Simultaneous measurement of X-ray specular reflection and off-specular diffuse scattering from liquid surfaces using a two-dimensional pixel array detector: the liquid-interface reflectometer of BL37XU at SPring-8

被引:16
作者
Yano, Yohko F. [1 ]
Uruga, Tomoya
Tanida, Hajime
Toyokawa, Hidenori
Terada, Yasuko
Yamada, Hironari [2 ]
机构
[1] Ritsumeikan Univ, Res Org Sci & Engn, Shiga, Japan
[2] Ritsumeikan Univ, Dept Photon, Shiga, Japan
关键词
X-ray reflectivity; off-specular diffuse scattering; PILATUS; air/water interface; ionic liquids; IONIC LIQUIDS; CAPILLARY WAVES; FLUCTUATIONS; MONOLAYERS;
D O I
10.1107/S0909049510013087
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An X-ray reflectometer for simultaneous measurement of specular and off-specular reflection of liquid surfaces is described. The reflectometer, equipped with a two-dimensional single X-ray photon-counting pixel array detector obtained the full range of X-ray specular and off-specular reflections in an extremely short time (1 s). Both the specular and off-specular reflection of water exhibited good agreement with the predicted capillary-wave theory within the appropriate instrumental resolution. The approach is also demonstrated on an aqueous solution of 1-dodecyl-3-methylimidazolium chloride. The monolayer in which the dodecyl chain faces upwards and the Cl- anion locates next to the imidazolium ring formed on the water surface was found to be laterally homogeneous. The use of a pixel array detector will be particularly powerful for in situ measurements to investigate both out-of-plane and in-plane structures simultaneously, not only for liquid surfaces but also for other thin films.
引用
收藏
页码:511 / 516
页数:6
相关论文
共 33 条
[11]  
FURUKAWA Y, 2004, 8 INT C SYNCHR RAD I, P1229
[12]   Performance of single-photon-counting PILATUS detector modules [J].
Kraft, P. ;
Bergamaschi, A. ;
Broennimann, Ch. ;
Dinapoli, R. ;
Eikenberry, E. F. ;
Henrich, B. ;
Johnson, I. ;
Mozzanica, A. ;
Schlepuetz, C. M. ;
Willmott, P. R. ;
Schmitt, B. .
JOURNAL OF SYNCHROTRON RADIATION, 2009, 16 :368-375
[13]   Feedback system for improving the performance of extended x-ray absorption fine structure measurements [J].
Kudo, Togo ;
Tanida, Hajime .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2007, 78 (03)
[14]   Correlation between surface tension and void fraction in ionic liquids [J].
Larriba, Carlos ;
Yoshida, Yukihiro ;
de la Mora, Juan Fernandez .
JOURNAL OF PHYSICAL CHEMISTRY B, 2008, 112 (39) :12401-12407
[15]   X-ray reflectivity of diblock copolymer monolayers at the air/water interface [J].
Li, Z ;
Zhao, MW ;
Quinn, J ;
Rafailovich, MH ;
Sokolov, J ;
Lennox, RB ;
Eisenberg, A ;
Wu, XZ ;
Kim, MW ;
Sinha, SK ;
Tolan, M .
LANGMUIR, 1995, 11 (12) :4785-4792
[16]   X-ray Scattering From Liquid Surfaces: Effect of Resolution [J].
Pershan, P. S. .
JOURNAL OF PHYSICAL CHEMISTRY B, 2009, 113 (12) :3639-3646
[17]   Effects of thermal roughness on X-ray studies of liquid surfaces [J].
Pershan, PS .
COLLOIDS AND SURFACES A-PHYSICOCHEMICAL AND ENGINEERING ASPECTS, 2000, 171 (1-3) :149-157
[18]   X-RAY-SCATTERING STUDY OF CAPILLARY-WAVE FLUCTUATIONS AT A LIQUID SURFACE [J].
SANYAL, MK ;
SINHA, SK ;
HUANG, KG ;
OCKO, BM .
PHYSICAL REVIEW LETTERS, 1991, 66 (05) :628-631
[19]   A synchrotron x-ray liquid surface spectrometer [J].
Schlossman, ML ;
Synal, D ;
Guan, YM ;
Meron, M ;
Shea-McCarthy, G ;
Huang, ZQ ;
Acero, A ;
Williams, SM ;
Rice, SA ;
Viccaro, PJ .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1997, 68 (12) :4372-4384
[20]   THERMAL DIFFUSE X-RAY-SCATTERING STUDIES OF THE WATER-VAPOR INTERFACE [J].
SCHWARTZ, DK ;
SCHLOSSMAN, ML ;
KAWAMOTO, EH ;
KELLOGG, GJ ;
PERSHAN, PS ;
OCKO, BM .
PHYSICAL REVIEW A, 1990, 41 (10) :5687-5690