Extra-articular migration of PEEK interference screw after anterior cruciate ligament reconstruction: a report of two cases

被引:12
|
作者
Fang, Chao-Hua [1 ,2 ]
Li, Ming [1 ]
Zhang, Yun-Feng [1 ]
Liu, Hua [1 ]
机构
[1] Ningbo 6 Hosp, Dept Joint Surg, 1059 Zhongshan Rd, Ningbo 315000, Zhejiang, Peoples R China
[2] Shanghai Jiao Tong Univ, Sch Biomed Engn, 1954 Huashan Rd, Shanghai 200240, Peoples R China
关键词
Polyetheretherketone; PEEK; Anterior cruciate ligament; Interference screw; Migration; Case report; INTRAARTICULAR MIGRATION; WEAR PARTICLES; POLYETHERETHERKETONE; REMOVAL; IMPLANT;
D O I
10.1186/s12891-021-04387-2
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background The interference screw is the most popular device that fixes the graft for anterior cruciate ligament reconstruction, reducing the incidence of windshield effect and bungee effect. For the screw, either metallic, "bioresorbable," or polyetheretherketone (PEEK) material is available. PEEK is popular and extensively used due to its stability, biocompatibility, radiolucency, and elastic modulus. Rare relevant complications were reported, but here, we report two cases of extra-articular migrations of PEEK interference screw from the tibial tunnel after anterior cruciate reconstruction. Case report An 18-year-old boy and a 56-year-old woman underwent anterior cruciate ligament reconstruction using a PEEK interference screw to fix the graft in the tibial tunnel. They suffered from screw extrusion from the tibial tunnel after 40 days and six months, respectively, with an incision rupture or palpable subcutaneous mass. They underwent a second operation and recovered well. Conclusions The exact incidence of extra-articular migrations of PEEK interference screws is unknown, but it seems to be quite low; despite this and its uncertain cause, the negative effects caused by the PEEK material need to be considered.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Extra-articular migration of PEEK interference screw after anterior cruciate ligament reconstruction: a report of two cases
    Chao-Hua Fang
    Ming Li
    Yun-Feng Zhang
    Hua Liu
    BMC Musculoskeletal Disorders, 22
  • [2] EXTRA-ARTICULAR RECONSTRUCTION ASSOCIATED WITH THE ANTERIOR CRUCIATE LIGAMENT IN BRAZIL
    Helito, Camilo Partezani
    do Amaral Junior, Carlos
    Camargo, Danilo Bordini
    Bonadio, Marcelo Batista
    Pecora, Jose Ricardo
    Demange, Marco Kawamura
    ACTA ORTOPEDICA BRASILEIRA, 2019, 27 (04): : 202 - 206
  • [3] Extra-Articular Tenodesis in Combination with Anterior Cruciate Ligament Reconstruction: An Overview
    Cerciello, Simone
    Batailler, Cecile
    Darwich, Nader
    Neyret, Philippe
    CLINICS IN SPORTS MEDICINE, 2018, 37 (01) : 87 - +
  • [4] Intra-articular pull out of an interference screw after anterior cruciate ligament reconstruction
    Moonot, P.
    Allen, P.
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2006, 14 (09) : 862 - 864
  • [5] Intra-articular pull out of an interference screw after anterior cruciate ligament reconstruction
    Moonot, P.
    Allen, P.
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2006, 14 (10) : 1004 - 1006
  • [6] Intra-articular pull out of an interference screw after anterior cruciate ligament reconstruction
    P. Moonot
    P. Allen
    Knee Surgery, Sports Traumatology, Arthroscopy, 2006, 14 : 862 - 864
  • [7] Intra-articular pull out of an interference screw after anterior cruciate ligament reconstruction
    P. Moonot
    P. Allen
    Knee Surgery, Sports Traumatology, Arthroscopy, 2006, 14 : 1004 - 1006
  • [8] Intra- and extra-articular hamstring reconstruction of anterior cruciate ligament tears
    Buscayret, C
    Buscayret, F
    Farenq, C
    REVUE DE CHIRURGIE ORTHOPEDIQUE ET REPARATRICE DE L APPAREIL MOTEUR, 2001, 87 (03): : 276 - 280
  • [9] Transcutaneous migration of a tibial bioabsorbable interference screw after anterior cruciate ligament reconstruction
    Sassmannshausen, G
    Caff, CF
    ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2003, 19 (09)
  • [10] Intraarticular migration of a broken biodegradable interference screw after anterior cruciate ligament reconstruction
    Baums, M. H.
    Zelle, B. A.
    Schultz, W.
    Ernstberger, T.
    Klinger, H. -M.
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2006, 14 (09) : 865 - 868