The dose response of PTW microDiamond and microSilicon in transverse magnetic field under small field conditions

被引:11
作者
Blum, Isabel [1 ]
Tekin, Tuba [1 ]
Delfs, Bjorn [1 ]
Schonfeld, Ann-Britt [1 ]
Kapsch, Ralf-Peter [2 ]
Poppe, Bjorn [1 ]
Looe, Hui Khee [1 ]
机构
[1] Carl von Ossietzky Univ Oldenburg, Univ Clin Med Radiat Phys, Pius Hosp, Med Campus, Oldenburg, Germany
[2] Phys Tech Bundesanstalt, Braunschweig, Germany
关键词
magnetic resonance guided radiation therapy; Lorentz force; magnetic field; small field dosimetry; dose response; small field output correction factors; MONTE-CARLO; ION CHAMBERS; DETECTORS; DOSIMETRY; DIAMOND; DIODE;
D O I
10.1088/1361-6560/ac0f2e
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The aim of the present work is to investigate the behavior of two diode-type detectors (PTW microDiamond 60019 and PTW microSilicon 60023) in transverse magnetic field under small field conditions. A formalism based on TRS 483 has been proposed serving as the framework for the application of these high-resolution detectors under these conditions. Measurements were performed at the National Metrology Institute of Germany (PTB, Braunschweig) using a research clinical linear accelerator facility. Quadratic fields corresponding to equivalent square field sizes S between 0.63 and 4.27 cm at the depth of measurement were used. The magnetic field strength was varied up to 1.4 T. Experimental results have been complemented with Monte Carlo simulations up to 1.5 T. Detailed simulations were performed to quantify the small field perturbation effects and the influence of detector components on the dose response. The does response of both detectors decreases by up to 10% at 1.5 T in the largest field size investigated. In S = 0.63 cm, this reduction at 1.5 T is only about half of that observed in field sizes S > 2 cm for both detectors. The results of the Monte Carlo simulations show agreement better than 1% for all investigated conditions. Due to normalization at the machine specific reference field, the resulting small field output correction factors for both detectors in magnetic field k(Qclin,Qmsr)(B) are smaller than those in the magnetic field-free case, where correction up to 6.2% at 1.5 T is required for the microSilicon in the smallest field size investigated. The volume-averaging effect of both detectors was shown to be nearly independent of the magnetic field. The influence of the enhanced-density components within the detectors has been identified as the major contributors to their behaviors in magnetic field. Nevertheless, the effect becomes weaker with decreasing field size that may be partially attributed to the deficiency of low energy secondary electrons originated from distant locations in small fields.
引用
收藏
页数:15
相关论文
共 50 条
  • [31] Monte Carlo simulations of patient dose perturbations in rotational-type radiotherapy due to a transverse magnetic field: A tomotherapy investigation
    Yang, Y. M.
    Geurts, M.
    Smilowitz, J. B.
    Sterpin, E.
    Bednarz, B. P.
    MEDICAL PHYSICS, 2015, 42 (02) : 715 - 725
  • [32] Experimental evaluation of the impact of low tesla transverse magnetic field on dose distribution in presence of tissue interfaces
    Cusumano, Davide
    Teodoli, Stefania
    Greco, Francesca
    Fidanzio, Andrea
    Boldrini, Luca
    Massaccesi, Mariangela
    Cellini, Francesco
    Valentini, Vincenzo
    Azario, Luigi
    De Spirito, Marco
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 53 : 80 - 85
  • [33] Controlled Magnetohydrodynamic Separation of Liquids in a Transverse Magnetic Field
    Shabatura, Yuriy
    Atamanyuk, Vitalij
    Kuznyetsov, Oleksiy
    Parashchuk, Lidiya
    2017 IEEE INTERNATIONAL YOUNG SCIENTISTS FORUM ON APPLIED PHYSICS AND ENGINEERING (YSF), 2017, : 255 - 258
  • [34] Wave beams in an inhomogeneous plasma in a transverse magnetic field
    Bagdoev, AG
    Sedrakian, DM
    ASTROPHYSICS, 2002, 45 (01) : 46 - 51
  • [35] Baxter-Wu model in a transverse magnetic field
    Capponi, Sylvain
    Jahromi, Saeed S.
    Alet, Fabien
    Schmidt, Kai Phillip
    PHYSICAL REVIEW E, 2014, 89 (06):
  • [36] Wave Beams in an Inhomogeneous Plasma in a Transverse Magnetic Field
    A. G. Bagdoev
    D. M. Sedrakian
    Astrophysics, 2002, 45 : 46 - 51
  • [37] Linear Optical Response of Silicon Nanotubes Under Axial Magnetic Field
    Chegel, Raad
    Behzad, Somayeh
    JOURNAL OF ELECTRONIC MATERIALS, 2013, 42 (01) : 58 - 70
  • [38] Limitations of analytical dose calculations for small field proton radiosurgery
    Geng, Changran
    Daartz, Juliane
    Lam-Tin-Cheung, Kimberley
    Bussiere, Marc
    Shih, Helen A.
    Paganetti, Harald
    Schuemann, Jan
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (01) : 246 - 257
  • [39] Effect of uniform magnetic field on dose distribution in the breast radiotherapy
    Esmaeeli, A. D.
    Pouladian, M.
    Monfared, A. S.
    Mahdavi, S. R.
    Moslemi, D.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2014, 12 (02): : 151 - 160
  • [40] Asymptotic solutions for the asymmetric flow in a channel with porous retractable walls under a transverse magnetic field
    Hongxia GUO
    Ping LIN
    Lin LI
    AppliedMathematicsandMechanics(EnglishEdition), 2018, 39 (08) : 1147 - 1164