Taking a look at the calibration of a CCD detector with a fiber-optic taper

被引:3
作者
Alkire, R. W. [1 ]
Rotella, F. J. [1 ]
Duke, N. E. C. [1 ]
Otwinowski, Zbyszek [2 ]
Borek, Dominika [2 ]
机构
[1] Argonne Natl Lab, Biosci Div, 9700 S Cass Ave, Argonne, IL 60439 USA
[2] Univ Texas SW Med Ctr Dallas, Dept Biophys, 5323 Harry Hines Blvd, Dallas, TX 75390 USA
关键词
CCD detectors; fiber-optic tapers; flat-field calibration; data scaling; anomalous signal; CHARGE-COUPLED-DEVICE; X-RAY-DETECTOR; PROTEIN CRYSTALLOGRAPHY; DIFFRACTION INTENSITIES; RADIATION-DAMAGE; AREA DETECTORS; SYNCHROTRON; RESOLUTION; INTEGRATION; REFINEMENT;
D O I
10.1107/S1600576716000431
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
At the Structural Biology Center beamline 19BM, located at the Advanced Photon Source, the operational characteristics of the equipment are routinely checked to ensure they are in proper working order. After performing a partial flat-field calibration for the ADSC Quantum 210r CCD detector, it was confirmed that the detector operates within specifications. However, as a secondary check it was decided to scan a single reflection across one-half of a detector module to validate the accuracy of the calibration. The intensities from this single reflection varied by more than 30% from the module center to the corner of the module. Redistribution of light within bent fibers of the fiber-optic taper was identified to be a source of this variation. The degree to which the diffraction intensities are corrected to account for characteristics of the fiber-optic tapers depends primarily upon the experimental strategy of data collection, approximations made by the data processing software during scaling, and crystal symmetry.
引用
收藏
页码:415 / 425
页数:11
相关论文
共 30 条
[1]   Design of a vacuum-compatible high-precision monochromatic beam-position monitor for use with synchrotron radiation from 5 to 25 keV [J].
Alkire, RW ;
Rosenbaum, G ;
Evans, G .
JOURNAL OF SYNCHROTRON RADIATION, 2000, 7 (02) :61-68
[2]  
[Anonymous], 1969, UCRL50174 LAWR LIV N
[3]  
Arvai A. S., 2007, AM CRYST ASS M 21 26
[4]   Calibration procedures for charge-coupled device x-ray detectors [J].
Barna, SL ;
Tate, MW ;
Gruner, SM ;
Eikenberry, EF .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1999, 70 (07) :2927-2934
[5]   Measurement errors and their consequences in protein crystallography [J].
Borek, D ;
Minor, W ;
Otwinowski, Z .
ACTA CRYSTALLOGRAPHICA SECTION D-BIOLOGICAL CRYSTALLOGRAPHY, 2003, 59 :2031-2038
[6]   Identification of patterns in diffraction intensities affected by radiation exposure [J].
Borek, Dominika ;
Dauter, Zbigniew ;
Otwinowski, Zbyszek .
JOURNAL OF SYNCHROTRON RADIATION, 2013, 20 :37-48
[7]   The many faces of radiation-induced changes [J].
Borek, Dominika ;
Ginell, Stephan L. ;
Cymborowski, Marcin ;
Minor, Wladek ;
Otwinowski, Zbyszek .
JOURNAL OF SYNCHROTRON RADIATION, 2007, 14 :24-33
[8]   Diffraction data analysis in the presence of radiation damage [J].
Borek, Dominika ;
Cymborowski, Marcin ;
Machius, Mischa ;
Minor, Wladek ;
Otwinowski, Zbyszek .
ACTA CRYSTALLOGRAPHICA SECTION D-STRUCTURAL BIOLOGY, 2010, 66 :426-436
[9]   Quantifying instrument errors in macromolecular X-ray data sets [J].
Diederichs, Kay .
ACTA CRYSTALLOGRAPHICA SECTION D-STRUCTURAL BIOLOGY, 2010, 66 :733-740
[10]   How good are my data and what is the resolution? [J].
Evans, Philip R. ;
Murshudov, Garib N. .
ACTA CRYSTALLOGRAPHICA SECTION D-BIOLOGICAL CRYSTALLOGRAPHY, 2013, 69 :1204-1214