摘要
A novel method, combining isotope dilution with standard additions, was developed for the analysis of eight elements (Mn, Fe, Co, Ni, Cu, Zn, Cd and Pb) in seawater. The method requires just 12 mL of sample and employs an off-line pre-concentration step using the commercially available chelating resin Toyopearl AF-Chelate-650M prior to determination by high resolution inductively coupled plasma magnetic sector mass spectrometry (ICP-MS). Acidified samples were spiked with a multi-element standard of six isotopes (Fe-57, Ni-52, Cu-65, Zn-68, Cd-111 and Pb-207) enriched over natural abundance. In addition, standard additions of a mixed Co and Mn standard were performed on sub-sets of the same sample. All samples were irradiated using a low power (119 mW cm(-2); 254 nm) IN system, to destroy organic ligands, before pre-concentration and extraction from the seawater matrix. Ammonium acetate was used to raise the pH of the 12 mL sub-samples (off-line) to pH 6.4 +/- 0.2 prior to loading onto the chelating resin. The extracted metals were eluted using 1.0 M Q-HNO3 and determined using ICP-MS. The method was verified through the analysis of certified reference material (NASS-5) and the SAFe inter-comparison samples (S1 and D2), the results of which are in good agreement with the certified and reported consensus values. We also present vertical profiles of the eight metals taken from the Bermuda Atlantic Time Series (BATS) station collected during the GEOTRACES inter-comparison cruise in June 2008. Published by Elsevier BM.
引用
收藏
页码:200 / 207
页数:8
相关论文
共 50 条
  • [31] Distribution and stoichiometry of Al, Mn, Fe, Co, Ni, Cu, Zn, Cd, and Pb in seawater around the Juan de Fuca Ridge
    Zheng, Linjie
    Minami, Tomoharu
    Takano, Shotaro
    Minami, Hideki
    Sohrin, Yoshiki
    JOURNAL OF OCEANOGRAPHY, 2017, 73 (05) : 669 - 685
  • [32] Distribution and stoichiometry of Al, Mn, Fe, Co, Ni, Cu, Zn, Cd, and Pb in seawater around the Juan de Fuca Ridge
    Linjie Zheng
    Tomoharu Minami
    Shotaro Takano
    Hideki Minami
    Yoshiki Sohrin
    Journal of Oceanography, 2017, 73 : 669 - 685
  • [33] Thermo-responsive polymer coated fiber-in-tube capillary microextraction and its application to on-line determination of Co, Ni and Cd by inductively coupled plasma mass spectrometry (ICP-MS)
    Zheng, Fei
    Hu, Bin
    TALANTA, 2011, 85 (02) : 1166 - 1173
  • [34] Quantification of Arsenic Species in Wheat Flour Samples by Ion Chromatography Coupled to High Resolution Inductively Coupled Plasma-Mass Spectrometry (IC-HR-ICP-MS)
    Thosago, Magadima Mary
    Botha, Angelique
    Ambushe, Abayneh Ataro
    Godeto, Taddese Wondimu
    ANALYTICAL LETTERS, 2022, 55 (18) : 2860 - 2875
  • [35] Preconcentration and fractionation of Cd, Co, Cu, Ni, Pb and Zn in natural water samples prior to analysis by inductively coupled plasma atomic emission spectrometry
    Pohl, P
    Bogdal, Z
    Prusisz, B
    MICROCHIMICA ACTA, 2005, 150 (3-4) : 253 - 259
  • [36] Validation of the Analytical Procedure for Quantitative Determination of Four Trace Metals (As, Cd, Pb, and Hg) in Fish Lipids Using Inductively Coupled Plasma-Mass Spectrometry (ICP-MS)
    Jayakody, Kasun S.
    Edirisinghe, Ranjith K. B.
    Senevirathne, Suchithra A.
    Senarathna, Lalith
    MASS SPECTROMETRY LETTERS, 2024, 15 (03) : 149 - 157
  • [37] Determination of V, Cr, Mn, Fe, Ni, Cu, Cd, Pb, and U in Zirconium Matrix by Inductively Coupled Plasma Mass Spectrometry Employing Microwave-Assisted Digestion and Matrix Separation by Synergistic Solvent Extraction
    Nagar, B. K.
    Deb, S. B.
    Saxena, M. K.
    Ramakumar, K. L.
    ATOMIC SPECTROSCOPY, 2011, 32 (06) : 213 - 222
  • [38] Determination of 237Np and 239Pu in urine using sector field inductively coupled plasma mass spectrometry (SF-ICP-MS)
    Xiao, Ge
    Liu, Yongzhong
    Jones, Robert L.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2020, 324 (02) : 887 - 896
  • [39] Automated determination of gallium in seawater using seaFAST pre-concentration and high-resolution inductively-coupled plasma mass spectrometry
    Gilbert, Melissa
    Ho, Peng
    Whitmore, Laura
    Shiller, Alan
    ANALYTICA CHIMICA ACTA, 2023, 1241
  • [40] An ICP-MS procedure to determine Cd, Co, Cu, Ni, Pb and Zn in oceanic waters using in-line flow-injection with solid-phase extraction for preconcentration
    O'Sullivan, Jeanette E.
    Watson, Roslyn J.
    Butler, Edward C. V.
    TALANTA, 2013, 115 : 999 - 1010