Biomechanical Analysis of Retrograde Superior Ramus Screw Fixation Constructs

被引:8
作者
Lucas, Justin F. [1 ]
Routt, Milton L. Chip, Jr. [2 ]
Eastman, Jonathan G. [3 ]
机构
[1] Santa Clara Valley Med Ctr, Dept Orthopaed Surg, 751 South Bascom Ave, San Jose, CA 95128 USA
[2] Univ Texas Houston, Dept Orthopaed Surg, Hlth Sci Ctr, Houston, TX USA
[3] Univ Calif Davis, Dept Orthoped Surg, Sacramento, CA USA
基金
美国国家卫生研究院;
关键词
superior ramus screw; ramus fracture; retrograde screw; pelvic fracture; percutaneous screw; ANTERIOR PELVIC RING; ILIOSACRAL SCREWS; FRACTURES;
D O I
10.1097/BOT.0000000000001951
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: To investigate the biomechanical performance of different size and length retrograde superior ramus screws. Materials and Methods: A vertical superior ramus fracture was created in osteoporotic composite hemipelvis bone models (Sawbones, Vashon Island, WA). After reduction, 4 fixation groups were created by inserting either a solid 4.5 mm (Depuy Synthes, Paoli, PA) or cannulated 7.0 mm screw (Zimmer, Warsaw, IN) of either 80 mm (short) or 140 mm (long) in length. An intact and an unstabilized osteotomy group were also created. Samples underwent cyclic loading for 5000 cycles with data acquisition at regular intervals. At the end of cyclic loading, load to failure was performed. Results: The displacement after 5000 cycles for 4.5 mm short screws was significantly greater than 4.5 mm long and 7.0 mm short screws. At 5000 cycles, the 4.5-mm short screws had a significantly lower stiffness and lower load to failure than all other screws and were not different from the osteotomy model. Conclusions: Short 4.5-mm screws demonstrated increased displacement, lower stiffness, and decreased load to failure compared with all other screws. The biomechanical performance of 4.5-mm short screws was no different than unstabilized controls. Longer bicortical screw fixation is suggested when possible. Additional biomechanical and clinical studies are needed to fully understand the significance of these findings.
引用
收藏
页码:187 / 191
页数:5
相关论文
共 25 条
[1]   Are two retrograde 3.5 mm screws superior to one 7.3 mm screw for anterior pelvic ring fixation in bones with low bone mineral density? [J].
Acklin, Y. P. ;
Zderic, I. ;
Grechenig, S. ;
Richards, R. G. ;
Schmitz, P. ;
Gueorguiev, B. .
BONE & JOINT RESEARCH, 2017, 6 (01) :8-13
[2]   Biomechanical comparison of plate and screw fixation in anterior pelvic ring fractures with low bone mineral density [J].
Acklin, Yves P. ;
Zderic, Ivan ;
Buschbaum, Jan ;
Varga, Peter ;
Inzana, Jason A. ;
Grechenig, Stephan ;
Richards, R. Geoff ;
Gueorguiev, Boyko ;
Schmitz, Paul .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2016, 47 (07) :1456-1460
[3]   Hip contact forces and gait patterns from routine activities [J].
Bergmann, G ;
Deuretzbacher, G ;
Heller, M ;
Graichen, F ;
Rohlmann, A ;
Strauss, J ;
Duda, GN .
JOURNAL OF BIOMECHANICS, 2001, 34 (07) :859-871
[4]   Differences of percutaneous retrograde screw fixation of anterior column acetabular fractures between male and female: A study of 164 virtual three-dimensional models [J].
Chen, Kai Ning ;
Wang, Gang ;
Cao, Liang Guo ;
Zhang, Mei Chao .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2009, 40 (10) :1067-1072
[5]   Minimally-invasive fixation for anterior pelvic ring disruptions [J].
Cole, Peter A. ;
Dyskin, Evgeny A. ;
Gilbertson, Jeffrey A. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2015, 46 :S27-S34
[6]   Anterior Pelvic External Fixator Versus Subcutaneous Internal Fixator in the Treatment of Anterior Ring Pelvic Fractures [J].
Cole, Peter A. ;
Gauger, Erich M. ;
Anavian, Jack ;
Ly, Thuan V. ;
Morgan, Robert A. ;
Heddings, Archie A. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2012, 26 (05) :269-277
[7]   Intramedullary Fixation Techniques for the Anterior Pelvic Ring [J].
Eastman, Jonathan G. ;
Routt, Milton L. Chip, Jr. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2018, 32 :S4-S13
[8]   Die bildwandergestützte, perkutane transiliosakrale Schraubenfixation instabiler BeckenverletzungenPercutaneous Iliosacral Screw Fixation of Unstable Pelvic Injuries by Conventional Fluoroscopy [J].
Axel Gänsslen ;
Tobias Hüfner ;
Christian Krettek .
Operative Orthopädie und Traumatologie, 2006, 18 (3) :225-244
[9]   Retrograde transpubic screw fixation of transpubic instabilities [J].
Gänsslen A. ;
Krettek C. .
Operative Orthopädie und Traumatologie, 2006, 18 (4) :330-340
[10]   Anterior Pelvic Reduction and Fixation Using a Subcutaneous Internal Fixator [J].
Gardner, Michael J. ;
Mehta, Samir ;
Mirza, Amer ;
Ricci, William M. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2012, 26 (05) :314-321