Impact of Length and Hydrophilic Coating of the Introducer Sheath on Radial Artery Spasm During Transradial Coronary Intervention A Randomized Study

被引:153
作者
Rathore, Sudhir [1 ]
Stables, Rodney H. [1 ]
Pauriah, Maheshwar [1 ]
Hakeem, Abdul [1 ]
Mills, Joseph D. [1 ]
Palmer, Nick D. [1 ]
Perry, Raphael A. [1 ]
Morris, John L. [1 ]
机构
[1] Liverpool Heart & Chest Hosp, Dept Cardiol, Liverpool L14 3PE, Merseyside, England
关键词
transradial coronary procedures; radial artery spasm; randomized study; radial artery occlusion; introducer sheaths; CARDIAC-CATHETERIZATION; VASCULAR COMPLICATIONS; FEMORAL APPROACH; ALLENS TEST; ANGIOPLASTY; ANGIOGRAPHY; DISCOMFORT; PATIENT;
D O I
10.1016/j.jcin.2010.03.009
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives The aim of this study was to assess the impact of length and hydrophilic coating of the introducer sheath on radial artery spasm, radial artery occlusion, and local vascular complications in patients undergoing transradial coronary procedures. Background Radial artery spasm is common during transradial procedures and the most common cause for procedural failure. Methods We randomly assigned, in a factorial design, 790 patients scheduled for a transradial coronary procedure to long (23-cm) or short (13-cm) and hydrophilic-coated or uncoated introducer sheaths. The primary outcome measure was clinical evidence of radial artery spasm, and secondary outcome measures were patient discomfort and local vascular complications. Results Procedural success was achieved in 96% of the cases, and radial artery spasm accounted for 17 of 33 failed cases. There was significantly less radial artery spasm (19.0% vs. 39.9%, odds ratio [OR]: 2.87; 95% confidence interval [CI]: 2.07 to 3.97, p < 0.001) and patient reported discomfort (15.1% vs. 28.5%, OR: 2.27; 95% CI: 1.59 to 3.23, p < 0.001) in patients receiving a hydrophilic-coated sheath. No difference was observed between long and short sheaths. Radial artery occlusion was observed in 9.5% of the patients and was not influenced by sheath length or coating. A local large hematoma or arterial dissection was seen in 2.6% of the patients with no difference in groups allocated at randomization. Younger age, female sex, diabetes, and lower body mass index were identified as independent predictors of radial artery spasm. Conclusions Hydrophilic sheath coating, but not sheath length, reduces the incidence of radial artery spasm during transradial coronary procedures. (J Am Coll Cardiol Intv 2010;3:475-83) (C) 2010 by the American College of Cardiology Foundation
引用
收藏
页码:475 / 483
页数:9
相关论文
共 45 条
[1]   Radial versus femoral approach for percutaneous coronary diagnostic and interventional procedures - Systematic overview and meta-analysis of randomized trials [J].
Agostoni, P ;
Biondi-Zoccai, GGL ;
De Benedictis, ML ;
Rigattieri, S ;
Turri, M ;
Anselmi, M ;
Vassanelli, C ;
Zardini, P ;
Louvard, Y ;
Hamon, M .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 44 (02) :349-356
[2]   Transulnar versus transradial artery approach for coronary angioplasty: The PCVI-CUBA study [J].
Aptecar, E ;
Pernes, JM ;
Chabane-Chaouch, M ;
Bussy, N ;
Catarino, G ;
Shahmir, A ;
Bougrini, K ;
Dupouy, P .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2006, 67 (05) :711-720
[3]   Evaluation of the ulnopalmar arterial arches with pulse oximetry and plethysmography:: Comparison with the Allen's test in 1010 patients [J].
Barbeau, GR ;
Arsenault, F ;
Dugas, L ;
Simard, S ;
Larivière, MM .
AMERICAN HEART JOURNAL, 2004, 147 (03) :489-493
[4]   PERCUTANEOUS RADIAL ARTERY APPROACH FOR CORONARY ANGIOGRAPHY [J].
CAMPEAU, L .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1989, 16 (01) :3-7
[5]  
Caputo RP, 2000, CATHETER CARDIO INTE, V51, P287, DOI 10.1002/1522-726X(200011)51:3<287::AID-CCD8>3.0.CO
[6]  
2-3
[7]   Vascular complications and clinical outcome after coronary angioplasty with platelet IIb/IIIa receptor blockade - Comparison of transradial vs transfemoral arterial access [J].
Choussat, R ;
Black, A ;
Bossi, I ;
Fajadet, J ;
Marco, J .
EUROPEAN HEART JOURNAL, 2000, 21 (08) :662-667
[8]   Reduction of discomfort at sheath removal during transradial coronary procedures with the use of a hydrophilic-coated sheath [J].
Dery, JP ;
Simard, S ;
Barbeau, GR .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2001, 54 (03) :289-294
[9]  
GHURAN A, 2005, HEART S, V91
[10]  
Grinfeld L., 1996, J. Am. Coll. Cardiol, V27, P901