Anterior vertebral body screw pullout testing - A comparison of Zielke, Kaneda, universal spine system, and Universal Spine System with pullout-resistant nut

被引:28
作者
Lieberman, IH
Khazim, R
Woodside, T
机构
[1] Cleveland Clin Fdn, Sect Spinal Surg A41, Div Orthopaed, Cleveland, OH 44195 USA
[2] Toronto Hosp, Orthopaed Biomech Lab, Dept Surg, Toronto, ON M5T 2S8, Canada
关键词
anterior vertebral body screws; biomechanical testing; spine;
D O I
10.1097/00007632-199804150-00012
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. A biomechanical study of pullout of anteriorly implanted screws in cadaveric vertebral bodies. Objectives. To investigate and compare the pullout strength of the Zielke, Kaneda, Universal Spine System (USS) pedicle screw, and USS pedicle screw with a new pullout-resistant nut. Summary of Background Data. A common problem with anterior purchase regardless of the implant system is screw pullout at the proximal and distal ends of multilevel constructs. There is limited information on a solution to this problem. Methods. The L1 to L4 vertebral bodies from four cadavers had one each of Zielke and Kaneda pedicle screws (Acromed Corp, Cleveland, OH), USS pedicle screw (Synthes Spine, Paoli, PA), and USS pedicle screw with pullout-resistant nut implanted transversely across the center of the vertebral body with bicortical purchase in a similar fashion as would be used clinically. The screws were extracted using a servohydraulic material testing system. The maximum axial forces were recorded. Results. The Zielke and Kaneda screws had no significant difference in mean pullout strength (P = 0.542). The USS screw alone was less strong (P = 0.009). The USS screw and pullout-resistant nut increased the pullout strength by twofold (P = 0.00006). In the screw pullout tests, the mode of failure was at the screw thread's interface, The USS screw and pullout-resistant nut failed by imploding the body around the nut. With the USS screw and pullout-resistant nut, the pullout strength was determined by the compressive strength of the bone. Conclusions. The addition of a pullout-resistant nut to an anterior vertebral body screw improves the pullout strength by twofold and changes the mode of failure to rely ultimately on the inherent vertebral body strength rather than the screw's characteristics. The addition of a pullout-resistant nut may be applicable to multilevel implant constructs to prevent screw pullout at the top and bottom.
引用
收藏
页码:908 / 910
页数:3
相关论文
共 6 条
[1]  
DeCoster T A, 1990, J Orthop Trauma, V4, P169, DOI 10.1097/00005131-199004020-00012
[2]   Strength of fixation of anterior vertebral body screws [J].
Horton, WC ;
Blackstock, SF ;
Norman, JT ;
Hill, CS ;
Feiertag, MA ;
Hutton, WC .
SPINE, 1996, 21 (04) :439-444
[3]   MECHANICAL PROPERTIES OF SURGICAL BONE SCREWS AND SOME ASPECTS OF INSERTION PRACTICE [J].
HUGHES, AN ;
JORDAN, BA .
INJURY-THE BRITISH JOURNAL OF ACCIDENT SURGERY, 1972, 4 (01) :25-&
[4]   BURST FRACTURES WITH NEUROLOGIC DEFICITS OF THE THORACOLUMBAR-LUMBAR SPINE - RESULTS OF ANTERIOR DECOMPRESSION AND STABILIZATION WITH ANTERIOR INSTRUMENTATION [J].
KANEDA, K ;
ABUMI, K ;
FUJIYA, M .
SPINE, 1984, 9 (08) :788-795
[5]  
*MATH SOFT INC, 1995, STAT SCI S PLUS GUID
[6]   VENTRAL DEROTATIONS-SPONDYLODESIS - RESULTS WITH 26 SURGICALLY TREATED CASES [J].
ZIELKE, K ;
STUNKAT, R ;
BEAUJEAN, F .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1976, 85 (03) :257-277