Characterization of pulsed ultrasound using optical detection in Raman-Nath regime

被引:6
作者
Jia, Lecheng [1 ]
Xue, Bin [2 ]
Chen, Shili [1 ]
Wu, Hanzhong [2 ]
Yang, Xiaoxia [3 ]
Zhai, Jingsheng [2 ]
Zeng, Zhoumo [1 ]
机构
[1] Tianjin Univ, State Key Lab Precis Measurement Technol & Instru, Tianjin 300072, Peoples R China
[2] Tianjin Univ, Sch Marine Sci & Technol, Tianjin 300072, Peoples R China
[3] Tianjin Univ Technol & Educ, Tianjin Key Lab Informat Sensing & Intelligent Co, Tianjin 300222, Peoples R China
基金
中国国家自然科学基金;
关键词
LIGHT; DIFFRACTION; HYDROPHONES; SENSITIVITY; CALIBRATION;
D O I
10.1063/1.5024882
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In this paper, we describe an optical detection method for the characterization of pulsed ultrasound based on acousto-optic interaction. We deduce the relationship between the ultrasound and the diffracted light from the principle of acousto-optic diffraction in the Raman-Nath regime, which is verified experimentally. Five ultrasonic transducers with different central frequencies and different focusing types are measured to show the method's performance regarding linearity, sound pressure measurement, phase measurement, frequency response, and spatial resolution. The experimental results show a good agreement with simulation data by CIVA (ultrasonic simulation software, M2M NDT, Inc.) and the pulse-echo method. Published by AIP Publishing.
引用
收藏
页数:8
相关论文
共 27 条
[1]  
Abeele K. V. D., 1990, J ACOUST SOC AM, V88, P2298
[2]   ACOUSTOOPTICAL RECONSTRUCTION OF A PULSED ULTRASONIC SIGNAL [J].
BLOMME, E ;
LEROY, O .
OPTICAL ENGINEERING, 1992, 31 (10) :2122-2128
[3]   Application of the acousto-optic effect to pressure measurements in ultrasound fields in water using a laser vibrometer [J].
Buick, JM ;
Cosgrove, JA ;
Douissard, PA ;
Greated, CA ;
Gilabert, B .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (10) :3203-3207
[4]   CIVA: An expertise platform for simulation and processing NDT data [J].
Calmon, P. ;
Mahaut, S. ;
Chatillon, S. ;
Raillon, R. .
ULTRASONICS, 2006, 44 :E975-E979
[5]   ULTRASONIC MEASUREMENT OF FETAL ABDOMEN CIRCUMFERENCE IN ESTIMATION OF FETAL WEIGHT [J].
CAMPBELL, S ;
WILKIN, D .
BRITISH JOURNAL OF OBSTETRICS AND GYNAECOLOGY, 1975, 82 (09) :689-697
[6]  
Gan W.S., 2012, ACOUSTICAL IMAGING T
[7]   Sensitivity measurements of piezoelectric polymer hydrophones from 0.2-2 MHz using a broadband-pulse technique [J].
Harris, GR ;
Gammell, PM .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1999, 105 (02) :725-731
[8]  
He L., 2016, REV SCI INSTRUM, V87, P231
[9]   DIFFRACTION OF LIGHT BY A FOCUSED ULTRASONIC WAVE [J].
HUANG, J ;
NISSEN, JA ;
BODEGOM, E .
JOURNAL OF APPLIED PHYSICS, 1992, 71 (01) :70-75
[10]  
[靳世久 Jin Shijiu], 2014, [电子测量与仪器学报, Journal of Electronic Measurement and Instrument], V28, P925