Comparison of Autof ms1000 and Bruker Biotyper MALDI-TOF MS Platforms for Routine Identification of Clinical Microorganisms

被引:13
|
作者
Park, Jae Hyeon [1 ,2 ]
Jang, Yujin [3 ]
Yoon, Inseon [2 ]
Kim, Taek Soo [1 ,2 ]
Park, Hyunwoong [1 ,4 ]
机构
[1] Seoul Natl Univ, Dept Lab Med, Coll Med, Seoul 03080, South Korea
[2] Seoul Natl Univ Hosp, Dept Lab Med, Seoul 03080, South Korea
[3] Seoul Natl Univ Hosp, Biomed Res Inst, Seoul 03080, South Korea
[4] Seoul Natl Univ, Dept Lab Med, Boramae Med Ctr, Seoul 07061, South Korea
关键词
D O I
10.1155/2021/6667623
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS) is widely used in clinical microbiology laboratories because it is cost-effective, reliable, and fast. This study is aimed at comparing the identification performance of the recently developed Autof ms1000 (Autobio, China) with that of the Bruker Biotyper (Bruker Daltonics, Germany). From January to June 2020, 205 preserved strains and 302 clinical isolates were used for comparison. Bacteria were tested with duplicates of the direct transfer method, and formic acid extraction was performed if the results were not at the species level. Fungi were tested with formic acid extraction followed by ethanol extraction methods. 16S rRNA or ITS region sequence analysis was performed on isolates that could not be identified by any of the instruments and on isolates that showed inconsistent results. The time to result of each instrument was also compared. Among preserved strains, species-level identification results were obtained in 202 (98.5%) strains by the Autof ms1000 and 200 (97.6%) strains by the Bruker Biotyper. Correct identification at the species/complex level was obtained for 200 (97.6%) strains by the Autof ms1000 and for 199 (97.1%) strains by the Bruker Biotyper. Among clinical isolates, species-level identification results were obtained in 301 (99.7%) strains and 300 (99.3%) strains by the Autof ms1000 and Bruker Biotyper, respectively. Correct identification at the species/complex level was achieved for 299 (99.0%) strains by the Autof ms1000 and for 300 (99.3%) strains by the Bruker Biotyper. The time to analyze 96 spots was approximately 14 min for the Autof ms1000 and approximately 27 min for the Bruker Biotyper. The two instruments showed comparable performance for the routine identification of clinical microorganisms. In addition, the Autof ms1000 has a short test time, making it convenient for use in clinical microbiology laboratories.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Comparison of Autof Ms1000 and EXS3000 MALDI-TOF MS Platforms for Routine Identification of Microorganisms
    Xiong, Lijuan
    Long, Xu
    Ni, Lijun
    Wang, Lili
    Zhang, Yang
    Cui, Lili
    Guo, Jian
    Yang, Chunying
    INFECTION AND DRUG RESISTANCE, 2023, 16 : 913 - 921
  • [2] The MALDI BioTyper; Identification and classification of microorganisms by MALDI-TOF MS profiling
    Maier, T.
    Kostrzewa, M.
    Gvozdyak, O.
    Pusch, W.
    Shi, G.
    MOLECULAR & CELLULAR PROTEOMICS, 2006, 5 (10) : S262 - S262
  • [3] Identification of Nocardia, Streptomyces, and Tsukamurella using MALDI-TOF MS with the Bruker Biotyper
    Yarbrough, Melanie L.
    Lainhart, William
    Burnham, Carey-Ann D.
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2017, 89 (02) : 92 - 97
  • [4] Identification of Anaerobic Bacteria Using MALDI-TOF MS Bruker Biotyper System
    Hsu, Y-M S.
    Burnham, C-A D.
    LABORATORY INVESTIGATION, 2012, 92 : 387A - 387A
  • [5] Identification of Anaerobic Bacteria Using MALDI-TOF MS Bruker Biotyper System
    Hsu, Y-M S.
    Burnham, C-A D.
    MODERN PATHOLOGY, 2012, 25 : 387A - 387A
  • [6] Comparison of the BluePoint MoldID oligonucleotide array and Bruker Biotyper MALDI-TOF MS for the identification of filamentous fungi
    Chou, Chia-Hua
    Chao, Qiao-Ting
    Lu, Yun-Shan
    Lee, Tai-Fen
    Hsueh, Po-Ren
    Huang, Yu-Tsung
    Liao, Chun-Hsing
    JOURNAL OF CLINICAL MICROBIOLOGY, 2025, 63 (01)
  • [7] Evaluation of the Bruker Biotyper and VITEK MS MALDI-TOF MS systems for the identification of unusual and/or difficult-to-identify microorganisms isolated from clinical specimens
    E. McElvania TeKippe
    C.-A. D. Burnham
    European Journal of Clinical Microbiology & Infectious Diseases, 2014, 33 : 2163 - 2171
  • [8] Evaluation of the Bruker Biotyper and VITEK MS MALDI-TOF MS systems for the identification of unusual and/or difficult-to-identify microorganisms isolated from clinical specimens
    TeKippe, E. McElvania
    Burnham, C. -A. D.
    EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2014, 33 (12) : 2163 - 2171
  • [9] Comparison of the Vitek MS and Bruker Microflex LT MALDI-TOF MS platforms for routine identification of commonly isolated bacteria and yeast in the clinical microbiology laboratory
    Deak, Eszter
    Charlton, Carmen L.
    Bobenchik, April M.
    Miller, Shelley A.
    Pollett, Simon
    McHardy, Ian H.
    Wu, Max T.
    Garner, Omai B.
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2015, 81 (01) : 27 - 33
  • [10] Evaluation of Autof MS2600 and MBT Smart MALDI-TOF MS Systems for Routine Identification of Clinical Bacteria and Yeasts
    De Carolis, Elena
    Ivagnes, Vittorio
    Magri, Carlotta
    Falasca, Benedetta
    Spanu, Teresa
    Sanguinetti, Maurizio
    MICROORGANISMS, 2024, 12 (02)