Effect of tracheal tube cuff shape on fluid leakage across the cuff: an in vitro study

被引:74
作者
Dave, M. H. [1 ]
Frotzler, A. [1 ]
Spielmann, N. [1 ]
Madjdpour, C. [1 ]
Weiss, M. [1 ]
机构
[1] Univ Childrens Hosp, Dept Anaesthesia, CH-8032 Zurich, Switzerland
关键词
airway; tracheal tube; VENTILATOR-ASSOCIATED PNEUMONIA; LARYNGEAL SEAL DESIGN; ENDOTRACHEAL-TUBE; PREVENT LEAKAGE; HIGH-VOLUME; PRESSURE; ASPIRATION; DIAGNOSIS; MODEL;
D O I
10.1093/bja/aeq202
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
This study compared the fluid leakage in the new 'tapered' shaped against the classic 'cylindrical' shaped tracheal tube cuffs when placed in different sized tracheas. The 7.5 mm internal diameter (ID) tracheal tube cuffs-Tapered Seal Guard (TSG), Standard Seal Guard (SSG), Hi-Lo, Microcuff, Ruesch, and Portex Profile-were compared in an in vitro apparatus. Vertical artificial tracheas with 16, 20, and 22 mm ID were intubated, 5 ml clear water was applied above the unlubricated tube cuffs, and fluid leakage was measured up to 60 min. Data of tapered vs non-tapered tube cuffs (16 observations) were compared for each tracheal diameter using the Mann-Whitney test. Median (range) fluid leakage (ml) at 60 min was 2.14 (0.05-4.88), 1.14 (0.00-4.84), and 0.13 (0.00-1.32), respectively, for 16, 20, and 22 mm tracheas in the TSG tube studies when compared with 4.58 (0.44-4.88), 2.21 (0.00-4.81), and 0.00 (0.00-4.81) in the SSG tube and 4.54 (1.54-4.82), 0.90 (0.00-4.49), and 4.85 (4.40-4.99) in the Microcuff tube studies. Leakage in all polyvinylchloride (PVC) tube cuffs was almost complete (5 ml) within 5 min (P < 0.001). The tapered PU tube cuff was as effective as the cylindrical PU cuffs in smaller tracheal diameters and was more efficient than the cylindrical Microcuff PU tube cuff in larger tracheal diameter in preventing subglottic fluid leakage across the tube cuff tested in this in vitro study. PVC tube cuffs leaked much more and faster than PU cuffs.
引用
收藏
页码:538 / 543
页数:6
相关论文
共 26 条
[1]   RATIONALIZATION OF THE SELECTION OF TRACHEAL TUBES [J].
CHANDLER, M ;
CRAWLEY, BE .
BRITISH JOURNAL OF ANAESTHESIA, 1986, 58 (01) :111-116
[2]   Epidemiology, etiology, and diagnosis of hospital-acquired pneumonia and ventilator-associated pneumonia in Asian countries [J].
Chawla, Rajesh .
AMERICAN JOURNAL OF INFECTION CONTROL, 2008, 36 (04) :S93-S100
[3]   Tracheal fluid leakage in benchtop trials: comparison of static versus dynamic ventilation model with and without lubrication [J].
Dave, Mital H. ;
Koepfer, Nelly ;
Madjdpour, Caveh ;
Frotzler, Angela ;
Weiss, Markus .
JOURNAL OF ANESTHESIA, 2010, 24 (02) :247-252
[4]   Fit and seal characteristics of a new paediatric tracheal tube with high volume-low pressure polyurethane cuff [J].
Dullenkopf, A ;
Gerber, AC ;
Weiss, M .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2005, 49 (02) :232-237
[5]   Air leakage around endotracheal tube cuffs [J].
Dullenkopf, A ;
Schmitz, A ;
Frei, M ;
Gerber, AC ;
Weiss, M .
EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2004, 21 (06) :448-453
[6]   Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube [J].
Dullenkopf, A ;
Gerber, A ;
Weiss, M .
INTENSIVE CARE MEDICINE, 2003, 29 (10) :1849-1853
[7]   NOSOCOMIAL PNEUMONIA IN VENTILATED PATIENTS - A COHORT STUDY EVALUATING ATTRIBUTABLE MORTALITY AND HOSPITAL STAY [J].
FAGON, JY ;
CHASTRE, J ;
HANCE, AJ ;
MONTRAVERS, P ;
NOVARA, A ;
GIBERT, C .
AMERICAN JOURNAL OF MEDICINE, 1993, 94 (03) :281-288
[8]  
Ho AMH, 2002, ANAESTHESIA, V57, P173
[9]   In Vivo and In Vitro Morphometry of the Human Trachea [J].
Kamel, Kirollos Salah ;
Lau, Gabriel ;
Stringer, Mark D. .
CLINICAL ANATOMY, 2009, 22 (05) :571-579
[10]   DESIGN AND DEVELOPMENT OF ULTRATHIN-WALLED, NONKINKING ENDOTRACHEAL-TUBES OF A NEW NO-PRESSURE LARYNGEAL SEAL DESIGN - A PRELIMINARY-REPORT [J].
KOLOBOW, T ;
TSUNO, K ;
ROSSI, N ;
APRIGLIANO, M .
ANESTHESIOLOGY, 1994, 81 (04) :1061-1067