Influence of beam efficiency through the patient-specific collimator on secondary neutron dose equivalent in double scattering and uniform scanning modes of proton therapy

被引:23
作者
Hecksel, D. [1 ]
Anferov, V. [2 ]
Fitzek, M. [1 ]
Shahnazi, K. [1 ]
机构
[1] Indiana Univ, Sch Med, Midwest Proton Radiotherapy Inst, Bloomington, IN 47408 USA
[2] Indiana Univ, Cyclotron Facil, Bloomington, IN 47408 USA
关键词
proton therapy; neutrons; dose equivalent; MONTE-CARLO SIMULATIONS; PROSTATE-CANCER; STRAY RADIATION; 2ND CANCER; RISK; RADIOTHERAPY; EXPOSURES; IRRADIATION; MOTHER; SYSTEM;
D O I
10.1118/1.3431575
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Conventional proton therapy facilities use double scattering nozzles, which are optimized for delivery of a few fixed field sizes. Similarly, uniform scanning nozzles are commissioned for a limited number of field sizes. However, cases invariably occur where the treatment field is significantly different from these fixed field sizes. The purpose of this work was to determine the impact of the radiation field conformity to the patient-specific collimator on the secondary neutron dose equivalent. Methods: Using a WENDI-II neutron detector, the authors experimentally investigated how the neutron dose equivalent at a particular point of interest varied with different collimator sizes, while the beam spreading was kept constant. The measurements were performed for different modes of dose delivery in proton therapy, all of which are available at the Midwest Proton Radiotherapy Institute (MPRI): Double scattering, uniform scanning delivering rectangular fields, and uniform scanning delivering circular fields. The authors also studied how the neutron dose equivalent changes when one changes the amplitudes of the scanned field for a fixed collimator size. Results: The secondary neutron dose equivalent was found to decrease linearly with the collimator area for all methods of dose delivery. The relative values of the neutron dose equivalent for a collimator with a 5 cm diameter opening using 88 MeV protons were 1.0 for the double scattering field, 0.76 for rectangular uniform field, and 0.6 for the circular uniform field. Furthermore, when a single circle wobbling was optimized for delivery of a uniform field 5 cm in diameter, the secondary neutron dose equivalent was reduced by a factor of 6 compared to the double scattering nozzle. Additionally, when the collimator size was kept constant, the neutron dose equivalent at the given point of interest increased linearly with the area of the scanned proton beam. Conclusions: The results of these experiments suggest that the patient-specific collimator is a significant contributor to the secondary neutron dose equivalent to a distant organ at risk. Improving conformity of the radiation field to the patient-specific collimator can significantly reduce secondary neutron dose equivalent to the patient. Therefore, it is important to increase the number of available generic field sizes in double scattering systems as well as in uniform scanning nozzles. (C) 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3431575]
引用
收藏
页码:2910 / 2917
页数:8
相关论文
共 30 条
  • [1] Secondary neutron and photon dose in proton therapy
    Agosteo, S
    Birattari, C
    Caravaggio, M
    Silari, M
    Tosi, G
    [J]. RADIOTHERAPY AND ONCOLOGY, 1998, 48 (03) : 293 - 305
  • [2] Combined X-Y scanning magnet for conformal proton radiation therapy
    Anferov, V
    [J]. MEDICAL PHYSICS, 2005, 32 (03) : 815 - 818
  • [3] Scan pattern optimization for uniform proton beam scanning
    Anferov, Vladimir A.
    [J]. MEDICAL PHYSICS, 2009, 36 (08) : 3560 - 3567
  • [4] Secondary dose exposures during 200 MeV proton therapy
    Binns, PJ
    Hough, JH
    [J]. RADIATION PROTECTION DOSIMETRY, 1997, 70 (1-4) : 441 - 444
  • [5] Clinical characterization of a proton beam continuous uniform scanning system with dose layer stacking
    Farr, J. B.
    Mascia, A. E.
    Hsi, W. -C.
    Allgower, C. E.
    Jesseph, F.
    Schreuder, A. N.
    Wolanski, M.
    Nichiporov, D. F.
    Anferov, V.
    [J]. MEDICAL PHYSICS, 2008, 35 (11) : 4945 - 4954
  • [6] Equivalent dose and effective dose from stray radiation during passively scattered proton radiotherapy for prostate cancer
    Fontenot, Jonas
    Taddei, Phillip
    Zheng, Yuanshui
    Mirkovic, Dragan
    Jordan, Thomas
    Newhauser, Wayne
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (06) : 1677 - 1688
  • [7] RISK OF SECONDARY MALIGNANT NEOPLASMS FROM PROTON THERAPY AND INTENSITY-MODULATED X-RAY THERAPY FOR EARLY-STAGE PROSTATE CANCER
    Fontenot, Jonas D.
    Lee, Andrew K.
    Newhauser, Wayne D.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 74 (02): : 616 - 622
  • [8] Gottschalk B., 2008, Proton and Charged Particle Radiotherapy
  • [9] Neutron dose in scattered and scanned proton beams:: In regard to Eric J!Hall (int j radiat oncol biol phys 2006;65:1-7)
    Gottschalk, Bernard
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (05): : 1594 - 1594
  • [10] Intensity-modulated radiation therapy, protons, and the risk of second cancers
    Hall, Eric J.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 65 (01): : 1 - 7