Disposable biosensors for determination of biogenic amines

被引:112
作者
Asuncion Alonso-Lomillo, M. [1 ]
Dominguez-Renedo, Olga [1 ]
Matos, Patricia [2 ]
Julia Arcos-Martinez, M. [1 ]
机构
[1] Univ Burgos, Fac Sci, Dept Analyt Chem, Burgos 09001, Spain
[2] Univ Beira Interior, Fac Sci, Covilha, Portugal
关键词
Biogenic amines; Screen-printed electrodes; Diazonium salt; Biosensor; AMPEROMETRIC BIOSENSOR; HORSERADISH-PEROXIDASE; PUTRESCINE OXIDASE; HISTAMINE; ELECTRODES;
D O I
10.1016/j.aca.2010.03.012
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This work reports monoamine oxidase (MAO)/horseradish peroxidase (HRP) and diamine oxidase (DAO)/horseradish peroxidase (HRP) based biosensors using screen-printed carbon electrodes for the determination of biogenic amines (BA). The enzymes have been covalently immobilized onto the carbon working electrode, previously modified by an aryl diazonium salt, using hydroxysuccinimide and carbodiimide. The detection has been performed by measuring the cathodic current due to the reduction of the mediator hydroxymethylferrocene at a low potential, 250 mV vs screen-printed Ag/AgCl reference electrode. The experimental conditions for the enzymes immobilization, as well as for the main variables that can influence the chronoamperometric current have been optimized by the experimental design methodology. Under these optimum conditions, the disposable biosensors have been characterized. A linear response range from 0.2 up to 1.6 mu M and from 0.4 to 2.4 mu M of histamine was obtained for DAO/HRP and MAO/HRP based biosensors, respectively. The biosensor construction was highly reproducible, yielding relative standard deviations of 10% and 11% in terms of sensitivity for DAO/HRP and MAO/HRP based biosensors, respectively. The capability of detection, 0.18 +/- 0.01 mu M in the case of DAO/HRP and 0.40 +/- 0.04 mu M (alpha = 0.05 and beta = 0.005) for MAO/HRP based biosensors, and the biosensor sensitivity towards different BA has also been analyzed. Finally, the developed biosensors have been applied to the determination of the total amine content in fish samples. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:26 / 31
页数:6
相关论文
共 35 条
[1]   CYP450 2B4 covalently attached to carbon and gold screen printed electrodes by diazonium salt and thiols monolayers [J].
Alonso-Lomillo, M. A. ;
Yardimci, C. ;
Dominguez-Renedo, O. ;
Arcos-Martinez, M. J. .
ANALYTICA CHIMICA ACTA, 2009, 633 (01) :51-56
[2]   Hydrogenase-coated carbon nanotubes for efficient H2 oxidation [J].
Alonso-Lomillo, M. Asuncion ;
Rudiger, Olaf ;
Maroto-Valiente, Angel ;
Velez, Marisela ;
Rodriguez-Ramos, Inmaculada ;
Munoz, F. Javier ;
Fernandez, Victor M. ;
De Lacey, Antonio L. .
NANO LETTERS, 2007, 7 (06) :1603-1608
[3]  
ANDREW TF, 2002, J POLYM SCI A, V40, P719
[4]   Horseradish peroxidase covalent grafting onto screen-printed carbon electrodes for levetiracetam chronoamperometric determination [J].
Asuncion Alonso-Lomillo, M. ;
Dominguez-Renedo, Olga ;
Hernandez-Martin, Alba ;
Julia Arcos-Martinez, M. .
ANALYTICAL BIOCHEMISTRY, 2009, 395 (01) :86-90
[5]   Amperometric biosensor for diamine using diamine oxidase purified from porcine kidney [J].
Bouvrette, P ;
Male, KB ;
Luong, JHT ;
Gibbs, BF .
ENZYME AND MICROBIAL TECHNOLOGY, 1997, 20 (01) :32-38
[6]  
Box G. E. P., 1999, ESTADISTICA INVESTIG
[7]   An interference free amperometric biosensor for the detection of biogenic amines in food products [J].
Carelli, Donatella ;
Centonze, Diego ;
Palermo, Carmen ;
Quinto, Maurizio ;
Rotunno, Taddeo .
BIOSENSORS & BIOELECTRONICS, 2007, 23 (05) :640-647
[8]   Amperometric biosensor based on horseradish peroxidase for biogenic amine determinations in biological samples [J].
Castilho, TJ ;
Sotomayor, MDT ;
Kubota, LT .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2005, 37 (04) :785-791
[9]   Bienzyme biosensors for glucose, ethanol and putrescine built on oxidase and sweet potato peroxidase [J].
Castillo, J ;
Gáspár, S ;
Sakharov, I ;
Csöregi, E .
BIOSENSORS & BIOELECTRONICS, 2003, 18 (5-6) :705-714
[10]   Development of screen-printed enzyme electrodes for the estimation of fish quality [J].
Chemnitius, GC ;
Bilitewski, U .
SENSORS AND ACTUATORS B-CHEMICAL, 1996, 32 (02) :107-113