Quality assurance of an image guided intracranial stereotactic positioning system for radiosurgery treatment with helical tomotherapy

被引:8
作者
Soisson, Emilie T. [1 ,2 ]
Hardcastle, Nick [2 ,3 ]
Tome, Wolfgang A. [1 ,2 ,3 ]
机构
[1] Univ Wisconsin, Sch Med & Publ Hlth, Dept Human Oncol, Madison, WI 53792 USA
[2] Univ Wisconsin, Sch Med & Publ Hlth, Dept Med Phys, Madison, WI 53792 USA
[3] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW, Australia
关键词
Stereotactic radiosurgery; Tomotherapy; Stereotactic localization; Image guidance; VERIFICATION; LOCALIZATION;
D O I
10.1007/s11060-010-0227-4
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The aim of this work was to determine the accuracy and precision of stereotactic localization and treatment delivery using a helical tomotherapy based stereotactic radiosurgery (SRS) system. A tomotherapy specific radiosurgery workflow was designed that exploits the system's on board megavotage CT (MVCT) imaging system so that it not only provides a pre-treatment volumetric verification image that can be used for stereotactic localization, eliminating the need for a patient-frame based coordinate system, but also supplies the treatment planning image. Using an imaging guidance based intracranial stereotactic positioning system, a head ring and tabletop docking device are used only for fixation, while image guidance is used for localization. Due to the unconventional workflow, a methodology for determining the localization accuracy was developed and results were compared to other linear accelerator based radiosurgery systems. In this work, the localization error using volumetric localization was found to be 0.45 mm +/- 0.17 mm, indicating a localization precision of 0.3 mm within a 95% confidence interval. In addition, procedures for testing the delivery accuracy of the Tomotherapy system are described. Results show that the accuracy of the delivery can be verified to within +/- 1 voxel dimension. These results are well within conventional SRS tolerances and compare favorably to other linear accelerator based techniques.
引用
收藏
页码:277 / 285
页数:9
相关论文
共 14 条
[1]   SU-GG-J-75: Dosimetric and Image Quality Analysis of a New Ultrafine Imaging Mode in TomoTherapy [J].
Bichay, T. ;
Chen, C. ;
Meadows, J. ;
Schippers, D. ;
Lucas, D. ;
Ruchala, K. ;
Chao, E. .
MEDICAL PHYSICS, 2008, 35 (06)
[2]   Automatic registration of megavoltage to kilovoltage CT images in helical tomotherapy:: An evaluation of the setup verification process for the special case of a rigid head phantom [J].
Boswell, Sarah ;
Tome, Wolfgang ;
Jeraj, Robert ;
Jaradat, Hazirn ;
Mackie, T. Rock .
MEDICAL PHYSICS, 2006, 33 (11) :4395-4404
[3]   STEREOTAXIC FRAME AND COMPUTER SOFTWARE FOR CT-DIRECTED NEUROSURGICAL LOCALIZATION [J].
BROWN, RA ;
ROBERTS, TS ;
OSBORN, AG .
INVESTIGATIVE RADIOLOGY, 1980, 15 (04) :308-312
[4]  
Friedman W A, 1992, Neurosurg Clin N Am, V3, P141
[5]   THE UNIVERSITY-OF-FLORIDA RADIOSURGERY SYSTEM [J].
FRIEDMAN, WA ;
BOVA, FJ .
SURGICAL NEUROLOGY, 1989, 32 (05) :334-342
[6]   Dosimetric verification of helical tomotherapy for total scalp irradiation [J].
Hardcastle, Nicholas ;
Soisson, Emilie ;
Metcalfe, Peter ;
Rosenfeld, Anatoly B. ;
Tome, Wolfgang A. .
MEDICAL PHYSICS, 2008, 35 (11) :5061-5068
[7]   Endo-rectal balloon cavity dosimetry in a phantom: Performance under IMRT and helical tomotherapy beams [J].
Hardcastle, Nicholas ;
Metcalfe, Peter E. ;
Rosenfeld, Anatoly B. ;
Tome, Wolfgang A. .
RADIOTHERAPY AND ONCOLOGY, 2009, 92 (01) :48-56
[8]   Intracranial stereotactic positioning systems: report of the American Association of Physicists in Medicine Radiation Therapy Committee Task Group no. 68 [J].
Lightstone, AW ;
Benedict, SH ;
Bova, FJ ;
Solberg, TD ;
Stern, RL .
MEDICAL PHYSICS, 2005, 32 (07) :2380-2398
[9]   A technique for the quantitative evaluation of dose distributions [J].
Low, DA ;
Harms, WB ;
Mutic, S ;
Purdy, JA .
MEDICAL PHYSICS, 1998, 25 (05) :656-661
[10]  
Schell M.C., 1995, STEREOTACTIC RADIOSU, DOI 10.37206/53