Capsule-Based dry powder inhaler evaluation using CFD-DEM simulations and next generation impactor data

被引:12
作者
Almeida, Lucilla C. [1 ]
Bharadwaj, Rahul [1 ]
Eliahu, Avi [2 ]
Wassgren, Carl R. [3 ]
Nagapudi, Karthik [4 ]
Muliadi, Ariel R. [4 ]
机构
[1] Engn Simulat & Sci Software, Rua Orlando Phillipi 100, BR-88032700 Florianopolis, SC, Brazil
[2] Genentech Inc, Device Dev, 1 DNA Way, South San Francisco, CA 94080 USA
[3] Purdue Univ, Sch Mech Engn, 585 Purdue Mall, W Lafayette, IN 47907 USA
[4] Genentech Inc, Small Mol Pharmaceut, Res & Early Dev gRED, 1 DNA Way, South San Francisco, CA 94080 USA
关键词
CFD-DEM; Dry powder inhalers; Discrete element method; Computational fluid dynamics; Inhalation drug delivery; COMPUTATIONAL FLUID-DYNAMICS; OROPHARYNGEAL DEPOSITION; PHARMACEUTICAL IMPACTOR; PARTICLE-SIZE; DEVICE DESIGN; PERFORMANCE; COMPARABILITY; FORMULATION; MODEL; FLOW;
D O I
10.1016/j.ejps.2022.106226
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
Capsule-based, single-dose dry powder inhalers (DPIs) are commonly-used devices to deliver medications to the lungs. This work evaluates the effect of the drug/excipient adhesive bonding and the DPI resistances on the aerosol performance using a combination of empirical multi-stage impactor data and a fully-coupled computational fluid dynamics (CFD) and discrete element method (DEM) model. Model-predicted quantities show that the primary modes of powder dispersion are a function of the device resistance. Lowering the device resistance increases its capacity to transport a wider range of particle size classes toward the outlet and generate more intense turbulence upstream therein. On the other hand, a higher device resistance increases the velocity of the tangential airflow along the device walls, which in turn increases the intensity of particle/device impaction. Correlating model data and experimental results shows that these differing powder dispersion mechanisms affect different formulations differently, with finer aerosols tending to result when pairing a lower resistance device with formulations that exhibit low API/excipient adhesion, or when pairing a high resistance device with more cohesive formulations.
引用
收藏
页数:19
相关论文
共 48 条
[41]   CFD-DEM investigation of the dispersion mechanisms in commercial dry powder inhalers [J].
Tong, Z. B. ;
Zheng, B. ;
Yang, R. Y. ;
Yu, A. B. ;
Chan, H. K. .
POWDER TECHNOLOGY, 2013, 240 :19-24
[42]   Numerical study of the effects of particle size and polydispersity on the agglomerate dispersion in a cyclonic flow [J].
Tong, Z. B. ;
Yang, R. Y. ;
Chu, K. W. ;
Yu, A. B. ;
Adi, S. ;
Chan, H. K. .
CHEMICAL ENGINEERING JOURNAL, 2010, 164 (2-3) :432-441
[43]  
Walton O, 1993, BUTTERWORTH HEINEMAN
[44]   Particle Aerosolisation and Break-Up in Dry Powder Inhalers: Evaluation and Modelling of Impaction Effects for Agglomerated Systems [J].
Wong, William ;
Fletcher, David F. ;
Traini, Daniela ;
Chan, Hak-Kim ;
Crapper, John ;
Young, Paul M. .
JOURNAL OF PHARMACEUTICAL SCIENCES, 2011, 100 (07) :2744-2754
[45]   Particle Aerosolisation and Break-up in Dry Powder Inhalers 1: Evaluation and Modelling of Venturi Effects for Agglomerated Systems [J].
Wong, William ;
Fletcher, David F. ;
Traini, Daniela ;
Chan, Hak-Kim ;
Crapper, John ;
Young, Paul M. .
PHARMACEUTICAL RESEARCH, 2010, 27 (07) :1367-1376
[46]   Dry Powder Aerosols Generated by Standardized Entrainment Tubes From Drug Blends With Lactose Monohydrate: 2. Ipratropium Bromide Monohydrate and Fluticasone Propionate [J].
Xu, Zhen ;
Mansour, Heidi M. ;
Mulder, Tako ;
McLean, Richard ;
Langridge, John ;
Hickey, Anthony J. .
JOURNAL OF PHARMACEUTICAL SCIENCES, 2010, 99 (08) :3415-3429
[47]   Agglomeration of fine particles subjected to centripetal compaction [J].
Yang, R. Y. ;
Yu, A. B. ;
Choi, S. K. ;
Coates, M. S. ;
Chan, H. K. .
POWDER TECHNOLOGY, 2008, 184 (01) :122-129
[48]   Flow and Particle Modelling of Dry Powder Inhalers: Methodologies, Recent Development and Emerging Applications [J].
Zheng, Zhanying ;
Leung, Sharon Shui Yee ;
Gupta, Raghvendra .
PHARMACEUTICS, 2021, 13 (02)