Slit slat collimator optimization with respect to MTF

被引:5
作者
Kamali-Asl, A [1 ]
Sarkar, S
Shahriari, M
Agha-Hosseini, H
机构
[1] Amir Kabir Univ, Dept Phys, Tehran, Iran
[2] Univ Tehran Med Sci, Dept Phys Med, Tehran, Iran
[3] Univ Tehran Med Sci, RCSTIM, Tehran, Iran
[4] Shahid Beheshtee Univ, Dept Sci, Tehran, Iran
关键词
slit slat collimator; optimization; SPECT metabolic imaging; Monte Carlo method;
D O I
10.1016/j.apradiso.2004.07.009
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Conventional single photon emission computed tomography (SPECT) collimators are not suitable for 511 keV imaging. In order to use a SPECT system for single positron emission tomography (SPET),,we need to design high-efficiency collimators capable of absorbing 511 keV photons. Slit slat (SS) collimators have been proposed for this purpose even though the optimization of such collimators has not yet been performed. In order to investigate the reliability of the system with SS collimator, it was decided to evaluate such a system. A gamma camera system with both parallel hole (PH) and SS collimators was simulated by Monte Carlo method. In these simulation parameters of the simulated system having (3)/(8) in NaI(Tl) crystal for 140 keV gamma-rays with PH collimator was compared with practical results and a good correlation was found between the simulated and practical results. In this Way, the validity of our simulation code was confirmed. The crystal thickness for simulated gamma-camera system with SS collimator for detection of 511 keV photons was set to be (5)/(8) in which resulted in an intrinsic spatial resolution of 4mm. The simulated SS collimators consisted of 50 lead plates of various height, thickness and spacing. The imaging was performed by rotating the SS collimator from 0degrees to 180degrees in 7 steps of 2.5degrees each. The physical parameters such as spatial resolution, efficiency and modulation transfer function (MTF) of this system for different SS collimator designs, (i.e. SS64,3,3, SS80,3,3, SS64,4,5, SS80,4,5, SS85,4,5, where in SSh,s,t is used to indicate the septal eight (h), septal separation (s), and septal thickness (t) in millimeters were evaluated. The value of these parameters were compared with the values obtained employing a 511 keV PH85,4,2.5 collimator. Based on MTF values,our result showed that the SS85,4,5 collimator has an optimum performance to the other SS collimators. The relative efficiency of the system with SS85,4,5 collimator was about 20 times higher than the efficiency of the system with the PH85,4,2.5 collimator whereas in this case the spatial resolution had deteriorated by almost 8% i.e. from 14.1 to 15.0 mm). The result of this study shows that the SS collimators have great advantages over the conventional PH collimators for 511 keV imaging, providing much higher efficiency While maintaining a comparable spatial resolution. Employing this collimator makes it possible to use a SPECT system for metabolic imaging by SPET system. (C) 2004 Elsevier Ltd. All rights reserved.
引用
收藏
页码:461 / 468
页数:8
相关论文
共 15 条
[1]  
*ADAC, 1999, ADAC LAB 540 ALD DRI
[2]  
Briesmeister J.F., 1986, MCNP GEN MONTE CARLO
[3]   Performance assessment of a slat gamma camera collimator for 511 keV imaging [J].
Britten, AJ ;
Klie, R .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (07) :1735-1741
[4]  
GUNTER DL, 1998, SCI BASIC NUCL MED C
[5]  
HOSSEINIASHRAFI ME, 1992, APPL RADIAT ISOTOPES, V43, P1449, DOI 10.1016/0883-2889(92)90171-A
[6]  
HOSSEINIASHRAFI ME, 1998, APPL RAD, V49, P934
[7]   THE EXPERIMENTAL EVALUATION OF A PROTOTYPE ROTATING SLAT COLLIMATOR FOR PLANAR GAMMA-CAMERA IMAGING [J].
LODGE, MA ;
BINNIE, DM ;
FLOWER, MA ;
WEBB, S .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (03) :427-448
[8]   A prototype rotating slat collimator for single photon emission computed tomography [J].
Lodge, MA ;
Webb, S ;
Flower, MA ;
Binnie, DM .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1996, 15 (04) :500-511
[9]  
MOORE SC, 1992, EUR J NUCL MED, V19, P138, DOI 10.1007/BF00184130
[10]  
NICOLAOU GE, 1986, APPL RADIAT ISOTOPES, V37, P1219, DOI 10.1016/0883-2889(86)90010-9