A Novel Mock Circuit to Test Full-Flow Extracorporeal Membrane Oxygenation

被引:3
作者
Caspari, Stefan [1 ]
Schwaerzel, Leonie S. [1 ]
Jungmann, Anna M. [1 ]
Schmoll, Nicole [1 ]
Seiler, Frederik [1 ]
Muellenbach, Ralf M. [2 ]
Krawczyk, Marcin [3 ,4 ]
Quoc Thai Dinh [1 ]
Bals, Robert [1 ]
Lepper, Philipp M. [1 ]
Omlor, Albert J. [1 ]
机构
[1] Univ Hosp Saarland, Dept Internal Med Pneumol Allergol & Intens Care, D-66424 Homburg, Germany
[2] Univ Southampton, Dept Anaesthesiol & Crit Care, Campus Kassel, D-34125 Kassel, Germany
[3] Univ Hosp Saarland, Dept Internal Med 2, D-66424 Homburg, Germany
[4] Med Univ Warsaw, Ctr Preclin Res, Dept Gen Transplant & Liver Surg, Lab Metab Liver Dis, PL-02091 Warsaw, Poland
关键词
ECMO; COPD; ARDS; mock circulation; recirculation loop; ARDS;
D O I
10.3390/membranes12050493
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Extracorporeal membrane oxygenation (ECMO) has become an important therapeutic approach in the COVID-19 pandemic. The development and research in this field strongly relies on animal models; however, efforts are being made to find alternatives. In this work, we present a new mock circuit for ECMO that allows measurements of the oxygen transfer rate of a membrane lung at full ECMO blood flow. The mock utilizes a large reservoir of heparinized porcine blood to measure the oxygen transfer rate of the membrane lung in a single passage. The oxygen transfer rate is calculated from blood flow, hemoglobin value, venous saturation, and post-membrane arterial oxygen pressure. Before the next measuring sequence, the blood is regenerated to a venous condition with a sweep gas of nitrogen and carbon dioxide. The presented mock was applied to investigate the effect of a recirculation loop on the oxygen transfer rate of an ECMO setup. The recirculation loop caused a significant increase in post-membrane arterial oxygen pressure (paO(2)). The effect was strongest for the highest recirculation flow. This was attributed to a smaller boundary layer on gas fibers due to the increased blood velocity. However, the increase in paO(2) did not translate to significant increases in the oxygen transfer rate because of the minor significance of physically dissolved oxygen for gas transfer. In conclusion, our results regarding a new ECMO mock setup demonstrate that recirculation loops can improve ECMO performance, but not enough to be clinically relevant.
引用
收藏
页数:9
相关论文
共 18 条
[1]   Efficacy of Flow Monitoring During ECMO [J].
Badheka, Aditya ;
Stucker, Sara E. ;
Turek, Joseph W. ;
Raghavan, Madhavan L. .
ASAIO JOURNAL, 2017, 63 (04) :496-500
[2]   Septic porcine blood does not further activate coagulation during in vitro membrane oxygenation [J].
Bleilevens, Christian ;
Grottke, Oliver ;
Groening, Sabine ;
Honickel, Markus ;
Kopp, Ruedger ;
Singh, Smriti ;
Arens, Jutta ;
Rossaint, Rolf .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2017, 51 (03) :449-456
[3]   Twelve Hours In Vitro Biocompatibility Testing of Membrane Oxygenators [J].
Bleilevens, Christian ;
Grottke, Oliver ;
Tillmann, Sabine ;
Honickel, Markus ;
Kopp, Ruedger ;
Arens, Jutta ;
Rossaint, Rolf .
ASAIO JOURNAL, 2015, 61 (05) :548-555
[4]   Extracorporeal Membrane Oxygenation for ARDS in Adults [J].
Brodie, Daniel ;
Bacchetta, Matthew .
NEW ENGLAND JOURNAL OF MEDICINE, 2011, 365 (20) :1905-1914
[5]   Testing of extracorporeal membrane oxygenation circuit related hemolysis using long-term stored packed red cells and fresh frozen plasma [J].
Burda, G ;
Trittenwein, H ;
Carole, H ;
Trittenwein, G .
ARTIFICIAL ORGANS, 2004, 28 (05) :496-499
[6]   Extracorporeal life support for adults with acute respiratory distress syndrome [J].
Combes, Alain ;
Schmidt, Matthieu ;
Hodgson, Carol L. ;
Fan, Eddy ;
Ferguson, Niall D. ;
Fraser, John F. ;
Jaber, Samir ;
Pesenti, Antonio ;
Ranieri, Marco ;
Rowan, Kathryn ;
Shekar, Kiran ;
Slutsky, Arthur S. ;
Brodie, Daniel .
INTENSIVE CARE MEDICINE, 2020, 46 (12) :2464-2476
[7]  
de Villiers Hugo Juan, 2017, J Extra Corpor Technol, V49, P257
[8]   Blood Recirculation Enhances Oxygenation Efficiency of Artificial Lungs [J].
Madhani, Shalv P. ;
May, Alexandra G. ;
Frankowski, Brian J. ;
Burgreen, Greg W. ;
Federspiel, William J. .
ASAIO JOURNAL, 2020, 66 (05) :565-570
[9]   Extracorporeal membrane oxygenation (ECMO) and the acute respiratory distress syndrome (ARDS): a systematic review of pre-clinical models [J].
Millar, Jonathan E. ;
Bartnikowski, Nicole ;
von Bahr, Viktor ;
Malfertheiner, Maximilian, V ;
Obonyo, Nchafatso G. ;
Belliato, Mirko ;
Suen, Jacky Y. ;
Combes, Alain ;
McAuley, Daniel F. ;
Lorusso, Roberto ;
Fraser, John F. .
INTENSIVE CARE MEDICINE EXPERIMENTAL, 2019, 7 (1)
[10]   ECMO for ARDS due to COVID-19 [J].
Namendys-Silva, Silvio A. .
HEART & LUNG, 2020, 49 (04) :348-349