Diamond-like carbon (DLC) films are gaining big interest in electrochemistry research area. DLC electrodes made with different ratio of sp(3)/sp(2) carbon hybridization or doped with different percentages of nickel were characterized electrochemically by cyclic voltammetry and by amperometric measurements towards hydrogen peroxide. SiCAr1 and SiCNi5% were chosen as sensitive transducers for the elaboration of amperometric glucose biosensors. Immobilization of glucose oxidase was carried out by cross-linking with glutaraldehyde. Measurements were made at a fixed potential + 1.0 V in 40 mM phosphate buffer pH 7.4. SiCAr1 seems to be more sensitive for glucose, 0.6875 mu A/mM, than SiCNi5%, 0.3654 mu A/mM. Detections limits were 20 RM and 30 mu M, respectively. Apparent Michaelis-Menten constants were found around 3 mM. Forty-eight percent and 79% of the original response for 0.5 mM glucose remained after 10 days for both biosensors, respectively. (C) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:310 / 314
页数:5
相关论文
共 18 条
[1]   Nickel-incorporated amorphous carbon film deposited by femtosecond pulsed laser ablation [J].
Benchikh, N ;
Garrelie, F ;
Donnet, C ;
Bouchet-Fabre, B ;
Wolski, K ;
Rogemond, F ;
Loir, AS ;
Subtil, JL .
THIN SOLID FILMS, 2005, 482 (1-2) :287-292
[2]   Biosensors for life quality -: Design, development and applications [J].
Castillo, J ;
Gáspár, S ;
Leth, S ;
Niculescu, M ;
Mortari, A ;
Bontidean, I ;
Soukharev, V ;
Dorneanu, SA ;
Ryabov, AD ;
Csöregi, E .
SENSORS AND ACTUATORS B-CHEMICAL, 2004, 102 (02) :179-194
[3]   Amperometric glucose biosensor based on sol-gel-derived metal oxide/Nafion composite films [J].
Choi, HN ;
Kim, MA ;
Lee, WY .
ANALYTICA CHIMICA ACTA, 2005, 537 (1-2) :179-187
[4]   Development and evaluation of electrochemical glucose enzyme biosensors based on carbon film electrodes [J].
Florescu, M ;
Brett, CMA .
TALANTA, 2005, 65 (02) :306-312
[5]   Improved operational stability of biosensors based on enzyme-polyelectrolyte complex adsorbed into a porous carbon electrode [J].
Gavalas, VG ;
Chaniotakis, NA ;
Gibson, TD .
BIOSENSORS & BIOELECTRONICS, 1998, 13 (11) :1205-1211
[6]   Development and comparison of biosensors for in-vivo applications [J].
Georganopoulou, DG ;
Carley, R ;
Jones, DA ;
Boutelle, MG .
FARADAY DISCUSSIONS, 2000, 116 :291-303
[7]   A glucose biosensor using methyl viologen redox mediator on carbon film electrodes [J].
Ghica, ME ;
Brett, CMA .
ANALYTICA CHIMICA ACTA, 2005, 532 (02) :145-151
[8]   DIAMOND-LIKE CARBON-COATED FILMS FOR ENZYME ELECTRODES - CHARACTERIZATION OF BIOCOMPATIBILITY AND SUBSTRATE DIFFUSION LIMITING PROPERTIES [J].
HIGSON, SPJ ;
VADGAMA, PM .
ANALYTICA CHIMICA ACTA, 1995, 300 (1-3) :77-83
[9]   DIAMOND-LIKE CARBON-FILMS FOR ENZYME ELECTRODES - CHARACTERIZATION OF NOVEL OVERLYING PERMSELECTIVE BARRIERS [J].
HIGSON, SPJ ;
VADGAMA, PM .
ANALYTICA CHIMICA ACTA, 1995, 300 (1-3) :85-90
[10]   Amperometric biosensor sensitive to glucose and lactose based on co-immobilization of ferrocene, glucose oxidase, β-galactosidase and mutarotase in β-cyclodextrin polymer [J].
Liu, HY ;
Li, HH ;
Ying, TL ;
Sun, K ;
Qin, YQ ;
Qi, DY .
ANALYTICA CHIMICA ACTA, 1998, 358 (02) :137-144