Digital overlay technique for documenting toric intraocular lens axis orientation

被引:38
作者
Nguyen, TM
Miller, KM
机构
[1] Univ Calif Los Angeles, Sch Med, Jules Stein Eye Inst, Los Angeles, CA 90095 USA
[2] Univ Calif Los Angeles, Sch Med, Dept Ophthalmol, Los Angeles, CA 90095 USA
关键词
D O I
10.1016/S0886-3350(00)00442-9
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To describe a digital overlay technique for documenting toric intraocular lens (IOL) axis alignment. Setting: Jules Stein Eye Institute and the Department of Ophthalmology, University of California at Los Angeles School of Medicine, Los Angeles, California, USA. Methods: Digital overlay imaging was used to evaluate the alignment of Staar toric IOLs in 4 eyes of 3 patients who had regular corneal astigmatism at the time of cataract surgery. Lens axes were determined by computerized analysis of digitally scanned retroillumination photographs. A stock digital image of a Staar AA4203TF toric IOL was super-imposed on corneal topography images to document IOL alignment with the steep corneal meridian. Results: Digital overlay images demonstrated that 3 IOLs in the sample group were within 5 degrees and 1 was within 20 degrees of the intended axis at the time of the final postoperative examination. Slitlamp lens axis estimations were 3 to 18 degrees different from that determined by this technique. Conclusion: Digital overlay imaging of correctly oriented toric IOLs on computerized corneal topography maps represents an intuitive, accurate, and visually appealing method of documenting toric IOL axis alignment. J Cataract Refract Surg 2000; 26:1496-1504 (C) 2000 ASCRS and ESCRS.
引用
收藏
页码:1496 / 1504
页数:9
相关论文
共 14 条
[1]   COMPARISON OF THE ACCURACY OF COMPUTERIZED VIDEOKERATOGRAPHY AND KERATOMETRY FOR USE IN INTRAOCULAR-LENS CALCULATIONS [J].
CUAYCONG, MJ ;
GAY, CA ;
EMERY, J ;
HAFT, EA ;
KOCH, DD .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1993, 19 :178-181
[2]   PRESCRIBING CYLINDERS - PROBLEM OF DISTORTION [J].
GUYTON, DL .
SURVEY OF OPHTHALMOLOGY, 1977, 22 (03) :177-188
[3]   Computerized videokeratography and keratometry in determining intraocular lens calculations [J].
Husain, SE ;
Kohnen, T ;
Maturi, R ;
Er, H ;
Koch, DD .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1996, 22 (03) :362-366
[4]   Decentration of 3-piece versus plate-haptic silicone intraocular lenses [J].
Hwang, IP ;
Clinch, TE ;
Moshifar, M ;
Malmquist, L ;
Cason, M ;
Crandall, AS .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1998, 24 (11) :1505-1508
[5]   COMPARISON OF THE INDUCED ASTIGMATISM AFTER TEMPORAL CLEAR CORNEAL TUNNEL INCISIONS OF DIFFERENT SIZES [J].
KOHNEN, T ;
DICK, B ;
JACOBI, KW .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1995, 21 (04) :417-424
[6]   Comparison of two plate-haptic intraocular lenses in a rabbit model [J].
Mamalis, N ;
Omar, O ;
Veiga, J ;
Tanner, D ;
Pirayesh, A ;
Fernquist, DS .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1996, 22 :1291-1295
[7]   EFFECT OF CATARACT WOUND INCISION SIZE ON ACUTE CHANGES IN CORNEAL TOPOGRAPHY [J].
MARTIN, RG ;
SANDERS, DR ;
MILLER, JD ;
COX, CC ;
BALLEW, C .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1993, 19 :170-177
[8]   Astigmatism and the toric intraocular lens and other vertex distance effects [J].
Novis, C .
SURVEY OF OPHTHALMOLOGY, 1997, 42 (03) :268-270
[9]   Three year prospective, randomized evaluation of intraocular lens implantation through 3.2 and 5.5 mm incisions [J].
Oshika, T ;
Nagahara, K ;
Yaguchi, S ;
Emi, K ;
Takenaka, H ;
Tsuboi, S ;
Yoshitomi, F ;
Nagamoto, T ;
Kurosaka, D .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1998, 24 (04) :509-514
[10]   WHEN KERATOMETRIC MEASUREMENTS DO NOT ACCURATELY REFLECT CORNEAL TOPOGRAPHY [J].
SANDERS, DR ;
GILLS, JP ;
MARTIN, RG .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1993, 19 :131-135