First assessment of simultaneous dual isotope (123I/99mTc) cardiac SPECT on two different CZT cameras: A phantom study

被引:21
作者
Blaire, Tanguy [1 ,2 ,3 ]
Bailliez, Alban [1 ,2 ,3 ]
Ben Bouallegue, Faycal [4 ]
Bellevre, Dimitri [5 ]
Agostini, Denis [2 ,5 ]
Manrique, Alain [2 ,5 ]
机构
[1] Grp Hop Inst Catholique Lille, Dept Nucl Med, UF 5881, Lomme Les Lille, France
[2] Normandie Univ, UNICAEN, Signalisat Electrophysiol & Imagerie Les Ischemie, FHU REMOD VHF, Caen, France
[3] Hop Prive Le Bois, Dept Nucl Med, Lille, France
[4] CHU Montpellier, Dept Nucl Med, Montpellier, France
[5] CHU Cote Nacre, Dept Nucl Med, Caen, France
关键词
CZT; DSPECT; DNM; 530c; mIBG; dual isotope; mismatch; LEFT-VENTRICULAR FUNCTION; HEART-FAILURE; GAMMA-CAMERA; DILATED CARDIOMYOPATHY; CLINICAL VALIDATION; SIMULTANEOUS TC-99M; PROGNOSTIC VALUE; INNERVATION; ASSOCIATION; PERFORMANCE;
D O I
10.1007/s12350-017-0841-z
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We studied the impact of simultaneous dual-isotope acquisition on I-123/Tc-99m mismatch assessment using two CZT cameras (DNM 530c, GE Healthcare and DSPECT, Biosensors International). We used an anthropomorphic torso phantom (respectively filled with a solution of I-123 alone, Tc-99m alone, and a mixture of I-123 and Tc-99m) and its cardiac insert with two defects mimicking two matched and mismatched defects. Mismatch extent and reconstructed image contrast were evaluated. The acquisition mode (single vs dual) significantly impacted (i) Tc-99m (but not I-123) reconstructed segmental activities using both camera (P < .001), and (ii) image contrast (using I-123 and DNM 530c, P < .0001; and using both I-123 and Tc-99m with DSPECT, P < .0001). However, the defect and mismatch size were not impacted by the type of acquisition. With both DNM 530c and DSPECT, Lin's concordance correlation coefficient and Bland-Altman analysis demonstrated an almost perfect concordance and agreement between single- and simultaneous dual-isotope segmental activity (I-123 and Tc-99m). This study found no impact of the acquisition mode (single vs dual) or the type of camera (DSPECT vs DNM 530c) on I-123 and Tc-99m defect size and mismatch, providing a new step toward simultaneous dual-isotope acquisition for combined innervation and perfusion assessment.
引用
收藏
页码:1692 / 1704
页数:13
相关论文
共 31 条
[1]   Global and Regional Myocardial Innervation Before and After Ablation of Drug-Refractory Ventricular Tachycardia Assessed with 123I-MIBG [J].
Abdulghani, Mohammed ;
Duell, John ;
Smith, Mark ;
Chen, Wengen ;
Bentzen, Soren M. ;
Asoglu, Ramazan ;
Klein, Tomas ;
Bob-Manuel, Tamunoinemi ;
Saliaris, Anastasios ;
See, Vincent ;
Shorofsky, Stephen ;
Dilsizian, Vasken ;
Dickfeld, Timm .
JOURNAL OF NUCLEAR MEDICINE, 2015, 56 :52S-58S
[2]   Performance of cardiac cadmium-zinc-telluride gamma camera imaging in coronary artery disease: a review from the cardiovascular committee of the European Association of Nuclear Medicine (EANM) [J].
Agostini, Denis ;
Marie, Pierre-Yves ;
Ben-Haim, Simona ;
Rouzet, Francois ;
Songy, Bernard ;
Giordano, Alessandro ;
Gimelli, Alessia ;
Hyafil, Fabien ;
Sciagra, Roberto ;
Bucerius, Jan ;
Verberne, Hein J. ;
Slart, Riemer H. J. A. ;
Lindner, Oliver .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2016, 43 (13) :2423-2432
[3]   Left Ventricular Function Assessment Using 2 Different Cadmium-Zinc-Telluride Cameras Compared with a γ-Camera with Cardiofocal Collimators: Dynamic Cardiac Phantom Study and Clinical Validation [J].
Bailliez, Alban ;
Lairez, Olivier ;
Merlin, Charles ;
Piriou, Nicolas ;
Legallois, Damien ;
Blaire, Tanguy ;
Agostini, Denis ;
Valette, Frederic ;
Manrique, Alain .
JOURNAL OF NUCLEAR MEDICINE, 2016, 57 (09) :1370-1375
[4]   Segmental and global left ventricular function assessment using gated SPECT with a semiconductor Cadmium Zinc Telluride (CZT) camera: Phantom study and clinical validation vs cardiac magnetic resonance [J].
Bailliez, Alban ;
Blaire, Tanguy ;
Mouquet, Frederic ;
Legghe, R. ;
Etienne, B. ;
Legallois, Damien ;
Agostini, Denis ;
Manrique, Alain .
JOURNAL OF NUCLEAR CARDIOLOGY, 2014, 21 (04) :712-722
[5]   First determination of the heart-to-mediastinum ratio using cardiac dual isotope (123I-MIBG/99mTc-tetrofosmin) CZT imaging in patients with heart failure: the ADRECARD study [J].
Bellevre, Dimitri ;
Manrique, Alain ;
Legallois, Damien ;
Bross, Samy ;
Baavour, Rafael ;
Roth, Nathaniel ;
Blaire, Tanguy ;
Desmonts, Cedric ;
Bailliez, Alban ;
Agostini, Denis .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 (12) :1912-1919
[6]   Quantitative assessment of myocardial perfusion abnormality on SPECT myocardial perfusion imaging is more reproducible than expert visual analysis [J].
Berman, Daniel S. ;
Kang, Xingping ;
Gransar, Heidi ;
Gerlach, James ;
Friedman, John D. ;
Hayes, Sean W. ;
Thomson, Louise E. J. ;
Hachamovitch, Rory ;
Shaw, Leslee J. ;
Slomka, Piotr J. ;
De Yang, Ling ;
Germano, Guido .
JOURNAL OF NUCLEAR CARDIOLOGY, 2009, 16 (01) :45-53
[7]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[8]   A fast cardiac gamma camera with dynamic SPECT capabilities: design, system validation and future potential [J].
Bocher, Moshe ;
Blevis, Ira M. ;
Tsukerman, Leonid ;
Shrem, Yigal ;
Kovalski, Gil ;
Volokh, Lana .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2010, 37 (10) :1887-1902
[9]   Cardiac Sympathetic Imaging With mIBG in Heart Failure [J].
Carrio, Ignasi ;
Cowie, Martin R. ;
Yamazaki, Junichi ;
Udelson, James ;
Camici, Paolo G. .
JACC-CARDIOVASCULAR IMAGING, 2010, 3 (01) :92-100
[10]   Absolute Quantification of Left Ventricular Global and Regional Function at Nuclear MPI Using Ultrafast CZT SPECT: Initial Validation Versus Cardiac MR [J].
Cochet, Hubert ;
Bullier, Eric ;
Gerbaud, Edouard ;
Durieux, Muriel ;
Godbert, Yann ;
Lederlin, Mathieu ;
Coste, Pierre ;
Barat, Jean-Louis ;
Laurent, Francois ;
Montaudon, Michel .
JOURNAL OF NUCLEAR MEDICINE, 2013, 54 (04) :556-563