Agreement of white-to-white measurements with the IOLMaster 700, Atlas 9000, and Sirius systems

被引:16
作者
Ferrer-Blasco, Teresa [1 ]
Esteve-Taboada, Jose J. [1 ]
Martinez-Albert, Noelia [1 ]
Alfonso, Jose F. [2 ,3 ]
Montes-Mico, Robert [1 ]
机构
[1] Univ Valencia, Dept Opt & Optometry & Visual Sci, C Dr Moliner 50, E-46100 Burjassot, Spain
[2] Fernandez Vega Ophthalmol Inst, Surg Dept, Oviedo, Spain
[3] Univ Oviedo, Sch Med, Oviedo, Spain
关键词
White-to-White; IOLMaster; 700; Sirius; Atlas; 9000; agreement studies; PARTIAL-COHERENCE INTERFEROMETRY; ANTERIOR SEGMENT MEASUREMENTS; OCULAR BIOMETRY MEASUREMENTS; IMPLANTABLE COLLAMER LENS; OPTICAL BIOMETER; PRECISION REPEATABILITY; STATISTICAL-METHODS; REFLECTOMETRY; REPRODUCIBILITY; KERATOMETRY;
D O I
10.1080/17434440.2018.1481745
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Current advances in ocular surgery require the evaluation of differences in automated white-to-white (WTW) measurements between systems. We aim to analyze three different commonly used devices in clinical practice: the IOLMaster 700, the Atlas 9000, and the Sirius. Methods: Ninety right eyes of ninety healthy subjects were included. Measurements of WTW distances were taken with each device for all subjects in the same session and by the same observer. Statistical comparison was carried out by means of the rANOVA procedure and the Bonferroni post-hoc test. In addition, the Bland-Altman analysis was applied in order to perform pairwise comparisons. Results: Average WTW values obtained by the IOLMaster 700, the Atlas 9000, and the Sirius devices were 12.180.40-mm, 12.32 +/- 0.37-mm and 11.90 +/- 0.37-mm, respectively. There were statistically significant differences in all pairwise comparisons. Mean differences revealed that the Atlas 9000 provided the highest WTW values, followed by the IOLMaster 700 values and finally by the Sirius results, which had the lowest values. The limits of agreement obtained in all pairwise comparisons were wide, with a range between 0.64 and 0.75mm. Conclusions: According to our results, these three devices should not be used interchangeably for WTW diameter measurements during the clinical practice.
引用
收藏
页码:453 / 459
页数:7
相关论文
共 35 条
[1]   Comparison of ocular biometry measurements by applanation and immersion A-scan techniques [J].
Ademola-Popoola, Dupe S. ;
Nzeh, Donald A. ;
Saka, Sadiat E. ;
Olokoba, Lateefat B. ;
Obajolowo, Tokunbo S. .
JOURNAL OF CURRENT OPHTHALMOLOGY, 2015, 27 (3-4) :110-114
[2]   Agreement and clinical comparison between a new swept-source optical coherence tomography-based optical biometer and an optical low-coherence reflectometry biometer [J].
Arriola-Villalobos, P. ;
Almendral-Gomez, J. ;
Garzon, N. ;
Ruiz-Medrano, J. ;
Fernandez-Perez, C. ;
Martinez-de-la-Casa, J. M. ;
Diaz-Valle, D. .
EYE, 2017, 31 (03) :437-442
[3]   Comparison of manual and automated methods to determine horizontal corneal diameter [J].
Baumeister, M ;
Terzi, E ;
Ekici, Y ;
Kohnen, T .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2004, 30 (02) :374-380
[4]  
Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
[5]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[6]   SOME THEOREMS ON QUADRATIC FORMS APPLIED IN THE STUDY OF ANALYSIS OF VARIANCE PROBLEMS .2. EFFECTS OF INEQUALITY OF VARIANCE AND OF CORRELATION BETWEEN ERRORS IN THE 2-WAY CLASSIFICATION [J].
BOX, GEP .
ANNALS OF MATHEMATICAL STATISTICS, 1954, 25 (03) :484-498
[7]   Comparison of the measurements of a novel optical biometry: Nidek AL-Scan with Sirius and a ultrasound biometry [J].
Caglar, Cagatay ;
Kocamis, Sucattin Ilker ;
Demir, Emre ;
Durmus, Mustafa .
INTERNATIONAL OPHTHALMOLOGY, 2017, 37 (03) :491-498
[8]   Comparison of Anterior Segment Measurements with Optical Low-coherence Reflectometry and Partial-coherence Interferometry Optical Biometers [J].
Can, Ertugrul ;
Duran, Mustafa ;
Cetinkaya, Tugba ;
Ariturk, Nursen .
MIDDLE EAST AFRICAN JOURNAL OF OPHTHALMOLOGY, 2016, 23 (04) :288-292
[9]   Scheimpflug-Placido topographer and optical low-coherence reflectometry biometer: Repeatability and agreement [J].
Chen, Wuhe ;
McAlinden, Colm ;
Pesudovs, Konrad ;
Wang, Qinmei ;
Lu, Fan ;
Feng, Yifan ;
Chen, Jie ;
Huang, Jinhai .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (09) :1626-1632
[10]   Comparison of the Orbscan II topographer and the iTrace aberrometer for the measurements of keratometry and corneal diameter in myopic patients [J].
Chen, Yao ;
Xia, Xiaobo .
BMC OPHTHALMOLOGY, 2016, 16