The dose response of PTW microDiamond and microSilicon in transverse magnetic field under small field conditions

被引:11
作者
Blum, Isabel [1 ]
Tekin, Tuba [1 ]
Delfs, Bjorn [1 ]
Schonfeld, Ann-Britt [1 ]
Kapsch, Ralf-Peter [2 ]
Poppe, Bjorn [1 ]
Looe, Hui Khee [1 ]
机构
[1] Carl von Ossietzky Univ Oldenburg, Univ Clin Med Radiat Phys, Pius Hosp, Med Campus, Oldenburg, Germany
[2] Phys Tech Bundesanstalt, Braunschweig, Germany
关键词
magnetic resonance guided radiation therapy; Lorentz force; magnetic field; small field dosimetry; dose response; small field output correction factors; MONTE-CARLO; ION CHAMBERS; DETECTORS; DOSIMETRY; DIAMOND; DIODE;
D O I
10.1088/1361-6560/ac0f2e
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The aim of the present work is to investigate the behavior of two diode-type detectors (PTW microDiamond 60019 and PTW microSilicon 60023) in transverse magnetic field under small field conditions. A formalism based on TRS 483 has been proposed serving as the framework for the application of these high-resolution detectors under these conditions. Measurements were performed at the National Metrology Institute of Germany (PTB, Braunschweig) using a research clinical linear accelerator facility. Quadratic fields corresponding to equivalent square field sizes S between 0.63 and 4.27 cm at the depth of measurement were used. The magnetic field strength was varied up to 1.4 T. Experimental results have been complemented with Monte Carlo simulations up to 1.5 T. Detailed simulations were performed to quantify the small field perturbation effects and the influence of detector components on the dose response. The does response of both detectors decreases by up to 10% at 1.5 T in the largest field size investigated. In S = 0.63 cm, this reduction at 1.5 T is only about half of that observed in field sizes S > 2 cm for both detectors. The results of the Monte Carlo simulations show agreement better than 1% for all investigated conditions. Due to normalization at the machine specific reference field, the resulting small field output correction factors for both detectors in magnetic field k(Qclin,Qmsr)(B) are smaller than those in the magnetic field-free case, where correction up to 6.2% at 1.5 T is required for the microSilicon in the smallest field size investigated. The volume-averaging effect of both detectors was shown to be nearly independent of the magnetic field. The influence of the enhanced-density components within the detectors has been identified as the major contributors to their behaviors in magnetic field. Nevertheless, the effect becomes weaker with decreasing field size that may be partially attributed to the deficiency of low energy secondary electrons originated from distant locations in small fields.
引用
收藏
页数:15
相关论文
共 50 条
  • [41] Asymptotic solutions for the asymmetric flow in a channel with porous retractable walls under a transverse magnetic field
    Guo, Hongxia
    Lin, Ping
    Li, Lin
    APPLIED MATHEMATICS AND MECHANICS-ENGLISH EDITION, 2018, 39 (08) : 1147 - 1164
  • [42] Asymptotic solutions for the asymmetric flow in a channel with porous retractable walls under a transverse magnetic field
    Hongxia Guo
    Ping Lin
    Lin Li
    Applied Mathematics and Mechanics, 2018, 39 : 1147 - 1164
  • [43] Chemical diffusion characteristics of Al-Si alloy melts under a transverse magnetic field
    Zhang, Kai
    Bian, Xiufang
    Li, Yumin
    Liu, Yang
    Yang, Chuncheng
    Zhao, Xiaolin
    PHYSICS LETTERS A, 2015, 379 (22-23) : 1464 - 1466
  • [44] Effects of Thermoelectric Magnetic Convection on the Solidification Structure During Directional Solidification under Lower Transverse Magnetic Field
    Xi LI
    Zhongming Ren
    Annie Gagnoud
    Olga Budebkova
    Yves Fautrelle
    Metallurgical and Materials Transactions A, 2011, 42 : 3459 - 3471
  • [45] Linear Optical Response of Silicon Nanotubes Under Axial Magnetic Field
    Raad Chegel
    Somayeh Behzad
    Journal of Electronic Materials, 2013, 42 : 58 - 70
  • [46] Measuring magnetic field texture in correlated electron systems under extreme conditions
    Yip, King Yau
    Ho, Kin On
    Yu, King Yiu
    Chen, Yang
    Zhang, Wei
    Kasahara, S.
    Mizukami, Y.
    Shibauchi, T.
    Matsuda, Y.
    Goh, Swee K.
    Yang, Sen
    SCIENCE, 2019, 366 (6471) : 1355 - +
  • [47] The use of small coolers in a magnetic field
    Green, M. A.
    Witte, H.
    ADVANCES IN CRYOGENIC ENGINEERING, VOLS 53A AND 53B, 2008, 985 : 1299 - +
  • [48] Deformation of Si under conditions of the combined influence of an electric current and a magnetic field
    Velikhanov A.R.
    Bulletin of the Russian Academy of Sciences: Physics, 2014, 78 (2) : 108 - 110
  • [49] Measurement of interaction force between small distances sandwiched with magnetic fluid under magnetic field
    Miyazaki, T
    Shibayama, A
    Sato, T
    Fujita, T
    JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 2002, 252 (1-3) : 256 - 258
  • [50] Impact of transverse magnetic fields on dose response of a nanoDot OSLD in megavoltage photon beams
    Ito, Shotaro
    Araki, Fujio
    Hoshida, Kento
    Ohno, Takeshi
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 70 : 153 - 160