The dose response of PTW microDiamond and microSilicon in transverse magnetic field under small field conditions

被引:11
|
作者
Blum, Isabel [1 ]
Tekin, Tuba [1 ]
Delfs, Bjorn [1 ]
Schonfeld, Ann-Britt [1 ]
Kapsch, Ralf-Peter [2 ]
Poppe, Bjorn [1 ]
Looe, Hui Khee [1 ]
机构
[1] Carl von Ossietzky Univ Oldenburg, Univ Clin Med Radiat Phys, Pius Hosp, Med Campus, Oldenburg, Germany
[2] Phys Tech Bundesanstalt, Braunschweig, Germany
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2021年 / 66卷 / 15期
关键词
magnetic resonance guided radiation therapy; Lorentz force; magnetic field; small field dosimetry; dose response; small field output correction factors; MONTE-CARLO; ION CHAMBERS; DETECTORS; DOSIMETRY; DIAMOND; DIODE;
D O I
10.1088/1361-6560/ac0f2e
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The aim of the present work is to investigate the behavior of two diode-type detectors (PTW microDiamond 60019 and PTW microSilicon 60023) in transverse magnetic field under small field conditions. A formalism based on TRS 483 has been proposed serving as the framework for the application of these high-resolution detectors under these conditions. Measurements were performed at the National Metrology Institute of Germany (PTB, Braunschweig) using a research clinical linear accelerator facility. Quadratic fields corresponding to equivalent square field sizes S between 0.63 and 4.27 cm at the depth of measurement were used. The magnetic field strength was varied up to 1.4 T. Experimental results have been complemented with Monte Carlo simulations up to 1.5 T. Detailed simulations were performed to quantify the small field perturbation effects and the influence of detector components on the dose response. The does response of both detectors decreases by up to 10% at 1.5 T in the largest field size investigated. In S = 0.63 cm, this reduction at 1.5 T is only about half of that observed in field sizes S > 2 cm for both detectors. The results of the Monte Carlo simulations show agreement better than 1% for all investigated conditions. Due to normalization at the machine specific reference field, the resulting small field output correction factors for both detectors in magnetic field k(Qclin,Qmsr)(B) are smaller than those in the magnetic field-free case, where correction up to 6.2% at 1.5 T is required for the microSilicon in the smallest field size investigated. The volume-averaging effect of both detectors was shown to be nearly independent of the magnetic field. The influence of the enhanced-density components within the detectors has been identified as the major contributors to their behaviors in magnetic field. Nevertheless, the effect becomes weaker with decreasing field size that may be partially attributed to the deficiency of low energy secondary electrons originated from distant locations in small fields.
引用
收藏
页数:15
相关论文
共 50 条
  • [1] Is the PTW 60019 microDiamond a suitable candidate for small field reference dosimetry?
    De Coste, Vanessa
    Francescon, Paolo
    Marinelli, Marco
    Masi, Laura
    Paganini, Lucia
    Pimpinella, Maria
    Prestopino, Giuseppe
    Russo, Serenella
    Stravato, Antonella
    Verona, Claudio
    Verona-Rinati, Gianluca
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (17): : 7036 - 7055
  • [2] Model-based machine learning for the recovery of lateral dose profiles of small photon fields in magnetic field
    Looe, Hui Khee
    Blum, Isabel
    Schoenfeld, Ann-Britt
    Tekin, Tuba
    Delfs, Bjoern
    Poppe, Bjoern
    PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (08):
  • [3] Thermoelectric figure of merit in transverse magnetic field under adiabatic and isothermal conditions
    Tarkhanyan, Roland H.
    Niarchos, Dimitris G.
    JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 2020, 493
  • [4] Effect of Transverse Steady Magnetic Field on MHD Flow Under Free Convection Conditions in Vertical Microchannels
    Hamdan, Mohammad A.
    Al-Assaf, Anwar H.
    Al-Nimr, Mohammad A.
    JORDAN JOURNAL OF MECHANICAL AND INDUSTRIAL ENGINEERING, 2018, 12 (02): : 131 - 139
  • [5] The role of the construction and sensitive volume of compact ionization chambers on the magnetic field-dependent dose response
    Delfs, Bjoern
    Blum, Isabel
    Tekin, Tuba
    Schoenfeld, Ann-Britt
    Kranzer, Rafael
    Poppinga, Daniela
    Giesen, Ulrich
    Langner, Frank
    Kapsch, Ralf-Peter
    Poppe, Bjoern
    Looe, Hui Khee
    MEDICAL PHYSICS, 2021, 48 (08) : 4572 - 4585
  • [6] Small field dosimetry correction factors for circular and MLC shaped fields with the CyberKnife M6 System: evaluation of the PTW 60023 microSilicon detector
    Francescon, P.
    Kilby, W.
    Noll, J. M.
    Satariano, N.
    Orlandi, C.
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (01):
  • [7] Calculations of magnetic field correction factors for ionization chambers in a transverse magnetic field using Monte Carlo code TOPAS
    Mao, Lingli
    Pei, Xi
    Chao, Tsi-Chian
    Xu, Xie George
    RADIATION PHYSICS AND CHEMISTRY, 2021, 183
  • [8] Characterization of a microSilicon diode detector for small-field photon beam dosimetry
    Akino, Yuichi
    Fujiwara, Masateru
    Okamura, Keita
    Shiomi, Hiroya
    Mizuno, Hirokazu
    Isohashi, Fumiaki
    Suzuki, Osamu
    Seo, Yuji
    Tamari, Keisuke
    Ogawa, Kazuhiko
    JOURNAL OF RADIATION RESEARCH, 2020, 61 (03) : 410 - 418
  • [9] Small field output correction factors of the microSilicon detector and a deeper understanding of their origin by quantifying perturbation factors
    Weber, Carolin
    Kranzer, Rafael
    Weidner, Jan
    Kroeninger, Kevin
    Poppe, Bjoern
    Looe, Hui Khee
    Poppinga, Daniela
    MEDICAL PHYSICS, 2020, 47 (07) : 3165 - 3173
  • [10] Fluid structure interaction of cantilever micro and nanotubes conveying magnetic fluid with small size effects under a transverse magnetic field
    Sadeghi-Goughari, Moslem
    Jeon, Soo
    Kwon, Hyock-Ju
    JOURNAL OF FLUIDS AND STRUCTURES, 2020, 94