Development of mini-spectrophotometer for determination of plasma glucose

被引:23
作者
Chaianantakul, Natpasit [1 ]
Wutthi, Kanchaporn [1 ]
Kamput, Nattanit [1 ]
Pramanpol, Nuttawan [2 ]
Janphuang, Pattanaphong [2 ]
Pummara, Watcharapon [2 ]
Phimon, Kantapon [2 ]
Phatthanakun, Rungrueang [2 ]
机构
[1] Naresuan Univ, Fac Allied Hlth Sci, Dept Med Technol, Phitsanulok 65000, Thailand
[2] Synchrotron Light Res Inst Publ Org, Nakhon Ratchasima 30000, Thailand
关键词
Mini-spectrophotometer; Light emitting diode; Fabricated microwell; Glucose determination; LINEAR-REGRESSION; OXIDASE METHOD; POLY(DIMETHYLSILOXANE); CAPILLARY; SYSTEM; MEDIA; BLOOD; ASSAY;
D O I
10.1016/j.saa.2018.06.107
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
This work demonstrates a novel compact spectrophotometer, "Mini-spectrophotometer", designed for plasma glucose detection. Unlike conventional spectrophotometer, a light source of the mini spectrophotometer is replaced by a light-emitting diode (LED), and a fabricated polymer-based microwell is used as a cuvette. To validate the downsizing spectrophotometer prototype, the efficiency and reliability for glucose determination are investigated. Using a certain light intensified from LED, the within-run precision of mini-spectrophotometer is found to be 3.9-8.4% while the between-run precision is 6.7-10.8%. The linearity for the quantification of glucose was up to 500 mg dL(-1) and the recovery 99.1 +/- 3.4% is obtained. The sensitive and selective detection of glucose has been observed; with limit of detection (LOD) of 13.5 mg dL(-1) and limit of quantification (LOQ) of 46.2 mg dL(-1), respectively. Hemoglobin and triglyceride at high concentration slightly interferes with the proposed instrument. From comparative studies, there are no significant differences between the glucose concentration measured by mini-spectrophotometer and Shimadzu (r(2) = 0.9862) or CECIL spectrophotometer (r(2) = 0.9853). Using Passing-Bablok regression analysis, the results obtained from mini-spectrophotometer are in close agreement with the two conventional spectrophotometers. Furthermore, using microwell, the sample volume and reagent used in the process can be reduced. The in-house developed mini-spectrophotometer is capable of detecting plasma glucose while maintaining a compact system, demonstrating the potential of high performance, cost-effective, and portable spectrophotometer for clinical chemistry analysis in small routine, research, and teaching medical laboratory technologist. (C) 2018 Elsevier B.V. All rights reserved.
引用
收藏
页码:670 / 676
页数:7
相关论文
共 41 条
[1]  
Amer Diabet Assoc, 2013, DIABETES CARE, V36, pS67, DOI [10.2337/dc13-S067, 10.2337/dc14-S081, 10.2337/dc11-S062, 10.2337/dc10-S011, 10.2337/dc13-S011, 10.2337/dc11-S011, 10.2337/dc10-S062, 10.2337/dc12-s011, 10.2337/dc12-s064]
[2]  
American Diabetes Association, 2007, Diabetes Care, V30 Suppl 1, pS42
[3]   Glucose oxidase - An overview [J].
Bankar, Sandip B. ;
Bule, Mahesh V. ;
Singhal, Rekha S. ;
Ananthanarayan, Laxmi .
BIOTECHNOLOGY ADVANCES, 2009, 27 (04) :489-501
[4]   IMPROVED COLOR REAGENT FOR DETERMINATION OF BLOOD-GLUCOSE BY OXIDASE SYSTEM [J].
BARHAM, D ;
TRINDER, P .
ANALYST, 1972, 97 (1151) :142-&
[5]  
Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
[6]   Sensitive fluorescent probes for determination of hydrogen peroxide and glucose based on enzyme-immobilized magnetite/silica nanoparticles [J].
Chang, Qing ;
Zhu, Lihua ;
Jiang, Guodong ;
Tang, Heqing .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2009, 395 (07) :2377-2385
[7]   Creatinine determination in urine samples by batchwise kinetic procedure and flow injection analysis using the Jaffe reaction:: chemometric study [J].
Falcó, PC ;
Genaro, LAT ;
Lloret, SM ;
Gomez, FB ;
Cabeza, AS ;
Legua, CM .
TALANTA, 2001, 55 (06) :1079-1089
[8]   QUANTITATIVE-DETERMINATION OF THE ABSORPTION-SPECTRA OF CHROMOPHORES IN STRONGLY SCATTERING MEDIA - A LIGHT-EMITTING-DIODE BASED TECHNIQUE [J].
FANTINI, S ;
FRANCESCHINI, MA ;
FISHKIN, JB ;
BARBIERI, B ;
GRATTON, E .
APPLIED OPTICS, 1994, 33 (22) :5204-5213
[9]   A multichannel photometer based on an array of light emitting diodes for use in multivariate calibration [J].
Fonseca, A ;
Raimundo, IM .
ANALYTICA CHIMICA ACTA, 2004, 522 (02) :223-229
[10]  
Goldstein D. E., DIABETES CARE S1, V27