Improvement of the prompt-gamma neutron activation facility at Brookhaven National Laboratory

被引:21
作者
Dilmanian, FA [1 ]
Lidofsky, LJ
Stamatelatos, I
Kamen, Y
Yasumura, S
Vartsky, D
Pierson, RN
Weber, DA
Moore, RI
Ma, R
机构
[1] Brookhaven Natl Lab, Dept Med, Upton, NY 11973 USA
[2] Columbia Univ, Dept Appl Phys, New York, NY 10027 USA
[3] St Lukes Roosevelt Med Ctr, New York, NY 10025 USA
[4] Soreq Nucl Res Ctr, IL-70600 Yavneh, Israel
关键词
D O I
10.1088/0031-9155/43/2/009
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The prompt-gamma neutron activation facility at Brookhaven National Laboratory was upgraded to improve both the precision and accuracy of its in vivo determinations of total body nitrogen. The upgrade, guided by Monte Carlo simulations, involved elongating and modifying the source collimator and its shielding, repositioning the system's two NaI(TI) detectors, and improving the neutron and gamma shielding of these detectors. The new source collimator has a graphite reflector around the (PuBe)-Pu-238 neutron source to enhance the low-energy region of the neutron spectrum incident on the patient. The gamma detectors have been relocated from positions close to the upward-emerging collimated neutron beam to positions close to and at the sides of the patient. These modifications substantially reduced spurious counts resulting from the capture of small-angle scattered neutrons in the NaI detectors. The pile-up background under the 10.8 MeV N-14(n, gamma)N-15 spectral peak has been reduced so that the nitrogen peak-to-background ratio has been increased by a factor of 2.8. The resulting reduction in the coefficient of variation of the total body nitrogen measurements from 3% to 2.2% has improved the statistical significance of the results possible for any given number of patient measurements. The new system also has a more uniform composite sensitivity.
引用
收藏
页码:339 / 349
页数:11
相关论文
共 30 条
[11]  
*ICRP, 1991, ICRP PUBL, V60, P6
[12]  
*ICRU, 1962, 84 NAT BUR STAND ICR
[13]  
KAMEN Y, 1992, THESIS COLUMBIA U
[14]   USE OF A HIGH REPETITION RATE NEUTRON GENERATOR FOR INVIVO BODY-COMPOSITION MEASUREMENTS VIA NEUTRON INELASTIC-SCATTERING [J].
KEHAYIAS, JJ ;
ELLIS, KJ ;
COHN, SH ;
WEINLEIN, JH .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1987, 24-5 :1006-1009
[15]  
KEHAYIAS JJ, 1987, IN VIVO BODY COMPOSI, P427
[16]   IMPROVED CLINICAL FACILITY FOR INVIVO NITROGEN MEASUREMENT [J].
KRISHNAN, SS ;
MCNEILL, KG ;
MERNAGH, JR ;
BAYLEY, AJ ;
HARRISON, JE .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (04) :489-499
[17]  
MA R, 1993, HUMAN BODY COMPOSITI, V60, P345
[18]   APPARATUS FOR THE MEASUREMENT OF TOTAL-BODY NITROGEN USING PROMPT NEUTRON-ACTIVATION ANALYSIS WITH CF-252 [J].
MACKIE, A ;
HANNAN, WJ ;
SMITH, MA ;
TOTHILL, P .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1988, 12 (04) :152-159
[19]  
*MCNP, 1992, MONT CARL NEUTR PHOT
[20]   INVIVO DETERMINATION OF NITROGEN USING PU-BE SOURCES [J].
MERNAGH, JR ;
HARRISON, JE ;
MCNEILL, KG .
PHYSICS IN MEDICINE AND BIOLOGY, 1977, 22 (05) :831-835