Heads-Up 3D Surgery under Low Light Intensity Conditions: New High-Sensitivity HD Camera for Ophthalmological Microscopes

被引:23
作者
Matsumoto, Celso Soiti [1 ,2 ,3 ]
Shibuya, Masayuki [2 ]
Makita, Jun [2 ]
Shoji, Takuhei [2 ]
Ohno, Hisato [4 ]
Shinoda, Kei [2 ]
Matsumoto, Harue [3 ]
机构
[1] Teikyo Univ, Dept Ophthalmol, Sch Med, Itabashi Ku, Kaga 2-11-1, Tokyo 1738605, Japan
[2] Saitama Med Univ, Dept Ophthalmol, Sch Med, Morohongo 38,Moroyama Machi, Saitama 3500495, Japan
[3] Matsumoto Eye Clin, Tokushima, Japan
[4] Ohno Eye Clin, Saitama, Japan
关键词
ENDOILLUMINATION; VITRECTOMY; SYSTEM;
D O I
10.1155/2019/5013463
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose. To determine the feasibility of performing intraocular surgeries in a heads-up position with low illuminance conditions by observing a display of the surgical field created by a three-dimensional imaging (3D) system. Methods. Seventy-four eyes of 56 patients underwent cataract surgery (72 eyes) with the heads-up 3D surgery system; 60 eyes with cataract surgery alone, 7 eyes with combined cataract and glaucoma microdevice implant surgery, 5 eyes with combined cataract and vitrectomy surgery, and two eyes with vitrectomy surgery alone were studied. The illuminance from the surgical microscope was set to be dimmer (Leica M822F40 main light 2%; otto-flex 6%) than the usual setting to minimize the discomfort and glare for the patient. The surgeries were performed under topical anesthesia. The luminance of the images observed through the eyepieces of the operating microscope and the image of a 3D system created by a high-sensitivity sensor Exmor R 3CMOS HD camera (Sony MCC-1000MD) were measured. Results. All surgeries were completed without any complications under the low illumination conditions. The surgical field on the display monitor was created by a 3D system using a high-sensitivity sensor camera and was observed in a heads-up position. The patients did not report any intolerable discomfort or glare during the surgery. Cataract surgeries were performed with a good view of the surgical field under the extremely low illumination from the surgical microscope. The high-sensitivity sensors and electronic amplifications of the image signals made the surgical field brighter and allowed the surgeon to perform the surgery confidently and safely. Conclusions. Heads-up, 3D-assisted intraocular surgeries can be performed safely and efficiently with low illuminance of the surgical field. This trial is registered with UMIN000037838.
引用
收藏
页数:6
相关论文
共 8 条
[1]  
Adam MK, 2017, RETINA-J RET VIT DIS, V37, P1746, DOI 10.1097/IAE.0000000000001420
[2]   HEADS-UP SURGERY FOR VITREORETINAL PROCEDURES An Experimental and Clinical Study [J].
Eckardt, Claus ;
Paulo, Erica B. .
RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2016, 36 (01) :137-147
[3]   Comparison of clinical outcomes between "heads-up" 3D viewing system and conventional microscope in macular hole surgeries: A pilot study [J].
Kumar, Atul ;
Hasan, Nasiq ;
Kakkar, Prateek ;
Mutha, Vineet ;
Karthikeya, R. ;
Sundar, Dheepak ;
Ravani, Raghav .
INDIAN JOURNAL OF OPHTHALMOLOGY, 2018, 66 (12) :1816-1819
[4]   Heads-Up Macular Surgery with a 27-Gauge Microincision Vitrectomy System and Minimal Illumination [J].
Kunikata, Hiroshi ;
Abe, Toshiaki ;
Nakazawa, Toru .
CASE REPORTS IN OPHTHALMOLOGY, 2016, 7 (03) :265-269
[5]   Endoillumination during vitrectomy and phototoxicity thresholds [J].
van den Biesen, PR ;
Berenschot, T ;
Verdaasdonk, RM ;
van Weelden, H ;
van Norren, D .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2000, 84 (12) :1372-1375
[6]  
Weinstock R., 2010, Surgical Techniques in Ophthalmology Cataract Surgery, P124
[7]  
WEINSTOCK RJ, 2011, CATARACT REFRACT SUR, V8, P66
[8]   8K ultra-high-definition microscopic camera for ophthalmic surgery [J].
Yamashita, Hiromasa ;
Tanioka, Kenkichi ;
Miyake, Goichiro ;
Ota, Ichiro ;
Noda, Toru ;
Miyake, Kensaku ;
Chiba, Toshio .
CLINICAL OPHTHALMOLOGY, 2018, 12 :1823-1828