Evaluation of the i-STAT portable clinical analyzer in chickens (Gallus gallus)

被引:47
作者
Steinmetz, Hanspeter W.
Vogt, Rainer
Kastner, Sabine
Riond, Barbara
Hatt, Jean-Michel
机构
[1] Univ Zurich, Vetsuisse Fac, Div Zoo Anim & Exot Pets, CH-8057 Zurich, Switzerland
[2] Univ Zurich, Vetsuisse Fac, Div Anesthesiol, Zurich, Switzerland
[3] Univ Zurich, Vetsuisse Fac, Div Clin Lab, Zurich, Switzerland
关键词
avian; blood gas; electrolytes; emergency; point-of-care;
D O I
10.1177/104063870701900407
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
The i-STAT portable clinical analyzer (PCA) was evaluated for performance in avian species. With the EG7+ cartridge, which provided results for hydrogen ion concentration, oxygen tension, carbon dioxide tension, sodium, potassium, ionized calcium, hematocrit, and various calculated parameters, analytical accuracy and precision were tested by comparing obtained values to those of established traditional blood gas and chemistry analyzers. Deming's regression and bias plots were used to compare i-STAT results with those obtained by laboratory professionals using benchtop analyzers. The reliability of the i-STAT PCA with EG7+ cartridges was good, with 0-5.7% system failures in measured Values. Regression statistics were good for all blood gas analytes and acceptable for electrolytes and calculated parameters, except for potassium and base excess, for which the regression data or the discrepancy between the methods was too large. The system was reliable and easy to use and had an overall acceptable accuracy in avian species. These features, together with portability and small required blood volumes, make the i-STAT suitable for point-of-care use in critical avian patients, although single values require careful interpretation.
引用
收藏
页码:382 / 388
页数:7
相关论文
共 19 条
[2]  
ERICKSON KA, 1993, CLIN CHEM, V39, P283
[3]  
Fudge Alan M., 2000, P9
[4]  
Genao A, 1996, LAB ANIM SCI, V46, P399
[5]  
GREGORY RT, 2005, BIRD ERYTHROCYTE SIZ
[6]   Temporal effects of 3 commonly used anticoagulants on hematologic and biochemical variables in blood samples from macaws and Burmese python']pythons [J].
Harr, KE ;
Raskin, RE ;
Heard, DJ .
VETERINARY CLINICAL PATHOLOGY, 2005, 34 (04) :383-388
[7]  
JACOBS E, 1993, CLIN CHEM, V39, P1069
[8]  
Jones R.B. Payne., 1997, Clinical investigation and statistics in laboratory medicine
[9]   Performance of a digital PCO2/SpO2 ear sensor [J].
Kocher S. ;
Rohling R. ;
Tschupp A. .
Journal of Clinical Monitoring and Computing, 2004, 18 (2) :75-79
[10]   Blood values of juvenile northern elephant seals (Mirounga angustirostris) obtained using a portable clinical analyzer [J].
Larsen, RS ;
Haulena, M ;
Grindem, CB ;
Gulland, FMD .
VETERINARY CLINICAL PATHOLOGY, 2002, 31 (03) :106-110