Comparison between X-ray fluorescence and inductively coupled plasma atomic emission spectrometry in the analysis of sediment samples

被引:15
|
作者
Somogyi, A
Braun, M
Posta, J
机构
[1] Lajos Kossuth Univ, Isotope Lab, H-4010 Debrecen, Hungary
[2] Lajos Kossuth Univ, Dept Inorgan & Analyt Chem, H-4010 Debrecen, Hungary
关键词
inductively coupled plasma emission spectrometry; sample preparation; X-ray fluorescence;
D O I
10.1016/S0584-8547(97)00086-4
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Two multielemental analytical techniques, X-ray fluorescence analysis (XRF) and inductively coupled plasma atomic emission spectrometry (ICP-AES) were used for the analysis of the elemental composition of sediment samples from a marsh and standard reference materials. The sediment samples were pretreated with different methods which are widely used in practice. A comparison was made not only between the concentrations obtained by the different methods, but also between the statistical conclusions derived from the processing of the experimental results. Good agreement was found for some elements, e.g. Mn, Zn and Sr, while the concentrations and the statistical conclusions were shown to depend on the analytical method used in the case of other elements, e.g. Fe and Zr. (C) 1997 Elsevier Science B.V.