Analysis of the Mechanical Behavior and Effect of Cyclic Fatigue on the Implant-Abutment Interface

被引:0
作者
German Sandoval, Ramon [1 ]
Ortiz Magdaleno, Marine [1 ]
Sanchez Robles, Paula [1 ]
Zavala Alonso, Norma [2 ]
Romo Ramirez, Gabriel Fernando [1 ]
机构
[1] Autonomous Univ San Luis Potosi, Fac Stomatol, Aesthet Cosmet Restorat & Implantol Dent, San Luis Potosi, San Luis Potosi, Mexico
[2] Autonomous Univ San Luis Potosi, Fac Stomatol, Dept Dent Sci Adv Educ, San Luis Potosi, San Luis Potosi, Mexico
关键词
Abutment; Implant; Microgap; Cyclic fatigue; Compressive stress; FINITE-ELEMENT-ANALYSIS; STRESS-DISTRIBUTION; MARGINAL FIT; OSSEOINTEGRATED IMPLANTS; ZIRCONIA ABUTMENTS; VERTICAL MISFIT; MUCOSAL BARRIER; TITANIUM; MICROLEAKAGE; CONNECTIONS;
D O I
10.15517/IJDS.2021.45011
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: The seal of the interface formed at the implant-abutment connection is essential for the long-term success of the implant-supported restoration. The aim of this study was to analyze the mechanical behavior and the effect of cyclic fatigue before and after in the marginal fit of implant-abutment according to the manufacturing technique of the abutment. Materials and methods: Machined titanium abutments (DENTIS), cast abutments with Nickel-Chromium alloy (VeraBond II), and manufacturing custom milled Zirconia abutments (Zirkonzahn) were evaluated. The implant-abutment assemblies were subjected to cyclic loads of 133 N at a frequency of 19.1 Hz for 200,000 cycles. The microgap was measured using Scanning Electronic Microscope and the distribution of compressive stress by the three-dimensional Finite Element (FE) method. Results: The microgap measurement values of the machined abutments were 1.62 mu m and 1.92 mu m, cast abutments were 14.14 mu m, and 28.44 mu m, and the milled abutments were 14.18 mu m and 20.15 mu m before and after cyclic fatigue, respectively. Only the cast abutments and the machined abutments showed a statistically significant difference before and after cyclic fatigue (p <= 0.05). The FE analysis showed that the critical areas of compressive stress were located at the implant-abutment connection, increasing in the cast abutments and decreasing in the milled and the machined abutments. Conclusion: Cyclic fatigue exerts an effect on the dimensions of the microgap at the implant-abutment interface before and after loading; this microgap depends of the type of abutment material and the manufacturing technique.
引用
收藏
页码:104 / 114
页数:11
相关论文
共 40 条
[31]  
Ozen J, 2007, QUINTESSENCE INT, V38, pE355
[32]   BACTERIAL-COLONIZATION OF THE INTERNAL PART OF 2-STAGE IMPLANTS - AN IN-VIVO STUDY [J].
QUIRYNEN, M ;
VANSTEENBERGHE, D .
CLINICAL ORAL IMPLANTS RESEARCH, 1993, 4 (03) :158-161
[33]   In vitro synchrotron-based radiography of micro-gap formation at the implant-abutment interface of two-piece dental implants [J].
Rack, A. ;
Rack, T. ;
Stiller, M. ;
Riesemeier, H. ;
Zabler, S. ;
Nelson, K. .
JOURNAL OF SYNCHROTRON RADIATION, 2010, 17 :289-294
[34]   Mechanical Evaluation of an Implant-Abutment Self-Locking Taper Connection: Finite Element Analysis and Experimental Tests [J].
Sannino, Gianpaolo ;
Barlattani, Alberto .
INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2013, 28 (01) :E17-E26
[35]   Fatigue survival and failure resistance of titanium versus zirconia implant abutments with various connection designs [J].
Sen, Nazmiye ;
Us, Yesim Olcer .
JOURNAL OF PROSTHETIC DENTISTRY, 2019, 122 (03) :315.e1-315.e7
[36]   Sealing capability and marginal fit of titanium versus zirconia abutments with different connection designs [J].
Sen, Nazmiye ;
Sermet, Ibrahim Bulent ;
Gurler, Nezahat .
JOURNAL OF ADVANCED PROSTHODONTICS, 2019, 11 (02) :105-111
[37]   Experimental research on the relationship between fit accuracy and fracture resistance of zirconia abutments [J].
Sui, Xinxin ;
Wei, Huasha ;
Wang, Dashan ;
Han, Yan ;
Deng, Jing ;
Wang, Yongliang ;
Wang, Junjun ;
Yang, Jianjun .
JOURNAL OF DENTISTRY, 2014, 42 (10) :1353-1359
[38]   Marginal fit and microgaps of implant-abutment interface with internal anti-rotation configuration [J].
Tsuge, Takuma ;
Hagiwara, Yoshiyuki ;
Matsumura, Hideo .
DENTAL MATERIALS JOURNAL, 2008, 27 (01) :29-34
[39]   The mucosal barrier at implant abutments of different materials [J].
Welander, Maria ;
Abrahamsson, Ingemar ;
Berglundh, Tord .
CLINICAL ORAL IMPLANTS RESEARCH, 2008, 19 (07) :635-641
[40]   Fatigue induced deformation of taper connections in dental titanium implants [J].