A Diagnostic Model for Screening Diabetic Retinopathy Using the Hand-Held Electroretinogram Device RETeval

被引:8
作者
Deng, Xiaowen [1 ,2 ]
Li, Zijing [1 ,2 ]
Zeng, Peng [1 ,2 ]
Wang, Jing [1 ,2 ]
Liang, Jiaqi [1 ,2 ]
Lan, Yuqing [1 ,2 ]
机构
[1] Sun Yat Sen Univ, Sun Yat Sen Mem Hosp, Dept Ophthalmol, Guangzhou, Peoples R China
[2] Sun Yat Sen Univ, Sun Yat Sen Mem Hosp, Guangdong Prov Key Lab Malignant Tumor Epigenet &, Guangzhou, Peoples R China
来源
FRONTIERS IN ENDOCRINOLOGY | 2021年 / 12卷
关键词
diabetic retinopathy; electroretinogram; diagnostic model; risk factor; decision tree; FUNDUS PIGMENTATION; ISCEV STANDARD; MYDRIASIS-FREE; HYPERGLYCEMIA;
D O I
10.3389/fendo.2021.632457
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Purpose: To construct a proper model to screen for diabetic retinopathy (DR) with the RETeval. Method: This was a cross-sectional study. Two hundred thirty-two diabetic patients and seventy controls were recruited. The DR risk assessment protocol was performed to obtain subjects' DR risk score using the RETeval. Afterwards, the receiver operating characteristic (ROC) curve was used to determine the best cutoff for diagnosing DR. Random forest and decision tree models were constructed. Results: With increasing DR severity, the DR score gradually increased. When the DR score was used to diagnose DR, the ROC curve had an area under the curve of 0.881 (95% confidence interval: 0.836-0.927, P < 0.001), with a best cutoff value of 22.95, a sensitivity of 74.3% (95 CI: 66.0%similar to 82.6%), and a specificity of 90.6% (95 CI: 83.7%similar to 94.8%). The top four risk factors selected by the random forest were used to construct the decision tree for diagnosing DR, which had a sensitivity of 93.3% (95% CI: 86.3%similar to 97.0%) and a specificity of 80.3% (95% CI: 72.1%similar to 86.6%). Conclusions: The DR risk assessment protocol combined with the decision tree model was innovatively used to evaluate the risk of DR, improving the sensitivity of diagnosis, which makes this method more suitable than the current protocol for DR screening.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Diagnostic Accuracy of Hand-Held Fundus Camera and Artificial Intelligence in Diabetic Retinopathy Screening
    Tomic, Martina
    Vrabec, Romano
    Hendelja, Durdica
    Kolaric, Vilma
    Bulum, Tomislav
    Rahelic, Dario
    BIOMEDICINES, 2024, 12 (01)
  • [2] Validity, Usefulness and Cost of RETeval System for Diabetic Retinopathy Screening
    Al-Otaibi, Humoud
    Al-Otaibi, Mohammed D.
    Khandekar, Rajiv
    Souru, Ches
    Al-Abdullah, Abdulelah A.
    Al-Dhibi, Hassan
    Stone, Donald U.
    Kozak, Igor
    TRANSLATIONAL VISION SCIENCE & TECHNOLOGY, 2017, 6 (03):
  • [3] Diagnostic test accuracy of diabetic retinopathy screening by physician graders using a hand-held non-mydriatic retinal camera at a tertiary level medical clinic
    Mapa Mudiyanselage Prabhath Nishantha Piyasena
    Jennifer L. Y. Yip
    David MacLeod
    Min Kim
    Venkata S. Murthy Gudlavalleti
    BMC Ophthalmology, 19
  • [4] Diagnostic test accuracy of diabetic retinopathy screening by physician graders using a hand-held non-mydriatic retinal camera at a tertiary level medical clinic
    Piyasena, Mapa Mudiyanselage Prabhath Nishantha
    Yip, Jennifer L. Y.
    MacLeod, David
    Kim, Min
    Gudlavalleti, Venkata S. Murthy
    BMC OPHTHALMOLOGY, 2019, 19 (1)
  • [5] Screening for diabetic retinopathy in diabetic patients with a mydriasis-free, full-field flicker electroretinogram recording device
    Zeng, Yunkao
    Cao, Dan
    Yang, Dawei
    Zhuang, Xuenan
    Yu, Honghua
    Hu, Yunyan
    Zhang, Yan
    Yang, Cheng
    He, Miao
    Zhang, Liang
    DOCUMENTA OPHTHALMOLOGICA, 2020, 140 (03) : 211 - 220
  • [6] Screening for diabetic retinopathy in diabetic patients with a mydriasis-free, full-field flicker electroretinogram recording device
    Yunkao Zeng
    Dan Cao
    Dawei Yang
    Xuenan Zhuang
    Honghua Yu
    Yunyan Hu
    Yan Zhang
    Cheng Yang
    Miao He
    Liang Zhang
    Documenta Ophthalmologica, 2020, 140 : 211 - 220
  • [7] Development and Validation of a Diabetic Retinopathy Screening Modality Using a Hand-Held Nonmydriatic Digital Retinal Camera by Physician Graders at a Tertiary-Level Medical Clinic: Protocol for a Validation Study
    Piyasena, Mapa Mudiyanselage Prabhath Nishantha
    Gudlavalleti, Venkata S. Murthy
    Gilbert, Clare
    Yip, Jennifer L. Y.
    Peto, Tunde
    MacLeod, David
    Fonseka, Charith
    Kulatunga, Aruna
    Bandutilake
    Dhanapala, Mangala
    Pathirana, Lalani
    Dissanayake, Heshani
    JMIR RESEARCH PROTOCOLS, 2018, 7 (12):
  • [8] Mapping of retinal function in diabetic retinopathy using the multifocal electroretinogram
    Palmowski, AM
    Sutter, EE
    Bearse, MA
    Fung, W
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1997, 38 (12) : 2586 - 2596
  • [9] Light and portable novel device for diabetic retinopathy screening
    Ting, Daniel S. W.
    Franzco, Mei Ling Tay-Kearney
    Kanagasingam, Yogesan
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2012, 40 (01) : E40 - E46
  • [10] Screening for Diabetic Retinopathy Using an Automated Diagnostic System Based on Deep Learning: Diagnostic Accuracy Assessment
    Rego, Silvia
    Dutra-Medeiros, Marco
    Soares, Filipe
    Monteiro-Soares, Matilde
    OPHTHALMOLOGICA, 2021, 244 (03) : 250 - 257