Comparison of a New Cardiac Output Ultrasound Dilution Method With Thermodilution Technique in Adult Patients Under General Anesthesia

被引:29
作者
Tsutsui, Masato [1 ]
Matsuoka, Nobuhiro [1 ]
Ikeda, Takehiko [1 ]
Sanjo, Yoshimitsu [1 ]
Kazama, Tomiei [1 ]
机构
[1] Natl Def Med Coll, Dept Anesthesiol, Tokorozawa, Saitama 3598513, Japan
关键词
cardiac output; ultrasound velocity dilution; thermodilution; PULMONARY-ARTERY THERMODILUTION; PULSE DYE DENSITOMETRY; WAVE-FORM ANALYSIS; BLOOD-VOLUME; ACCESS FLOW; INDOCYANINE GREEN; VALIDATION; HEMODIALYSIS; SURGERY; DEVICE;
D O I
10.1053/j.jvca.2009.03.007
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Objective: The purpose of this study was to investigate the reliability of cardiac output (CO) measured by a new ultrasound dilution method (COud) in comparison with CO by pulmonary artery thermodilution (COtd) in adult patients undergoing surgery. Design: A prospective study. Setting: A university hospital, single institutional. Participants: Twenty-nine adult patients undergoing abdominal surgery. Measurements and Main Results: After approval of the institutional ethics review board, 29 adult patients were evaluated. After induction, radial and pulmonary artery catheters were inserted. A disposable extracorporeal AV loop was connected between existing arterial and central venous catheters. Reusable ultrasound sensors that measure changes in blood ultrasound velocity after dilution by isotonic saline were clamped onto the arterial and venous limbs of the loop. Ultrasound dilution (UD) measurements (COstatus; Transonic Systems, Inc, Ithaca, NY) were obtained by injecting 30 mL of body-temperature isotonic saline into the venous limb of the AV loop. An average of 3 COud and 5 COtd was obtained for comparison. Bland-Altman plot and correlation analysis were used for statistical comparison. A total of 142 comparison measurements were obtained. The correlation coefficient between the 2 techniques was r = 0.91. Bland-Altman analysis did not produce any significant bias (bias = 0.02, standard deviation = 0.56). The percentage error of these data was 23.53%. Conclusions: COud measurements agreed well with COtd. The results of this study indicated that COud might be interchangeable with conventional COtd in perioperative adult patients. (C) 2009 Elsevier Inc. All rights reserved.
引用
收藏
页码:835 / 840
页数:6
相关论文
共 27 条
  • [1] Cardiac output measurement by pulse dye densitometry in cardiac surgery
    Baulig, W
    Bernhard, EO
    Bettex, D
    Schmidlin, D
    Schmid, ER
    [J]. ANAESTHESIA, 2005, 60 (10) : 968 - 973
  • [2] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [3] Continuous cardiac output monitoring via arterial pressure waveform analysis following severe hemorrhagic shock in dogs
    Cooper, Edward S.
    Muir, William W.
    [J]. CRITICAL CARE MEDICINE, 2007, 35 (07) : 1724 - 1729
  • [4] A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques
    Critchley, LAH
    Critchley, JAJH
    [J]. JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (02) : 85 - 91
  • [5] Validation of a new arterial pulse contour-based cardiac output device
    de Waal, Eric E. C.
    Kalkman, Cor J.
    Rex, Steffen
    Buhre, Wolfgang F.
    [J]. CRITICAL CARE MEDICINE, 2007, 35 (08) : 1904 - 1909
  • [6] POWER AND SAMPLE-SIZE CALCULATIONS - A REVIEW AND COMPUTER-PROGRAM
    DUPONT, WD
    PLUMMER, WD
    [J]. CONTROLLED CLINICAL TRIALS, 1990, 11 (02): : 116 - 128
  • [7] FRIEDMAN LM, 1998, FUNDAMENTALS CLIN TR, P111
  • [8] Lithium dilution cardiac output measurement: A clinical assessment of central venous and peripheral venous indicator injection
    Garcia-Rodriguez, C
    Pittman, J
    Cassell, CH
    Sum-Ping, J
    El-Moalem, H
    Young, C
    Mark, JB
    [J]. CRITICAL CARE MEDICINE, 2002, 30 (10) : 2199 - 2204
  • [9] Reproducibility of double indicator dilution measurements of intrathoracic blood volume compartments, extravascular lung water, and liver function
    Godje, O
    Peyerl, M
    Seebauer, T
    Dewald, O
    Reichart, B
    [J]. CHEST, 1998, 113 (04) : 1070 - 1077
  • [10] ANALYSIS OF FACTORS AFFECTING THE VARIABILITY OF FICK VERSUS INDICATOR DILUTION MEASUREMENTS OF CARDIAC-OUTPUT
    HILLIS, LD
    FIRTH, BG
    WINNIFORD, MD
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1985, 56 (12) : 764 - 768