In vivo urethral dose measurements:: A method as verify high dose rate prostate treatments

被引:31
作者
Brezovich, IA [1 ]
Duan, J [1 ]
Pareek, PN [1 ]
Fiveash, J [1 ]
Ezekiel, M [1 ]
机构
[1] Univ Alabama, Dept Radiat Oncol, Birmingham, AL 35233 USA
关键词
prostate; brachytherapy; HDR; TLD; dose;
D O I
10.1118/1.1312811
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Radiation doses delivered in high dose rate (HDR) brachytherapy are susceptible to many inaccuracies and errors, including imaging, planning and delivery. Consequently, the dose delivered to the patient may deviate substantially from the treatment plan. We investigated the feasibility of using TLD measurements in the urethra to estimate the discrepancy in treatments for prostate cancer. The dose response of the 1 mm diam, 6 mm long LiF rods that we used for the in vivo measurements was calibrated with the Ir-192 HDR source, as well as a Co-60 teletherapy unit. A train of 20 rods contained in a sterile plastic tube was inserted into the urethral (Foley) catheter for the duration of a treatment fraction, and the measured doses were compared to the treatment plan. Initial results from a total of seven treatments in four patients show good agreement between theory and experiment. Analysis of any one treatment showed agreement within 11.7%+/-6.2% for the highest dose encountered in the central prostatic urethra, and within 10.4%+/-4.4% for the mean dose. Taking the average over all seven treatments shows agreement within 1.7% for the maximum urethral dose, and within 1.5% for the mean urethral dose. Based on these initial findings it seems that planned prostate doses can be accurately reproduced in the clinic. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)02510-4].
引用
收藏
页码:2297 / 2301
页数:5
相关论文
共 4 条
[1]   High dose-rate afterloading 192Iridium prostate brachytherapy:: Feasibility report [J].
Mate, TP ;
Gottesman, JE ;
Hatton, J ;
Gribble, M ;
Van Hollebeke, L .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 41 (03) :525-533
[2]  
MATE TP, 1994, HIGH DOSE RATE BRACH, P355
[3]   TLD dose measurement: A simplified accurate technique for the dose range from 0.5 cGy to 1000 cGy [J].
Yu, C ;
Luxton, G .
MEDICAL PHYSICS, 1999, 26 (06) :1010-1016
[4]   The impact of edema on planning 125I and 103Pd prostate implants [J].
Yue, N ;
Dicker, AP ;
Nath, R ;
Waterman, FM .
MEDICAL PHYSICS, 1999, 26 (05) :763-767