Carotid Arterial Wall MRI at 3T Using 3D Variable-Flip-Angle Turbo Spin-Echo (TSE) with Flow-Sensitive Dephasing (FSD)

被引:129
作者
Fan, Zhaoyang [1 ,2 ]
Zhang, Zhuoli [1 ]
Chung, Yiu-Cho [3 ]
Weale, Peter [3 ]
Zuehlsdorff, Sven [3 ]
Carr, James [1 ]
Li, Debiao [1 ,2 ]
机构
[1] Northwestern Univ, Dept Radiol, Chicago, IL 60611 USA
[2] Northwestern Univ, Dept Biomed Engn, Evanston, IL 60208 USA
[3] Siemens Med Solut USA Inc, Chicago, IL USA
关键词
carotid artery; magnetic resonance imaging; atherosclerosis; 3D SPACE; black-blood; RAPID EXTENDED COVERAGE; BLACK-BLOOD MRI; ATHEROSCLEROTIC PLAQUE; MAGNETIC-RESONANCE; 1.5; T; ARTIFACTS; ANGIOGRAPHY; SEQUENCES; CONTRAST; LESIONS;
D O I
10.1002/jmri.22058
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate the effectiveness of flow-sensitive dephasing (FSD) magnetization preparation in improving blood signal suppression of three-dimensional (3D) turbo spin-echo (TSE) sequence (SPACE) for isotropic high-spatial-resolution carotid arterial wall imaging at 3T. Materials and Methods: The FSD-prepared SPACE sequence (FSD-SPACE) was implemented by adding two identical FSD gradient pulses right before and after the first refocusing 180 degrees-pulse of the SPACE sequence in all three orthogonal directions. Nine healthy volunteers were imaged at 3T with SPACE, FSD-SPACE, and multislice T2-weighted 2D TSE coupled with saturation band (SB-TSE). Apparent carotid wall-lumen contrast-to-noise ratio (aCNR(w-1)) and apparent lumen area (aLA) at the locations with residual-blood (rb) signal shown on SPACE images were compared between SPACE and FSD-SPACE. Carotid aCNR(w-1) and lumen (LA) and wall area (WA) measured from FSD-SPACE were compared to those measured from SB-TSE. Results: Plaque-mimicking flow artifacts identified in seven carotids on SPACE images were eliminated on FSD-SPACE images. The FSD preparation resulted in slightly reduced aCNR(w-1) (P = 0.025), but significantly improved aCNR between the wall and rb regions (P < 0.001) and larger aLA (P < 0.001). Compared to SB-TSE, FSD-SPACE offered comparable aCNR(w-1) with much higher spatial resolution, shorter imaging time, and larger artery coverage. The LA and WA measurements from the two techniques were in good agreement based on intraclasss correlation coefficient (0.988 and 0.949, respectively; P < 0.001) and Bland-Altman analyses. Conclusion: FSD-SPACE is a time-efficient 3D imaging technique for carotid arterial wall with superior spatial resolution and blood signal suppression.
引用
收藏
页码:645 / 654
页数:10
相关论文
共 42 条
[1]   Intracranial black-blood MR angiography with high-resolution 3D fast spin echo [J].
Alexander, AL ;
Buswell, HR ;
Sun, Y ;
Chapman, BE ;
Tsuruda, TS ;
Parker, DL .
MAGNETIC RESONANCE IN MEDICINE, 1998, 40 (02) :298-310
[2]   Comparison between 2D and 3D high-resolution black-blood techniques for carotid artery wall Imaging in clinically significant atherosclerosis [J].
Balu, Niranjan ;
Chu, Baocheng ;
Hatsukami, Thomas S. ;
Yuan, Chun ;
Yarnykh, Vasily L. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 27 (04) :918-924
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   Fast spin echo sequences with very long echo trains:: Design of variable refocusing flip angle schedules and generation of clinical T2 contrast [J].
Busse, RF ;
Hariharan, H ;
Vu, A ;
Brittain, JH .
MAGNETIC RESONANCE IN MEDICINE, 2006, 55 (05) :1030-1037
[5]   Classification of human carotid atherosclerotic lesions with in vivo multicontrast magnetic resonance imaging [J].
Cai, JM ;
Hastukami, TS ;
Ferguson, MS ;
Small, R ;
Polissar, NL ;
Yuan, C .
CIRCULATION, 2002, 106 (11) :1368-1373
[6]  
CHUNG YC, 2007, P 15 ANN M ISMRM BER
[7]  
CHUNG YC, 2008, P 11 ANN M SCMR LOS
[8]   3D volume-selective turbo spin echo for carotid artery wall imaging with navigator detection of swallowing [J].
Crowe, LA ;
Keegan, J ;
Gatehouse, PD ;
Mohiaddin, RH ;
Varghese, A ;
Symmonds, K ;
Cannell, TM ;
Yang, GZ ;
Firmin, DN .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2005, 22 (04) :583-588
[9]   Volume-selective 3D turbo spin echo imaging for vascular wall imaging and distensibility measurement [J].
Crowe, LA ;
Gatehouse, P ;
Yang, GZ ;
Mohiaddin, RH ;
Varghese, A ;
Charrier, C ;
Keegan, J ;
Firmin, DN .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2003, 17 (05) :572-580
[10]   FAST SELECTIVE BLACK BLOOD MR IMAGING [J].
EDELMAN, RR ;
CHIEN, D ;
KIM, D .
RADIOLOGY, 1991, 181 (03) :655-660