Spatial Resolution and Image Qualities of Zr-89 on Siemens Biograph TruePoint PET/CT

被引:16
作者
Lee, Young Sub [1 ,2 ]
Kim, Jin Su [1 ]
Kim, Jung Young [1 ]
Kim, Byung Il [1 ]
Lim, Sang Moo [1 ]
Kim, Hee-Joung [2 ]
机构
[1] Korea Inst Radiol & Med Sci, Mol Imaging Res Ctr, Seoul 139706, South Korea
[2] Yonsei Univ, Res Inst Hlth Sci, Dept Radiat Convergence Engn, Wonju, South Korea
关键词
radioimmunotherapy (RIT); PET; spatial resolution; image quality; Zr-89; I-131; RADIOIMMUNOTHERAPY; PERFORMANCE-MEASUREMENT; DOSIMETRY; VIEW;
D O I
10.1089/cbr.2014.1709
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Zirconium-89 (t(1/2)=78.41 hours) is an ideal metallic radioisotope for immuno-positron emission tomography (PET), given that its physical half-life closely matches the biological half-life of monoclonal antibodies. In this study, the authors measured the spatial resolution and image quality of Zr-89 PET and compared the results against those obtained using F-18 PET, which is widely regarded as the gold standard for comparison of imaging characteristics. Materials and Methods: The spatial resolution and image qualities of Zr-89 were measured on the Siemens Biograph Truepoint TrueV PET/CT scanner, partly according to NEMA NU2-2007 standards. For spatial resolution measurement, the Zr-89 point source was located at the center of the axial field of view (FOV) and offset 1/4 axial FOV from the center. For image quality measurements, an NEMA IEC Phantom was used. The NEMA IEC Phantom consists of six hot spheres that were filled with Zr-89 solution. Spatial resolution and image quality (%contrast, %background variability [BV], and source to background ratio [SBR]) were assessed to compare the imaging characteristics of F-18 with those of Siemens Biograph Truepoint TrueV. Results: The transverse and axial spatial resolutions at 1 cm were 4.5 and 4.7 mm for Zr-89, respectively. The %contrast of Zr-89 was 25.5% for the smallest 10 mm sized sphere and 89.8% for the largest 37 mm sized sphere, and for F-18, it was 32.5% for the smallest 10 mm sized sphere and 103.9% for the largest 37 mm sized sphere using the ordered subset expectation maximization (OSEM) reconstruction method. The %BV of F-18 PET was 6.4% for the smallest 10 mm sized sphere and 3.5% for the largest 37 mm sized sphere using the OSEM reconstruction. The SBR of Zr-89 was 1.8 for the smallest 10 mm sized sphere and 3.7 for the largest 37 mm sized sphere, and for F-18, it was 2.0 for the smallest 10 mm sized sphere and 4.1 for the largest 37 mm sized sphere using the OSEM reconstruction method. Conclusions: This study assessed Zr-89 imaging characteristics using a Siemens Biograph Truepoint TrueV PET/CT scanner and compared the results with those obtained for F-18 PET. Although spatial resolution and image quality of Zr-89 PET were lower compared with F-18 PET, due to longer positron range and low positron branching ratio, Zr-89 is advantageous for immuno-PET due to well-matched half-life with monoclonal antibodies.
引用
收藏
页码:27 / 32
页数:6
相关论文
共 18 条
  • [1] Use of Integrated SPECT/CT Imaging for Tumor Dosimetry in I-131 Radioimmunotherapy: A Pilot Patient Study
    Dewaraja, Yuni K.
    Wilderman, Scott J.
    Koral, Kenneth F.
    Kaminski, Mark S.
    Avram, Anca M.
    [J]. CANCER BIOTHERAPY AND RADIOPHARMACEUTICALS, 2009, 24 (04) : 417 - 426
  • [2] Image-Quality Assessment for Several Positron Emitters Using the NEMA NU 4-2008 Standards in the Siemens Inveon Small-Animal PET Scanner
    Disselhorst, Jonathan A.
    Brom, Maarten
    Laverman, Peter
    Slump, Cornelius H.
    Boerman, Otto C.
    Oyen, Wim J. G.
    Gotthardt, Martin
    Visser, Eric P.
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2010, 51 (04) : 610 - 617
  • [3] Dosimetric impact of correcting count losses due to deadtime in clinical radioimmunotherapy trials involving iodine-131 scintigraphy
    Ferrer, L
    Delpon, G
    Lisbona, A
    Bardiès, M
    [J]. CANCER BIOTHERAPY AND RADIOPHARMACEUTICALS, 2003, 18 (01) : 117 - 124
  • [4] Standardized methods for the production of high specific-activity zirconium-89
    Holland, Jason P.
    Sheh, Yiauchung
    Lewis, Jason S.
    [J]. NUCLEAR MEDICINE AND BIOLOGY, 2009, 36 (07) : 729 - 739
  • [5] Physical and clinical performance of the mCT time-of-flight PET/CT scanner
    Jakoby, B. W.
    Bercier, Y.
    Conti, M.
    Casey, M. E.
    Bendriem, B.
    Townsend, D. W.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (08) : 2375 - 2389
  • [6] Performance Characteristics of a New LSO PET/CT Scanner With Extended Axial Field-of-View and PSF Reconstruction
    Jakoby, Bjoern W.
    Bercier, Yanic
    Watson, Charles C.
    Bendriem, Bernard
    Townsend, David W.
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (03) : 633 - 639
  • [7] Jönsson L, 2005, J NUCL MED, V46, P1679
  • [8] Performance measurement of the microPET focus 120 scanner
    Kim, Jin Su
    Lee, Jae Sung
    Im, Ki Chun
    Kim, Su Jin
    Kim, Seog-Young
    Lee, Dong Soo
    Moon, Dae Hyuk
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2007, 48 (09) : 1527 - 1535
  • [9] Performance measurement of PSF modeling reconstruction (True X) on Siemens Biograph TruePoint TrueV PET/CT
    Lee, Young Sub
    Kim, Jin Su
    Kim, Kyeong Min
    Kang, Joo Hyun
    Lim, Sang Moo
    Kim, Hee-Joung
    [J]. ANNALS OF NUCLEAR MEDICINE, 2014, 28 (04) : 340 - 348
  • [10] Imaging Characteristics of 124I Between 3D and 2D on Siemens ECAT HR plus PET Scanner
    Lee, Young Sub
    Kim, Jin Su
    Kim, Hee-Joung
    Woo, Sang-Keun
    Kim, Jong Guk
    Park, Ji Ae
    Choi, Chang Woon
    Lim, Sang Moo
    Kim, Kyeong Min
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2013, 60 (02) : 797 - 801