Spatial Resolution and Image Qualities of Zr-89 on Siemens Biograph TruePoint PET/CT

被引:16
作者
Lee, Young Sub [1 ,2 ]
Kim, Jin Su [1 ]
Kim, Jung Young [1 ]
Kim, Byung Il [1 ]
Lim, Sang Moo [1 ]
Kim, Hee-Joung [2 ]
机构
[1] Korea Inst Radiol & Med Sci, Mol Imaging Res Ctr, Seoul 139706, South Korea
[2] Yonsei Univ, Res Inst Hlth Sci, Dept Radiat Convergence Engn, Wonju, South Korea
关键词
radioimmunotherapy (RIT); PET; spatial resolution; image quality; Zr-89; I-131; RADIOIMMUNOTHERAPY; PERFORMANCE-MEASUREMENT; DOSIMETRY; VIEW;
D O I
10.1089/cbr.2014.1709
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Zirconium-89 (t(1/2)=78.41 hours) is an ideal metallic radioisotope for immuno-positron emission tomography (PET), given that its physical half-life closely matches the biological half-life of monoclonal antibodies. In this study, the authors measured the spatial resolution and image quality of Zr-89 PET and compared the results against those obtained using F-18 PET, which is widely regarded as the gold standard for comparison of imaging characteristics. Materials and Methods: The spatial resolution and image qualities of Zr-89 were measured on the Siemens Biograph Truepoint TrueV PET/CT scanner, partly according to NEMA NU2-2007 standards. For spatial resolution measurement, the Zr-89 point source was located at the center of the axial field of view (FOV) and offset 1/4 axial FOV from the center. For image quality measurements, an NEMA IEC Phantom was used. The NEMA IEC Phantom consists of six hot spheres that were filled with Zr-89 solution. Spatial resolution and image quality (%contrast, %background variability [BV], and source to background ratio [SBR]) were assessed to compare the imaging characteristics of F-18 with those of Siemens Biograph Truepoint TrueV. Results: The transverse and axial spatial resolutions at 1 cm were 4.5 and 4.7 mm for Zr-89, respectively. The %contrast of Zr-89 was 25.5% for the smallest 10 mm sized sphere and 89.8% for the largest 37 mm sized sphere, and for F-18, it was 32.5% for the smallest 10 mm sized sphere and 103.9% for the largest 37 mm sized sphere using the ordered subset expectation maximization (OSEM) reconstruction method. The %BV of F-18 PET was 6.4% for the smallest 10 mm sized sphere and 3.5% for the largest 37 mm sized sphere using the OSEM reconstruction. The SBR of Zr-89 was 1.8 for the smallest 10 mm sized sphere and 3.7 for the largest 37 mm sized sphere, and for F-18, it was 2.0 for the smallest 10 mm sized sphere and 4.1 for the largest 37 mm sized sphere using the OSEM reconstruction method. Conclusions: This study assessed Zr-89 imaging characteristics using a Siemens Biograph Truepoint TrueV PET/CT scanner and compared the results with those obtained for F-18 PET. Although spatial resolution and image quality of Zr-89 PET were lower compared with F-18 PET, due to longer positron range and low positron branching ratio, Zr-89 is advantageous for immuno-PET due to well-matched half-life with monoclonal antibodies.
引用
收藏
页码:27 / 32
页数:6
相关论文
共 18 条
[1]   Use of Integrated SPECT/CT Imaging for Tumor Dosimetry in I-131 Radioimmunotherapy: A Pilot Patient Study [J].
Dewaraja, Yuni K. ;
Wilderman, Scott J. ;
Koral, Kenneth F. ;
Kaminski, Mark S. ;
Avram, Anca M. .
CANCER BIOTHERAPY AND RADIOPHARMACEUTICALS, 2009, 24 (04) :417-426
[2]   Image-Quality Assessment for Several Positron Emitters Using the NEMA NU 4-2008 Standards in the Siemens Inveon Small-Animal PET Scanner [J].
Disselhorst, Jonathan A. ;
Brom, Maarten ;
Laverman, Peter ;
Slump, Cornelius H. ;
Boerman, Otto C. ;
Oyen, Wim J. G. ;
Gotthardt, Martin ;
Visser, Eric P. .
JOURNAL OF NUCLEAR MEDICINE, 2010, 51 (04) :610-617
[3]   Dosimetric impact of correcting count losses due to deadtime in clinical radioimmunotherapy trials involving iodine-131 scintigraphy [J].
Ferrer, L ;
Delpon, G ;
Lisbona, A ;
Bardiès, M .
CANCER BIOTHERAPY AND RADIOPHARMACEUTICALS, 2003, 18 (01) :117-124
[4]   Standardized methods for the production of high specific-activity zirconium-89 [J].
Holland, Jason P. ;
Sheh, Yiauchung ;
Lewis, Jason S. .
NUCLEAR MEDICINE AND BIOLOGY, 2009, 36 (07) :729-739
[5]   Physical and clinical performance of the mCT time-of-flight PET/CT scanner [J].
Jakoby, B. W. ;
Bercier, Y. ;
Conti, M. ;
Casey, M. E. ;
Bendriem, B. ;
Townsend, D. W. .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (08) :2375-2389
[6]   Performance Characteristics of a New LSO PET/CT Scanner With Extended Axial Field-of-View and PSF Reconstruction [J].
Jakoby, Bjoern W. ;
Bercier, Yanic ;
Watson, Charles C. ;
Bendriem, Bernard ;
Townsend, David W. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (03) :633-639
[7]  
Jönsson L, 2005, J NUCL MED, V46, P1679
[8]   Performance measurement of the microPET focus 120 scanner [J].
Kim, Jin Su ;
Lee, Jae Sung ;
Im, Ki Chun ;
Kim, Su Jin ;
Kim, Seog-Young ;
Lee, Dong Soo ;
Moon, Dae Hyuk .
JOURNAL OF NUCLEAR MEDICINE, 2007, 48 (09) :1527-1535
[9]   Performance measurement of PSF modeling reconstruction (True X) on Siemens Biograph TruePoint TrueV PET/CT [J].
Lee, Young Sub ;
Kim, Jin Su ;
Kim, Kyeong Min ;
Kang, Joo Hyun ;
Lim, Sang Moo ;
Kim, Hee-Joung .
ANNALS OF NUCLEAR MEDICINE, 2014, 28 (04) :340-348
[10]   Imaging Characteristics of 124I Between 3D and 2D on Siemens ECAT HR plus PET Scanner [J].
Lee, Young Sub ;
Kim, Jin Su ;
Kim, Hee-Joung ;
Woo, Sang-Keun ;
Kim, Jong Guk ;
Park, Ji Ae ;
Choi, Chang Woon ;
Lim, Sang Moo ;
Kim, Kyeong Min .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2013, 60 (02) :797-801