The dose response of PTW microDiamond and microSilicon in transverse magnetic field under small field conditions

被引:11
作者
Blum, Isabel [1 ]
Tekin, Tuba [1 ]
Delfs, Bjorn [1 ]
Schonfeld, Ann-Britt [1 ]
Kapsch, Ralf-Peter [2 ]
Poppe, Bjorn [1 ]
Looe, Hui Khee [1 ]
机构
[1] Carl von Ossietzky Univ Oldenburg, Univ Clin Med Radiat Phys, Pius Hosp, Med Campus, Oldenburg, Germany
[2] Phys Tech Bundesanstalt, Braunschweig, Germany
关键词
magnetic resonance guided radiation therapy; Lorentz force; magnetic field; small field dosimetry; dose response; small field output correction factors; MONTE-CARLO; ION CHAMBERS; DETECTORS; DOSIMETRY; DIAMOND; DIODE;
D O I
10.1088/1361-6560/ac0f2e
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The aim of the present work is to investigate the behavior of two diode-type detectors (PTW microDiamond 60019 and PTW microSilicon 60023) in transverse magnetic field under small field conditions. A formalism based on TRS 483 has been proposed serving as the framework for the application of these high-resolution detectors under these conditions. Measurements were performed at the National Metrology Institute of Germany (PTB, Braunschweig) using a research clinical linear accelerator facility. Quadratic fields corresponding to equivalent square field sizes S between 0.63 and 4.27 cm at the depth of measurement were used. The magnetic field strength was varied up to 1.4 T. Experimental results have been complemented with Monte Carlo simulations up to 1.5 T. Detailed simulations were performed to quantify the small field perturbation effects and the influence of detector components on the dose response. The does response of both detectors decreases by up to 10% at 1.5 T in the largest field size investigated. In S = 0.63 cm, this reduction at 1.5 T is only about half of that observed in field sizes S > 2 cm for both detectors. The results of the Monte Carlo simulations show agreement better than 1% for all investigated conditions. Due to normalization at the machine specific reference field, the resulting small field output correction factors for both detectors in magnetic field k(Qclin,Qmsr)(B) are smaller than those in the magnetic field-free case, where correction up to 6.2% at 1.5 T is required for the microSilicon in the smallest field size investigated. The volume-averaging effect of both detectors was shown to be nearly independent of the magnetic field. The influence of the enhanced-density components within the detectors has been identified as the major contributors to their behaviors in magnetic field. Nevertheless, the effect becomes weaker with decreasing field size that may be partially attributed to the deficiency of low energy secondary electrons originated from distant locations in small fields.
引用
收藏
页数:15
相关论文
共 38 条
[1]   On the Monte Carlo simulation of small-field micro-diamond detectors for megavoltage photon dosimetry [J].
Andreo, Pedro ;
Palmans, Hugo ;
Marteinsdottir, Maria ;
Benmakhlouf, Hamza ;
Carlsson-Tedgren, Asa .
PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (01) :L1-L10
[2]   Detector to detector corrections: A comprehensive experimental study of detector specific correction factors for beam output measurements for small radiotherapy beams [J].
Azangwe, Godfrey ;
Grochowska, Paulina ;
Georg, Dietmar ;
Izewska, Joanna ;
Hopfgartner, Johannes ;
Lechner, Wolfgang ;
Andersen, Claus E. ;
Beierholm, Anders R. ;
Helt-Hansen, Jakob ;
Mizuno, Hideyuki ;
Fukumura, Akifumi ;
Yajima, Kaori ;
Gouldstone, Clare ;
Sharpe, Peter ;
Meghzifene, Ahmed ;
Palmans, Hugo .
MEDICAL PHYSICS, 2014, 41 (07)
[3]   Evaluation of a synthetic single-crystal diamond detector for relative dosimetry measurements on a CyberKnife™ [J].
Chalkley, A. ;
Heyes, G. .
BRITISH JOURNAL OF RADIOLOGY, 2014, 87 (1035)
[4]   MR-guidance in clinical reality: current treatment challenges and future perspectives [J].
Corradini, S. ;
Alongi, F. ;
Andratschke, N. ;
Belka, C. ;
Boldrini, L. ;
Cellini, F. ;
Debus, J. ;
Guckenberger, M. ;
Hoerner-Rieber, J. ;
Lagerwaard, F. J. ;
Mazzola, R. ;
Palacios, M. A. ;
Philippens, M. E. P. ;
Raaijmakers, C. P. J. ;
Terhaard, C. H. J. ;
Valentini, V. ;
Niyazi, M. .
RADIATION ONCOLOGY, 2019, 14 (1)
[5]   Monte Carlo calculated correction factors for diodes and ion chambers in small photon fields [J].
Czarnecki, D. ;
Zink, K. .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (08) :2431-2444
[6]   First Water Calorimetric Dw Measurement and Direct Measurement of Magnetic Field Correction Factors, KQ, B, in a 1.5 T B-Field of An MRI Linac [J].
de Prez, L. ;
de Pooter, J. ;
Jansen, B. ;
Wolthaus, J. ;
van Asselen, B. ;
Woodings, S. ;
Soest, T. ;
Kok, J. ;
Raaymakers, B. .
MEDICAL PHYSICS, 2016, 43 (06) :3874-3874
[7]   The role of the construction and sensitive volume of compact ionization chambers on the magnetic field-dependent dose response [J].
Delfs, Bjoern ;
Blum, Isabel ;
Tekin, Tuba ;
Schoenfeld, Ann-Britt ;
Kranzer, Rafael ;
Poppinga, Daniela ;
Giesen, Ulrich ;
Langner, Frank ;
Kapsch, Ralf-Peter ;
Poppe, Bjoern ;
Looe, Hui Khee .
MEDICAL PHYSICS, 2021, 48 (08) :4572-4585
[8]   The 1D lateral dose response functions of photon-dosimetry detectors in magnetic fields-measurement and Monte-Carlo simulation [J].
Delfs, Bjoern ;
Poppinga, Daniela ;
Ulrichs, Ann-Britt ;
Kapsch, Ralf-Peter ;
Harder, Dietrich ;
Poppe, Bjoern ;
Looe, Hui Khee .
PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (19)
[9]   Origins of the changing detector response in small megavoltage photon radiation fields [J].
Fenwick, John D. ;
Georgiou, Georgios ;
Rowbottom, Carl G. ;
Underwood, Tracy S. A. ;
Kumar, Sudhir ;
Nahum, Alan E. .
PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (12)
[10]   Assessment of the setup dependence of detector response functions for mega-voltage linear accelerators [J].
Fox, Christopher ;
Simon, Tom ;
Simon, Bill ;
Dempsey, James F. ;
Kahler, Darren ;
Palta, Jatinder R. ;
Liu, Chihray ;
Yan, Guanghua .
MEDICAL PHYSICS, 2010, 37 (02) :477-484