Photon beam energy dependent single-arc volumetric modulated arc optimization

被引:1
作者
Momin, Shadab [1 ,2 ]
Grafe, James L. [2 ]
Georgiou, Konstantinos [3 ]
Khan, Rao F. [1 ]
机构
[1] Washington Univ, Sch Med, Dept Radiat Oncol, St Louis, MO 63130 USA
[2] Ryerson Univ, Dept Phys, Toronto, ON, Canada
[3] Ryerson Univ, Dept Math, Toronto, ON, Canada
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2021年 / 82卷
关键词
Single arc VMAT; Multiple photon energy optimization; Convex relaxation; IMRT; RADIATION-THERAPY; INTENSITY; IMRT;
D O I
10.1016/j.ejmp.2021.02.004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this work was to present a new single-arc mixed photon (6&18MV) VMAT (SAMP) optimization framework that concurrently optimizes for two photon energies with corresponding partial arc lengths. & nbsp; Methods and materials: Owing to simultaneous optimization of energy dependent intensity maps and corresponding arc locations, the proposed model poses nonlinearity. Unique relaxation constraints based on McCormick approximations were introduced for linearization. Energy dependent intensity maps were then decomposed to generate apertures. Feasibility of the proposed framework was tested on a sample of ten prostate cancer cases with lateral separation ranging from 34 cm (case no.1) to 52 cm (case no.6). The SAMP plans were compared against single energy (6MV) VMAT (SE) plans through dose volume histograms (DVHs) and radiobiological parameters including normal tissue complication probability (NTCP) and equivalent uniform dose (EUD). Results: The contribution of higher energy photon beam optimized by the algorithm demonstrated an increase for cases with a lateral separation >40 cm. SAMP?VMAT notably improved bladder and rectum sparing in large size cases. Compared to single energy, SAMP?VMAT plans reduced bladder and rectum NTCP in cases with large lateral separation. With the exception of one case, SAMP?VMAT either improved or maintained femoral heads compared to SE?VMAT. SAMP?VMAT reduced the nontarget tissue integral dose in all ten cases. Conclusions: A single-arc VMAT optimization framework comprising mixed photon energy partial arcs was presented. Overall results underline the feasibility and potential of the proposed approach for improving OAR sparing in large size patients without compromising the target homogeneity and coverage.
引用
收藏
页码:122 / 133
页数:12
相关论文
共 34 条
[1]  
Burman CM, 2002, FRONT RADIAT THER ON, V37, P151
[2]   A generalized inverse planning tool for volumetric-modulated arc therapy [J].
Cao, Daliang ;
Afghan, Muhammad K. N. ;
Ye, Jinsong ;
Chen, Fan ;
Shepard, David M. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (21) :6725-6738
[3]   The tradeoff between treatment plan quality and required number of monitor units in intensity-modulated radiotherapy [J].
Craft, David ;
Suess, Philipp ;
Bortfeld, Thomas .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 67 (05) :1596-1605
[4]   Multicriteria VMAT optimization [J].
Craft, David ;
McQuaid, Dualta ;
Wala, Jeremiah ;
Chen, Wei ;
Salari, Ehsan ;
Bortfeld, Thomas .
MEDICAL PHYSICS, 2012, 39 (02) :686-696
[5]   State of dose prescription and compliance to international standard (ICRU-83) in intensity modulated radiation therapy among academic institutions [J].
Das, Indra J. ;
Andersen, Aaron ;
Chen, Zhe ;
Dimofte, Andrea ;
Glatstein, Eli ;
Hoisak, Jeremy ;
Huang, Long ;
Langer, Mark P. ;
Lee, Choonik ;
Pacella, Matthew ;
Popple, Richard A. ;
Rice, Roger ;
Smilowitz, Jennifer ;
Sponseller, Patricia ;
Zhu, Timothy .
PRACTICAL RADIATION ONCOLOGY, 2017, 7 (02) :E145-E155
[6]   CERR: A computational environment for radiotherapy research [J].
Deasy, JO ;
Blanco, AI ;
Clark, VH .
MEDICAL PHYSICS, 2003, 30 (05) :979-985
[7]   4π Non-Coplanar Liver SBRT: A Novel Delivery Technique [J].
Dong, Peng ;
Lee, Percy ;
Ruan, Dan ;
Long, Troy ;
Romeijn, Edwin ;
Yang, Yingli ;
Low, Daniel ;
Kupelian, Patrick ;
Sheng, Ke .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 85 (05) :1360-1366
[8]  
Grant M., 2009, CVX-Matlab software for disciplinedconvex programming
[9]  
Gu Z, 2018, GUROBI OPTIMIZER VER
[10]   Proportion of Cancer Cases Attributable to Excess Body Weight by US State, 2011-2015 [J].
Islami, Farhad ;
Sauer, Ann Goding ;
Gapstur, Susan M. ;
Jemal, Ahmedin .
JAMA ONCOLOGY, 2019, 5 (03) :384-392