
Knutson, Nels ;
Sun, Baozhou ;
Mutic, Sasa ;
Zhao, Tianyu .
MEDICAL PHYSICS, 2020, 47 (09) :4348-4355
[10]   FLASH radiotherapy: ultra-high dose rates to spare healthy tissue [J].
de Kruijff, R. M. .
INTERNATIONAL JOURNAL OF RADIATION BIOLOGY, 2020, 96 (04) :419-423