CT Metal Artifact Reduction Using MR Image Patches

被引:0
作者
Nielsen, Jonathan S. [1 ,2 ]
Edmund, Jens M. [2 ,3 ]
Van Leemput, Koen [1 ,4 ]
机构
[1] Tech Univ Denmark, Dept Appl Math & Comp Sci, Lyngby, Denmark
[2] Gentofte & Herlev Hosp, Dept Oncol, Radiotherapy Res Unit, Herlev, Denmark
[3] Univ Copenhagen, Niels Bohr Inst, Copenhagen, Denmark
[4] Harvard Med Sch, Massachusetts Gen Hosp, Dept Radiol, Boston, MA USA
来源
MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING | 2018年 / 10573卷
关键词
Computed Tomography; Metal Artifact Reduction; Bayesian modeling; Radiation Therapy; ATTENUATION CORRECTION; COMPUTED-TOMOGRAPHY; PSEUDO-CT; RECONSTRUCTION; ALGORITHM;
D O I
10.1117/12.2293815
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Metal implants give rise to metal artifacts in computed tomography (CT) images, which may lead to diagnostic errors and erroneous CT number estimates when the CT is used for radiation therapy planning. Methods for reducing metal artifacts by exploiting the anatomical information provided by coregistered magnetic resonance (MR) images are of great potential value, but remain technically challenging due to the poor contrast between bone and air on the MR image. In this paper, we present a novel MR-based algorithm for automatic CT metal artifact reduction (MAR), referred to as kerMAR. It combines kernel regression on known CT value/MR patch pairs in the uncorrupted patient volume with a forward model of the artifact corrupted values to estimate CT replacement values. In contrast to pseudo-CT generation that builds on multi-patient modelling, the algorithm requires no MR intensity normalisation or atlas registration. Image results for 7 head-and-neck radiation therapy patients with T1-weighted images acquired in the same fixation as the RT planning CT suggest a potential for more complete MAR close to the metal implants than the oMAR algorithm (Philips) used clinically. Our results further show improved performance in air and bone regions as compared to other MR-based MAR algorithms. In addition, we experimented with using kerMAR to define a prior for iterative reconstruction with the maximum likelihood transmission reconstruction algorithm, however with no apparent improvements
引用
收藏
页数:10
相关论文
共 31 条
[1]   Metal artifact reduction strategies for improved attenuation correction in hybrid PET/CT imaging [J].
Abdoli, Mehrsima ;
Dierckx, Rudi A. J. O. ;
Zaidi, Habib .
MEDICAL PHYSICS, 2012, 39 (06) :3343-3360
[2]  
Anderla AA, 2013, INT C TELECOMMUN MO, P337, DOI 10.1109/TELSKS.2013.6704943
[3]   Patch-based generation of a pseudo CT from conventional MRI sequences for MRI-only radiotherapy of the brain [J].
Andreasen, Daniel ;
Van Leemput, Koen ;
Hansen, Rasmus H. ;
Andersen, Jon A. L. ;
Edmund, Jens M. .
MEDICAL PHYSICS, 2015, 42 (04) :1596-1605
[4]  
[Anonymous], EXPECTATION MA UNPUB
[5]  
[Anonymous], 2008, Computed tomography: from photonstatistics to modern cone-beam CT
[6]  
Bishop Christopher M, 2016, Pattern recognition and machine learning
[7]   MR-driven metal artifact reduction in PET/CT [J].
Delso, G. ;
Wollenweber, S. ;
Lonn, A. ;
Wiesinger, F. ;
Veit-Haibach, P. .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (07) :2267-2280
[8]  
DeMan B, 2001, THESIS
[9]   Metal artifacts in computed tomography for radiation therapy planning: dosimetric effects and impact of metal artifact reduction [J].
Giantsoudi, Drosoula ;
De Man, Bruno ;
Verburg, Joost ;
Trofimov, Alexei ;
Jin, Yannan ;
Wang, Ge ;
Gjesteby, Lars ;
Paganetti, Harald .
PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (08) :R49-R80
[10]   MRI-Based Attenuation Correction for PET/MRI: A Novel Approach Combining Pattern Recognition and Atlas Registration [J].
Hofmann, Matthias ;
Steinke, Florian ;
Scheel, Verena ;
Charpiat, Guillaume ;
Farquhar, Jason ;
Aschoff, Philip ;
Brady, Michael ;
Schoelkopf, Bernhard ;
Pichler, Bernd J. .
JOURNAL OF NUCLEAR MEDICINE, 2008, 49 (11) :1875-1883