Evaluation of the Vitek 2 system for rapid identification of clinical isolates of gram-negative bacilli and members of the family Streptococcaceae

被引:29
作者
Gavin, PJ
Warren, JR
Obias, AA
Collins, SM
Peterson, LR
机构
[1] Evanston NW Healthcare, Dept Pathol, Evanston, IL 60201 USA
[2] NW Mem Hosp, Clin Microbiol Div, Dept Pathol, Chicago, IL 60611 USA
[3] Northwestern Univ, Sch Med, Chicago, IL 60611 USA
关键词
D O I
10.1007/s10096-002-0826-x
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
Accuracy of the Vitek 2 automated system (bioMerieux Vitek, USA) for rapid identification of bacteria was evaluated using a collection of 858 epidemiologically unrelated gram-negative and 99 gram-positive clinical isolates. Isolates were tested after subculturing to ensure purity. Conventional agar-based biochemical tests (Steers replicator) were used as a reference method of identification. Gram-negative bacteria were identified to the species level with 95.3% accuracy by the system (Enterobacteriaceae, 95.9%; and non-Enterobacteriaceae, 92.5%), and gram-positive isolates with 72% accuracy. Although Vitek 2 identified routine clinical isolates of gram-negative bacilli and Enterococcus faecalis and Enterococcus faecium reliably, rapidly, and reproducibly, improvement is required in the identification of less common species of enterococci and viridans group streptococci.
引用
收藏
页码:869 / 874
页数:6
相关论文
共 10 条
[1]   Clinical and financial benefits of rapid bacterial identification and antimicrobial susceptibility testing [J].
Barenfanger, J ;
Drake, C ;
Kacich, G .
JOURNAL OF CLINICAL MICROBIOLOGY, 1999, 37 (05) :1415-1418
[2]   CLINICAL IMPACT OF RAPID IN-VITRO SUSCEPTIBILITY TESTING AND BACTERIAL IDENTIFICATION [J].
DOERN, GV ;
VAUTOUR, R ;
GAUDET, M ;
LEVY, B .
JOURNAL OF CLINICAL MICROBIOLOGY, 1994, 32 (07) :1757-1762
[3]   Evaluation of the VITEK 2 system for rapid identification of medically relevant gram-negative rods [J].
Funke, G ;
Monnet, D ;
deBernardis, C ;
von Graevenitz, A ;
Freney, J .
JOURNAL OF CLINICAL MICROBIOLOGY, 1998, 36 (07) :1948-1952
[4]  
Garcia-Garrote F, 2000, J CLIN MICROBIOL, V38, P2108
[5]   Evaluation of an automated system for identification of Enterobacteriaceae and nonfermenting bacilli [J].
Jossart, MF ;
Courcol, RJ .
EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 1999, 18 (12) :902-907
[6]   Evaluation of the VITEK 2 system for the identification and susceptibility testing of three species of nonfermenting gram-negative rods frequently isolated from clinical samples [J].
Joyanes, P ;
Conejo, MD ;
Martínez-Martínez, L ;
Perea, EJ .
JOURNAL OF CLINICAL MICROBIOLOGY, 2001, 39 (09) :3247-3253
[7]   Evaluation of VITEK 2 rapid identification and susceptibility testing system against Gram-negative clinical isolates [J].
Ling, TKW ;
Tam, PC ;
Liu, ZK ;
Cheng, AFB .
JOURNAL OF CLINICAL MICROBIOLOGY, 2001, 39 (08) :2964-2966
[8]  
Murray P.R., 1999, MANUAL CLIN MICROBIO, V7th
[9]   AUTOMATED SYSTEMS FOR IDENTIFICATION OF MICROORGANISMS [J].
STAGER, CE ;
DAVIS, JR .
CLINICAL MICROBIOLOGY REVIEWS, 1992, 5 (03) :302-327
[10]   CLINICAL IMPACT OF RAPID IDENTIFICATION AND SUSCEPTIBILITY TESTING OF BACTERIAL BLOOD CULTURE ISOLATES [J].
TRENHOLME, GM ;
KAPLAN, RL ;
KARAKUSIS, PH ;
STINE, T ;
FUHRER, J ;
LANDAU, W ;
LEVIN, S .
JOURNAL OF CLINICAL MICROBIOLOGY, 1989, 27 (06) :1342-1345